ABSTRACT
Clinical MRI systems use magnetic fields of at least 0.5T to take advantage of the increase in signal-to-noise ratio (SNR) with B(0). Low-field MRI apparatus is less expensive and offers the potential benefit of improved T(1) contrast between tissues. The poor inherent SNR at low field can be offset by incorporating prepolarizing field pulses with the MRI pulse sequence. The prepolarizing field does not need to be as homogeneous as the detection field, so it can be generated by a relatively inexpensive electromagnet. Prepolarizing hardware for a 0.01T MRI system was developed together with a prepolarized MRI pulse sequence that incorporates fast imaging techniques to reduce acquisition times by a factor of 5 relative to standard methods. Comparison images of test objects show that most of the enhanced SNR is retained with the fast method. Low-field images of a human wrist acquired using the fast prepolarized method are also shown.
Subject(s)
Magnetic Resonance Imaging/methods , Wrist/anatomy & histology , Equipment Design , Humans , Magnetic Resonance Imaging/instrumentation , Phantoms, ImagingABSTRACT
A portable, nuclear magnetic resonance (NMR) probe is described which utilises the intrinsic inhomogeneity of the field produced by a single-sided magnet to provide spatial encoding of the NMR signal. The probe uses a longitudinally magnetized hollow cylinder, and a figure-8 radiofrequency (RF) surface coil. The system has been used to measure NMR relaxation times and one-dimensional NMR profiles of rubber phantoms.
Subject(s)
Nuclear Magnetic Resonance, Biomolecular/instrumentation , Equipment Design , Phantoms, Imaging , Rubber , Signal Processing, Computer-AssistedABSTRACT
We describe the use of liquid helium-cooled (4.2 K) gradiometer coils and a DC superconducting quantum interference device (SQUID) preamplifier to improve the SNR of magnetic resonance imaging (MRI) at 0.01 T. Gradiometer windings are used both to reduce lossy interactions with the MRI system's room temperature magnet and gradient coils and also to reject interference from more distant sources, which reduces the need for RF shielding. We have tested both axial and planar (figure-of-eight) gradiometer configurations. The figure-of-eight gradiometer has a more rapid fall-off in sensitivity with increasing distance from its windings than the axial gradiometer, but this is compensated for by reduced lossy interactions and improved interference rejection. We have used the system to image the human arm.