ABSTRACT
A vortex-assisted liquid-liquid micro-extraction coupled with high-performance liquid chromatography, with UV-vis, is proposed to pre-concentrate methyl methacrylate and to improve separation in biological matrices. The use of 1-octanol as extracting phase, its volume, the need for a dispersant agent, the agitation conditions and the cooling time before phase separation were evaluated. In optimum conditions, enrichment factors of 20 (±0.5) and enrichment recovery of 99% were obtained. The straightforward association of this extraction process with the HPLC method, previously regulated by the International Organization for Standardization, afforded a detection limit of 122 ng/mL and a quantification limit of 370 ng/mL. The within-batch precision, relative standard deviation, was 3% for a sample with 1.49 µg/mL and 4% for a sample with 13.4 µg/mL. The results showed a between batch-precision of 21% for experiments performed on five different days, for a sample with a concentration of 1.10 µg/mL in methyl methacrylate.