Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 5 de 5
Filter
Add more filters










Database
Language
Publication year range
2.
Med Phys ; 42(4): 1528-37, 2015 Apr.
Article in English | MEDLINE | ID: mdl-25832043

ABSTRACT

PURPOSE: The dosimetric properties of the OCTAVIUS Detector 1500 (OD1500) ionization chamber array (PTW-Freiburg, Freiburg, Germany) have been investigated. A comparative study was carried out with the OCTAVIUS Detector 729 and OCTAVIUS Detector 1000 SRS arrays. METHODS: The OD1500 array is an air vented ionization chamber array with 1405 detectors in a 27 × 27 cm(2) measurement area arranged in a checkerboard pattern with a chamber-to-chamber distance of 10 mm in each row. A sampling step width of 5 mm can be achieved by merging two measurements shifted by 5 mm, thus fulfilling the Nyquist theorem for intensity modulated dose distributions. The stability, linearity, and dose per pulse dependence were investigated using a Semiflex 31013 chamber (PTW-Freiburg, Freiburg, Germany) as a reference detector. The effective depth of measurement was determined by measuring TPR curves with the array and a Roos chamber type 31004 (PTW-Freiburg, Freiburg, Germany). Comparative output factor measurements were performed with the array, the Semiflex 31010 ionization chamber and the Diode 60012 (both PTW-Freiburg, Freiburg, Germany). The energy dependence of the OD1500 was measured by comparing the array's readings to those of a Semiflex 31010 ionization chamber for varying mean photon energies at the depth of measurement, applying to the Semiflex chamber readings the correction factor kNR for nonreference conditions. The Gaussian lateral dose response function of a single array detector was determined by searching the convolution kernel suitable to convert the slit beam profiles measured with a Diode 60012 into those measured with the array's central chamber. An intensity modulated dose distribution measured with the array was verified by comparing a OD1500 measurement to TPS calculations and film measurements. RESULTS: The stability and interchamber sensitivity variation of the OD1500 array were within ±0.2% and ±0.58%, respectively. Dose linearity was within 1% over the range from 5 to 1000 MU. The effective point of measurement of the OD1500 for dose measurements in RW3 phantoms was determined to be (8.7 ± 0.2) mm below its front surface. Output factors showed deviations below 1% for field sizes exceeding 4 × 4 cm(2). The dose per pulse dependence was smaller than 0.4% for doses per pulse from 0.2 to 1 mGy. The energy dependence of the array did not exceed ±0.9%. The parameter σ of the Gaussian lateral dose response function was determined as σ6MV = (2.07 ± 0.02) mm for 6 MV and σ15MV = (2.09 ± 0.02) mm for 15 MV. An IMRT verification showed passing rates well above 90% for a local 3 mm/3% criterion. CONCLUSIONS: The OD1500 array's dosimetric properties showed the applicability of the array for clinical dosimetry with the possibility to increase the spatial sampling frequency and the coverage of a dose distribution with the sensitive areas of ionization chambers by merging two measurements.


Subject(s)
Photons/therapeutic use , Radiotherapy, Intensity-Modulated/instrumentation , Phantoms, Imaging , Radiometry/instrumentation , Radiotherapy Dosage
3.
Med Phys ; 40(8): 082106, 2013 Aug.
Article in English | MEDLINE | ID: mdl-23927343

ABSTRACT

PURPOSE: In this work, the properties of the two-dimensional liquid filled ionization chamber array Octavius 1000SRS (PTW-Freiburg, Germany) for use in clinical photon-beam dosimetry are investigated. METHODS: Measurements were carried out at an Elekta Synergy and Siemens Primus accelerator. For measurements of stability, linearity, and saturation effects of the 1000SRS array a Semiflex 31013 ionization chamber (PTW-Freiburg, Germany) was used as a reference. The effective point of measurement was determined by TPR measurements of the array in comparison with a Roos chamber (type 31004, PTW-Freiburg, Germany). The response of the array with varying field size and depth of measurement was evaluated using a Semiflex 31010 ionization chamber as a reference. Output factor measurements were carried out with a Semiflex 31010 ionization chamber, a diode (type 60012, PTW-Freiburg, Germany), and the detector array under investigation. The dose response function for a single detector of the array was determined by measuring 1 cm wide slit-beam dose profiles and comparing them against diode-measured profiles. Theoretical aspects of the low pass properties and of the sampling frequency of the detector array were evaluated. Dose profiles measured with the array and the diode detector were compared, and an intensity modulated radiation therapy (IMRT) field was verified using the Gamma-Index method and the visualization of line dose profiles. RESULTS: The array showed a short and long term stability better than 0.1% and 0.2%, respectively. Fluctuations in linearity were found to be within ±0.2% for the vendor specified dose range. Saturation effects were found to be similar to those reported in other studies for liquid-filled ionization chambers. The detector's relative response varied with field size and depth of measurement, showing a small energy dependence accounting for maximum signal deviations of ±2.6% from the reference condition for the setup used. The σ-values of the Gaussian dose response function for a single detector of the array were found to be (0.72±0.25) mm at 6 MV and (0.74±0.25) mm at 15 MV and the corresponding low pass cutoff frequencies are 0.22 and 0.21 mm(-1), respectively. For the inner 5×5 cm2 region and the outer 11×11 cm2 region of the array the Nyquist theorem is fulfilled for maximum sampling frequencies of 0.2 and 0.1 mm(-1), respectively. An IMRT field verification with a Gamma-Index analysis yielded a passing rate of 95.2% for a 3 mm∕3% criterion with a TPS calculation as reference. CONCLUSIONS: This study shows the applicability of the Octavius 1000SRS in modern dosimetry. Output factor and dose profile measurements illustrated the applicability of the array in small field and stereotactic dosimetry. The high spatial resolution ensures adequate measurements of dose profiles in regular and intensity modulated photon-beam fields.


Subject(s)
Radiometry/instrumentation , Linear Models , Photons , Radiotherapy, Intensity-Modulated
4.
Med Phys ; 39(6Part11): 3729, 2012 Jun.
Article in English | MEDLINE | ID: mdl-28517124

ABSTRACT

PURPOSE: The finite extension of an ionization chamber gives rise to a spatial averaging effect, known as the "volume effect". In order to provide the appropriate corrections, the response functions along its lateral and longitudinal directions are characterized using Gaussian distributions, whose standard deviations slat and slong have been determined for a large set of clinical dosimeters. METHODS: Nine cylindrical ionization chambers, two parallel-plate chambers and two 2D ionization chamber arrays have been examined by scanning rectangular photon fields along their short axes. The true profiles D(x) were known from scans with a small Si diode. The ionization chambers were aligned with their symmetry axes either perpendicular or parallel to the scan direction in order to obtain slat and slong separately. In a search process, D(x) was numerically convolved with normalized one-dimensional Gaussian kernels K(x) of varying s. The best fit between the convolution product D(x) * K(x) and the measured profile M(x) of the ionization chamber was used to determine parameters slat and slong of the Gaussian kernels. RESULTS: For both the lateral and longitudinal directions, very good agreement was found between M(x) and the convolution products of D(x) with Gaussian kernels K(x). For all chambers, their 2s values are similar to the cavity dimensions, which means that the "tails" of the Gaussian response functions reach into the exterior of the chambers, - an effect of the ranges of the secondary electrons. At higher photon energies response functions K(x) are slightly wider, but no detectable depth dependence has been observed. CONCLUSIONS: We have shown that the response functions of ionization chambers can be described by Gaussian distributions, confirming earlier observations, and we determined their standard deviations in both the lateral and longitudinal directions. Using these response functions, appropriate correction methods determined to eliminate the volume effect can be applied.

5.
Med Phys ; 39(6Part11): 3729, 2012 Jun.
Article in English | MEDLINE | ID: mdl-28517138

ABSTRACT

PURPOSE: The volume effect of ionization chambers gives rise to a spatial averaging effect that can be expressed mathematically as the convolution of the true dose profile with the detector's response function. The latter has been shown to be best described by Gaussian distribution. Based on this knowledge, the volume effect correction factor kV is derived. METHODS: To derive kV, a sixth degree polynomial is fitted to the true dose profile: D(x) = a0 + a2×2 + a4×4 + a6×6. The measured dose profile M(x) is calculated as the convolution product of D(x) with a one-dimensional normalized Gauss function with standard deviation s. Therefore kV at the dose maximum has the value D(0)/M(0), which is a function of the coefficients a0,2,4,6 and the detector specific s. In the case where D(x) is unknown, kV can be derived analogously from M(x) so that M(x) = b0 + b2×2 + b4×4 + b6×6, where kV can now be expressed as a function of the coefficients b0,2,4,6 and s. RESULTS: The magnitudes of kV,lat and kV,long were calculated for 1 to 5 cm dose profiles using measured s values, both in the lateral and the longitudinal directions, for a set of common ionization chambers. At field widths above 2 cm, the values of kV,lat fall below 1.01 for all the chambers evaluated, whereas it needs field widths above 4 cm to get all values of kV,long below 1.01. Since the detector's signal is integrated over the sensitive volume, the total kV can be calculated as kV,total = kV,lat . kV,long. CONCLUSIONS: In this work, a correction is developed to eliminate the volume effect of ionization chambers when they are positioned in the maxima of dose profiles, particularly for the performance of output factor measurements for the calibration of narrow photon beams.

SELECTION OF CITATIONS
SEARCH DETAIL