Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 15 de 15
Filter
Add more filters










Publication year range
1.
Biosens Bioelectron ; 16(9-12): 621-9, 2001 Dec.
Article in English | MEDLINE | ID: mdl-11679237

ABSTRACT

A device providing frequent, automatic, and non-invasive glucose measurements for persons with diabetes has been developed: the GlucoWatch biographer. This device extracts glucose through intact skin via reverse iontophoresis where it is detected by an amperometric biosensor. The biographer can provide glucose readings every 20 min for 12 h. The performance of this device was evaluated in two large clinical studies in a controlled clinical environment (n=231), and the home environment (n=124). Accuracy of the biographer was evaluated by comparing the automatic biographer readings to serial finger-stick blood glucose (BG) measurements. Biographer performance was comparable in both environments. Mean difference between biographer and finger-stick measurements was -0.01 and 0.26 mmol l(-1) for the clinical and home environments, respectively. The mean absolute value of the relative difference was 1.06 and 1.18 mmol l(-1) for the same studies. Correlation coefficient (r) between biographer and finger-stick measurements was 0.85 and 0.80 for the two studies. In both studies, over 94% of the biographer readings were in the clinically acceptable A+B region of the Clarke Error Grid. A slight positive bias is observed for the biographer readings at low BG levels. Biographer accuracy is relatively constant over all rates of BG changes, except when BG decreases more than 10 mmol l(-1) h(-1), which occurred for only 0.2% of points in the home environment study. Biographer precision, as measured by CV%, is approx. 10%. Skin irritation, characterized by erythema and edema, was either non-existent or mild in >90% of subjects and resolved in virtually all subjects without treatment in several days.


Subject(s)
Blood Glucose Self-Monitoring/instrumentation , Diabetes Mellitus/blood , Blood Glucose Self-Monitoring/statistics & numerical data , Diabetes Mellitus/metabolism , Equipment Design , Glucose/metabolism , Humans , Iontophoresis , Skin/metabolism
2.
Diabetes Care ; 24(5): 881-5, 2001 May.
Article in English | MEDLINE | ID: mdl-11347748

ABSTRACT

OBJECTIVE: Hypoglycemia is a common acute complication of diabetes therapy. The GlucoWatch biographer provides frequent and automatic glucose measurements with an adjustable low-glucose alarm. We have analyzed the performance of the biographer low-glucose alarm relative to hypoglycemia as defined by blood glucose < or = 3.9 mmol/l. RESEARCH DESIGN AND METHODS: The analysis was based on 1,091 biographer uses from four clinical trials. which generated 14,487 paired (biographer and blood glucose) readings. RESULTS: The results show that as the low-glucose alert level of the biographer is increased, the number of true positive alerts (alarm sounds and blood glucose < or = 3.9 mmol/l) and false positive alerts (alarm sounds but blood glucose >3.9 mmol/l) increased. When analyzed as a function of varying low-glucose alert levels, the results show receiver operator characteristic curves consistent with a highly useful diagnostic tool. Setting the alert level from 1.1 to 1.7 mmol/l above the level of concern is likely to optimize the trade-off between true positives and false positives for each user. When the same blood glucose data are analyzed for typical monitoring practices (two or four measurements per day), the results show that fewer hypoglycemic events are detected than those detected with the biographer.


Subject(s)
Blood Glucose Self-Monitoring/instrumentation , Blood Glucose/analysis , Hypoglycemia/diagnosis , Monitoring, Ambulatory/instrumentation , Automation , Blood Glucose Self-Monitoring/methods , False Negative Reactions , False Positive Reactions , Humans , Hypoglycemia/blood , Hypoglycemia/prevention & control , Hypoglycemic Agents/adverse effects , Monitoring, Ambulatory/methods , ROC Curve , Reproducibility of Results , Sensitivity and Specificity
3.
Biomaterials ; 21(18): 1837-45, 2000 Sep.
Article in English | MEDLINE | ID: mdl-10919687

ABSTRACT

This study investigated the in vitro degradation of porous poly(DL-lactic-co-glycolic acid) (PLGA) foams during a 20-week period in pH 7.4 phosphate-buffered saline (PBS) at 37 degrees C and their in vivo degradation following implantation in rat mesentery for up to 8 weeks. Three types of PLGA 85 : 15 and three types of 50 : 50 foams were fabricated using a solvent-casting, particulate-leaching technique. The two types had initial salt weight fraction of 80 and 90%, and a salt particle size of 106-150 microm, while the third type had 90% initial weight fraction of salt in the size range 0-53 microm. The porosities of the resulting foams were 0.82, 0.89, and 0.85 for PLGA 85 : 15, and 0.73, 0.87, and 0.84 for PLGA 50 : 50 foams, respectively. The corresponding median pore diameters were 30, 50, and 17 microm for PLGA 85: 15, and 19, 17, and 17 microm for PLGA 50 : 50. The in vitro and in vivo degradation kinetics of PLGA 85: 15 foams were independent of pore morphology with insignificant variation in foam weight, thickness, pore distribution, compressive creep behavior, and morphology during degradation. The in vitro foam half-lives based on the weight average molecular weight were 11.1 +/- 1.8 (80%, 106-150 microm), 12.0 +/- 2.0 (90%, 106-150 microm), and 11.6 +/- 1.3 (90%, 0-53 microm) weeks, similar to the corresponding values of 9.4 +/- 2.2, 14.3 +/- 1.5, and 13.7 +/- 3.3 weeks for in vivo degradation. In contrast, all PLGA 50 : 50 foams exhibited significant change in foam weight, water absorption, and pore distribution after 6-8 weeks of incubation with PBS. The in vitro foam half-lives were 3.3 +/- 0.3 (80%, 106-150 microm), 3.0 +/- 0.3 (90%, 106-150 microm), and 3.2 +/- 0.1 (90%, 0-53 microm) weeks, and the corresponding in vivo half-lives were 1.9 micro 0.1, 2.2 +/- 0.2, and 2.4 +/- 0.2 weeks. The significantly shorter half-lives of PLGA 50: 50 compared to 85: 15 foams indicated their faster degradation both in vitro and in vivo. In addition, PLGA 50: 50 foams exhibited significantly faster degradation in vivo as compared to in vitro conditions due to an autocatalytic effect of the accumulated acidic degradation products in the medium surrounding the implants. These results suggest that the polymer composition and environmental conditions have significant effects on the degradation rate of porous PLGA foams.


Subject(s)
Biocompatible Materials/chemistry , Biocompatible Materials/pharmacokinetics , Lactic Acid/chemistry , Lactic Acid/pharmacokinetics , Polyglycolic Acid/chemistry , Polyglycolic Acid/pharmacokinetics , Polymers/chemistry , Polymers/pharmacokinetics , Animals , Biodegradation, Environmental , Half-Life , Hot Temperature , Hydrogen-Ion Concentration , Kinetics , Mercury , Mesentery/cytology , Mesentery/pathology , Polylactic Acid-Polyglycolic Acid Copolymer , Prostheses and Implants , Rats , Thermodynamics
4.
Biomaterials ; 21(15): 1595-605, 2000 Aug.
Article in English | MEDLINE | ID: mdl-10885732

ABSTRACT

This study investigated the in vitro degradation of porous poly(L-lactic acid) (PLLA) foams during a 46-week period in pH 7.4 phosphate-buffered saline at 37 degrees C. Four types of PLLA foams were fabricated using a solvent-casting, particulate-leaching technique. The three types had initial salt weight fraction of 70, 80, and 90%, and a salt particle size of 106-150 microm, while the fourth type had 90% initial weight fraction of salt in the size range 0-53 microm. The porosities of the resulting foams were 0.67, 0.79, 0.91, and 0.84, respectively. The corresponding median pore diameters were 33, 52, 91, and 34 microm. The macroscopic degradation of PLLA foams was independent of pore morphology with insignificant variation in foam weight, thickness, pore distribution, compressive creep behavior, and morphology during degradation. However, decrease in melting temperature and slight increase in crystallinity were observed at the end of degradation. The foam half-lives based on the weight average molecular weight were 11.6+/-0.7 (70%, 106-150 microm), 15.8+/-1.2 (80%, 106-150 microm), 21.5+/-1.5 (90%, 106-150 microm), and 43.0+/-2.7 (90%, 0-53 microm) weeks. The thicker pore walls of foams prepared with 70 or 80% salt weight fraction as compared to those with 90% salt weight fraction contributed to an autocatalytic effect resulting in faster foam degradation. Also, the increased pore surface/volume ratio of foams prepared with salt in the range 0-53 microm enhanced the release of degradation products thus diminishing the autocatalytic effect and resulting in slower foam degradation compared to those with salt in the range 106-150 microm. Formation and release of crystalline PLLA particulates occurred for foams fabricated with 90% salt weight fraction at early stages of degradation. These results suggest that the degradation rate of porous foams can be engineered by varying the pore wall thickness and pore surface/volume ratio.


Subject(s)
Biocompatible Materials/chemistry , Lactic Acid/chemistry , Polymers/chemistry , Calorimetry, Differential Scanning , Chromatography, Gel , Drug Delivery Systems , Drug Stability , Microscopy, Electron, Scanning , Molecular Weight , Polyesters
5.
Ann Med ; 32(9): 632-41, 2000 Dec.
Article in English | MEDLINE | ID: mdl-11209971

ABSTRACT

The GlucoWatch (Cygnus, Inc, Redwood City, CA, USA) biographer provides automatic, frequent and noninvasive blood glucose measurements for up to 12 h. The device extracts glucose through intact skin where it is measured by an amperometric biosensor. Clinical trials in a variety of environments have shown that the biographer provides accurate and precise glucose measurements when compared with serial fingerstick blood glucose measurements. Mean difference between these measurements was 0.26 mmol/L in the home environment (r = 0.80). Over 94% of biographer readings were in the clinically acceptable A+B region of the Clarke Error Grid. A slight positive bias is observed for the biographer readings at low glucose levels. Biographer precision, as measured by coefficient of variation (CV)%, is approximately 10%. The low glucose alert function of the biographer was able to detect up to 75% of hypoglycaemic episodes with a low false alert level. Skin irritation, characterized by erythema and oedema was either nonexistent or mild in over 87% of subjects and resolved in virtually all subjects without treatment in several days. The GlucoWatch biographer has been shown to be a safe and effective method to track glucose level trends and patterns, which should enable improved glycaemic control for many patients.


Subject(s)
Blood Glucose Self-Monitoring , Diabetes Mellitus/blood , Analysis of Variance , Biosensing Techniques , Blood Glucose Self-Monitoring/instrumentation , Clinical Trials as Topic , Equipment Design , Female , Humans , Hypoglycemia/diagnosis , Male , Middle Aged
6.
Diabetes Technol Ther ; 2(2): 199-207, 2000.
Article in English | MEDLINE | ID: mdl-11469259

ABSTRACT

BACKGROUND: Improved glycemic control significantly reduces long-term microvascular complications of diabetes mellitus associated with chronic hyperglycemia. The GlucoWatch biographer is designed to facilitate intensive diabetes management by providing automatic, frequent, and noninvasive glucose readings up to three times per hour for as long as 12 hours. METHODS: The device extracts glucose through intact skin using reverse iontophoresis and measures the extracted glucose with an electrochemical biosensor. A clinical trial was performed to assess the effect of acetaminophen, a potential interference for traditional blood glucose meters, on the accuracy of the GlucoWatch biographer in adult subjects with diabetes (n = 18). One thousand milligram doses of acetaminophen were administered to subjects in two groups: one to achieve Cmax (peak acetominophen concentration) at the time of biographer calibration and the other to achieve Cmax during the measurement period. The biographer readings were compared to serial fingerstick blood glucose measurements. RESULTS: Time profiles over 9 hours show close tracking of the biographer glucose results with fingerstick blood glucose measurements for all groups. The mean difference between the two measurements is between 8 and 12 mg/dL for all groups. The mean absolute value of the relative difference is less than 20%, and more than 93% of the points were in the clinically acceptable (A+B) region of the Clarke Error Grid. No statistically significant differences were found for any accuracy measurement across all groups. CONCLUSIONS: The GlucoWatch Biographer provides frequent measurements of glucose over a 12-hour period with high accuracy. No effect of therapeutic dosage of acetaminophen on the accuracy of the glucose readings was found.


Subject(s)
Acetaminophen/pharmacology , Blood Glucose Self-Monitoring/methods , Blood Glucose/analysis , Monitoring, Ambulatory/instrumentation , Monitoring, Ambulatory/methods , Adult , Analysis of Variance , Automation , Biosensing Techniques , Blood Glucose Self-Monitoring/instrumentation , Calibration , Electrochemistry , Equipment Design , Humans , Reproducibility of Results , Time Factors , United States , White People
7.
JAMA ; 282(19): 1839-44, 1999 Nov 17.
Article in English | MEDLINE | ID: mdl-10573275

ABSTRACT

CONTEXT: Intensive diabetes management using frequent blood glucose measurements to guide therapy has been shown to significantly improve short- and long-term outcomes. Development of a device that makes possible frequent, automatic, painless, and accurate measurements of glucose would facilitate intensive management. OBJECTIVE: To determine the accuracy of the GlucoWatch automatic glucose biographer (Cygnus Inc) compared with that of serial blood glucose measurements. DESIGN: Multicenter comparative study of the GlucoWatch biographer and the HemoCue blood glucose analyzer (Aktiebolaget Leo) performed between August 29 and October 17, 1998. Participants wore up to 2 biographers during the 15-hour study session and performed 2 fingersticks per hour for comparative blood glucose measurements. The biographers were calibrated with a single HemoCue measurement after a 3-hour warm-up period. Diet and insulin were manipulated to produce a broad glycemic range during the study. SETTING: Controlled clinical environment at 2 diabetes centers and 3 contract research organizations in the United States. PARTICIPANTS: A total of 92 subjects (mean [SD] age, 42.1 [15.1] years; 59.8% women) with type 1 or 2 diabetes requiring treatment with insulin. MAIN OUTCOME MEASURES: Mean error, mean absolute error, correlation, slope, and intercept using Deming regression, and clinical significance of differences between biographer readings and blood glucose measurements using the Clarke error grid. RESULTS: Results showed close tracking of blood glucose over a range of 2.2 to 22.2 mmol/L (40-400 mg/dL) for up to 12 hours using a single point calibration. The biographer readings lagged behind serial blood glucose values by a mean of 18 minutes. An analysis of 2167 data pairs shows a linear relationship (r = 0.88; slope = 1.03; intercept = -0.33 mmol/L [-6 mg/dL]) between biographer readings and serial glucose measurements. The mean absolute error between the 2 measurements was 15.6% (mean error [SD], -0.07 [1.82] mmol/L [-1 [33] mg/dL]), and 96.8% of the data fell in the therapeutically relevant regions of the error grid analysis. CONCLUSION: These results demonstrate close agreement between GlucoWatch biographer readings and blood glucose measurements using repeated fingerstick blood samples. The automatic, frequent, and noninvasive measurements obtained with the biographer provides more information about glucose levels than the current standard of care.


Subject(s)
Biosensing Techniques , Blood Glucose Self-Monitoring/instrumentation , Adult , Blood Glucose/analysis , Diabetes Mellitus/prevention & control , Female , Humans , Male
8.
Diabetes Care ; 22(10): 1708-14, 1999 Oct.
Article in English | MEDLINE | ID: mdl-10526740

ABSTRACT

OBJECTIVE: The purpose of this study was to compare measurements of glucose obtained via iontophoretic extraction with the GlucoWatch automatic glucose biographer (Cygnus, Inc., Redwood City, CA) with capillary blood glucose values that were determined 1) in a controlled outpatient clinic setting and 2) in a home setting. RESEARCH DESIGN AND METHODS: There were 76 GlucoWatch biographers used on 28 different young adults (21 women and 7 men) with type 1 diabetes (age 30.9 +/- 6.9 years and duration of diabetes 18.4 +/- 8.1 years [mean +/- SD]) in a controlled outpatient clinic setting. Some subjects participated on multiple days. Subjects wore two GlucoWatch biographers, each on the forearm (ventral aspect). Comparisons were made to HemoCue blood glucose analyzer (Aktiebolgat Leo, Helsingborg, Sweden) capillary blood glucose measurements. In addition, GlucoWatch biographers (one each day for 3 consecutive days) were used by 12 subjects (8 women, 4 men) in a home setting. Comparisons were made to capillary blood glucose values determined using the One Touch Profile meter (Johnson & Johnson, New Brunswick, NJ). RESULTS: GlucoWatch biographer glucose values correlated well with capillary blood glucose values determined using the HemoCue analyzer in the clinic setting (r = 0.90, 1,554 paired data points) and using the One Touch Profile meter in the home setting (r = 0.85, 204 paired data points). When 36 subjects wore two biographers simultaneously, the correlation between the two biographers was r = 0.94. The error grid analysis demonstrated that > 96% of biographer glucose values determined in the clinic or home setting were in the clinically acceptable A and B regions. CONCLUSIONS: This study confirms the accuracy and precision of glucose values as determined using the GlucoWatch biographer in clinic and home settings.


Subject(s)
Blood Glucose Self-Monitoring/instrumentation , Blood Glucose Self-Monitoring/methods , Diabetes Mellitus, Type 1/blood , Adult , Capillaries , Equipment Design , Female , Fingers/blood supply , Humans , Male , Regression Analysis , Reproducibility of Results
10.
Nat Med ; 1(11): 1198-201, 1995 Nov.
Article in English | MEDLINE | ID: mdl-7584995

ABSTRACT

Results from the Diabetes Care and Complications Trial show that tight blood glucose control significantly reduces the long-term complications of diabetes mellitus. In that study, frequent self-testing of glucose and insulin administration resulted in a significant reduction in long-term complications. This protocol, however, also resulted in a threefold increase in the frequency of hypoglycaemic incidents. Currently, self-testing requires a drop of blood for each measurement. The pain and inconvenience of self-testing, along with the fear and danger of hypoglycaemia has led to poor patient acceptance of a tight control regimen, despite the clear long-term advantages. A continuously worn, noninvasive method to periodically measure glucose would provide a convenient and comfortable means of frequent self-testing. A continuously worn device could also alert the user of low glucose levels, thereby reducing the incidence of hypoglycaemia. Guy et al. demonstrated a noninvasive method to transport glucose through the skin using low-level electrical current. To provide a quantitative measurement, the flux of glucose extracted across the skin must correlate with serum glucose in a predictive manner. The results presented here show a quantitative relationship between serum and transdermally extracted glucose in diabetics.


Subject(s)
Blood Glucose Self-Monitoring/methods , Blood Glucose/analysis , Diabetes Mellitus/blood , Iontophoresis/methods , Adult , Diabetes Mellitus, Type 1/blood , Diabetes Mellitus, Type 2/blood , Female , Humans , Hyperglycemia/prevention & control , Hypoglycemia/prevention & control , Male , Middle Aged , Reproducibility of Results , Skin/metabolism
11.
J Biomed Mater Res ; 28(3): 387-95, 1994 Mar.
Article in English | MEDLINE | ID: mdl-8077254

ABSTRACT

Intracranial studies to analyze the degradation kinetics of the bioerodible polymer poly[bis(p-carboxyphenoxy)propane-sebacic acid] [p(CPP-SA) 20:80] copolymer wafers were conducted in a rat model. Rats were separated into four groups: those receiving 1) polymer, 2) polymer loaded with the chemotherapeutic agent BCNU, 3) drug-loaded polymer with previous tumor implantation, and 4) polymer and an absorbable hemostatic material. A polymer wafer was surgically implanted into the brain of each animal. Residual polymer was harvested at varying times for chromatographic analysis. In vitro effects of pH, mixing, and water availability on degradation were also studied. The results of in vitro and in vivo studies were compared to understand the behavior of polymers in a clinical setting. We found that degradation of p(CPP-SA) initially occurred more slowly in vivo than in vitro. The presence of BCNU, tumor, and absorbable hemostatic material did not affect the ultimate time of polymer degradation in vivo, and the intrinsic polymer degradation time of 1 mm thick p(CPP-SA) 20:80 disks in vivo was 6-8 weeks.


Subject(s)
Brain Neoplasms/drug therapy , Decanoic Acids/pharmacokinetics , Drug Carriers/pharmacokinetics , Gliosarcoma/drug therapy , Polyesters/pharmacokinetics , Animals , Biodegradation, Environmental , Carmustine/administration & dosage , Carmustine/therapeutic use , Cellulose, Oxidized , Delayed-Action Preparations , Drug Evaluation, Preclinical , Drug Implants , Foreign-Body Reaction/etiology , Inflammation , Male , Rats , Rats, Inbred F344
12.
Proc Natl Acad Sci U S A ; 90(2): 552-6, 1993 Jan 15.
Article in English | MEDLINE | ID: mdl-8421690

ABSTRACT

Degradable polymers are beginning to play an increasing role as materials for environmental and medical applications. Understanding factors that control erosion, such as bond cleavage and the dissolution and diffusion of degradation products, will be critical to the future development of these materials. Erosion kinetics, photomicroscopy, and infrared spectroscopy were used to understand the erosion mechanism of two families of degradable polymers, polyanhydrides and polyesters. Polyanhydrides exhibit behavior more characteristic of surface erosion, whereas the polyesters exhibit bulk erosion patterns. Control of erosion times from a few days to several years can be achieved by a judicious choice of monomer units and bond selection.


Subject(s)
Polymers/chemistry , Refuse Disposal , Anhydrides/chemistry , Polyesters/chemistry
13.
Pharm Res ; 9(10): 1279-83, 1992 Oct.
Article in English | MEDLINE | ID: mdl-1448426

ABSTRACT

We report a study of the effects of polymer molecular weight on the erosion of polyanhydride copolymer matrices composed of 1,3-bis(p-carboxyphenoxy)-propane (CPP) and sebacic acid (SA) in aqueous solution. The erosion profile characteristically displays an induction period during which the erosion rate is relatively slow. The length of this period depends on the initial molecular weight of the polymer. The induction period may be characterized as a time during which a rapid decrease in polymer molecular weight occurs, the end of this period correlating with the time required for the polymer molecular weight to decrease to below a value of approximately 5000 (MW).


Subject(s)
Dicarboxylic Acids , Polymers/chemistry , Decanoic Acids/chemistry , Drug Delivery Systems , Hydroxybenzoate Ethers , Hydroxybenzoates/chemistry , Kinetics , Molecular Weight , Spectrophotometry, Ultraviolet
14.
Proc Natl Acad Sci U S A ; 88(21): 9377-81, 1991 Nov 01.
Article in English | MEDLINE | ID: mdl-1946348

ABSTRACT

The stability of protein-based pharmaceuticals (e.g., insulin) is important for their production, storage, and delivery. To gain an understanding of insulin's aggregation mechanism in aqueous solutions, the effects of agitation rate, interfacial interactions, and insulin concentration on the overall aggregation rate were examined. Ultraviolet absorption spectroscopy, high-performance liquid chromatography, and quasielastic light scattering analyses were used to monitor the aggregation reaction and identify intermediate species. The reaction proceeded in two stages; insulin stability was enhanced at higher concentration. Mathematical modeling of proposed kinetic schemes was employed to identify possible reaction pathways and to explain greater stability at higher insulin concentration.


Subject(s)
Insulin/chemistry , Adsorption , Animals , Cattle , In Vitro Techniques , Kinetics , Models, Chemical , Polytetrafluoroethylene , Protein Binding , Protein Denaturation , Solubility , Thermodynamics , Water
15.
Biomaterials ; 11(9): 738-45, 1990 Nov.
Article in English | MEDLINE | ID: mdl-2090313

ABSTRACT

Biomaterials have made a great impact on medicine. However, numerous challenges remain. This paper discusses three representative areas involving important medical problems. First, drug delivery systems; major considerations include drug-polymer interactions, drug transformation, diffusion properties of drugs and, if degradation occurs, of polymer degradation products through polymer matrices developing a more complete understanding of matrix degradation in the case of erodible polymers and developing new engineered polymers designed for specific purposes such as vaccination or pulsatile release. Second, cell-polymer interactions, including the fate of inert polymers, the use of polymers as templates for tissue regeneration and the study of polymers which aid cell transplantation. Third, orthopaedic biomaterials, including basic research in the behaviour of chondrocytes, osteocytes and connective tissue-free interfaces and applied research involving computer-aided design of biomaterials and the creation of orthopaedic biomaterials.


Subject(s)
Biocompatible Materials , Chemical Engineering/trends , Delayed-Action Preparations , Drug Implants , Equipment Design , Forecasting , Materials Testing , Orthopedic Equipment/trends
SELECTION OF CITATIONS
SEARCH DETAIL
...