Results 1 - 20 de 76
Filter
1.
Eur Radiol ; 32(2): 1227-1237, 2022 Feb.
Article in English | MEDLINE | ID: mdl-34327581

ABSTRACT

OBJECTIVES: To assess task-based image quality for two abdominal protocols on various CT scanners. To establish a relationship between diagnostic reference levels (DRLs) and task-based image quality. METHODS: A protocol for the detection of focal liver lesions was used to scan an anthropomorphic abdominal phantom containing 8- and 5-mm low-contrast (20 HU) spheres at five CTDIvol levels (4, 8, 12, 16, and 20 mGy) on 12 CTs. Another phantom with high-contrast calcium targets (200 HU) was scanned at 2, 4, 6, 10, and 15 mGy using a renal stones protocol on the same CTs. To assess the detectability, a channelized Hotelling observer was used for low-contrast targets and a non-prewhitening observer with an eye filter was used for high contrast targets. The area under the ROC curve and signal to noise ratio were used as figures of merit. RESULTS: For the detection of 8-mm spheres, the image quality reached a high level (mean AUC over all CTs higher than 0.95) at 11 mGy. For the detection of 5-mm spheres, the AUC never reached a high level of image quality. Variability between CTs was found, especially at low dose levels. For the search of renal stones, the AUC was nearly maximal even for the lowest dose level. CONCLUSIONS: Comparable task-based image quality cannot be reached at the same dose level on all CT scanners. This variability implies the need for scanner-specific dose optimization. KEY POINTS: • There is an image quality variability for subtle low-contrast lesion detection in the clinically used dose range. • Diagnostic reference levels were linked with task-based image quality metrics. • There is a need for specific dose optimization for each CT scanner and clinical protocol.


Subject(s)
Diagnostic Reference Levels , Tomography, X-Ray Computed , Algorithms , Humans , Phantoms, Imaging , Radiation Dosage , Signal-To-Noise Ratio , Tomography Scanners, X-Ray Computed
2.
Radiat Prot Dosimetry ; 195(3-4): 289-295, 2021 Oct 12.
Article in English | MEDLINE | ID: mdl-33647105

ABSTRACT

Nationwide surveys on radiation dose to the population from medical imaging are recommended in order to follow trends in population exposure. The goal of the 2018 survey was to investigate the current exposure. The invoice coding information was collected in five university hospitals and large clinics. To improve the estimation of the effective dose delivered in computed tomography (CT), we collected dose data from different Dose Archiving Communication Systems. On average, we found that 1.2 radiological examinations per year and per inhabitant were performed. Dental radiography was the most frequent examination (48% of all the X-ray examinations), followed by conventional radiography (36%) and CT (11%). The average annual effective dose was estimated to be 1.48 mSv per inhabitant, with CT representing 64% of that dose. Our results show that the exposure of the Swiss population from medical imaging has remained stable since 2013, despite a 15% increase in the number of CT examinations.


Subject(s)
Diagnostic Tests, Routine , Tomography, X-Ray Computed , Humans , Radiation Dosage , Radiography , X-Rays
3.
J Med Imaging (Bellingham) ; 7(4): 045501, 2020 Jul.
Article in English | MEDLINE | ID: mdl-32743016

ABSTRACT

Purpose: Visual search using volumetric images is becoming the standard in medical imaging. However, we do not fully understand how eye movement strategies mediate diagnostic performance. A recent study on computed tomography (CT) images showed that the search strategies of radiologists could be classified based on saccade amplitudes and cross-quadrant eye movements [eye movement index (EMI)] into two categories: drillers and scanners. Approach: We investigate how the number of times a radiologist scrolls in a given direction during analysis of the images (number of courses) could add a supplementary variable to use to characterize search strategies. We used a set of 15 normal liver CT images in which we inserted 1 to 5 hypodense metastases of two different signal contrast amplitudes. Twenty radiologists were asked to search for the metastases while their eye-gaze was recorded by an eye-tracker device (EyeLink1000, SR Research Ltd., Mississauga, Ontario, Canada). Results: We found that categorizing radiologists based on the number of courses (rather than EMI) could better predict differences in decision times, percentage of image covered, and search error rates. Radiologists with a larger number of courses covered more volume in more time, found more metastases, and made fewer search errors than those with a lower number of courses. Our results suggest that the traditional definition of drillers and scanners could be expanded to include scrolling behavior. Drillers could be defined as scrolling back and forth through the image stack, each time exploring a different area on each image (low EMI and high number of courses). Scanners could be defined as scrolling progressively through the stack of images and focusing on different areas within each image slice (high EMI and low number of courses). Conclusions: Together, our results further enhance the understanding of how radiologists investigate three-dimensional volumes and may improve how to teach effective reading strategies to radiology residents.

4.
Phys Med ; 76: 28-37, 2020 Aug.
Article in English | MEDLINE | ID: mdl-32574999

ABSTRACT

PURPOSE: We aimed to thoroughly characterize image quality of a novel deep learning image reconstruction (DLIR), and investigate its potential for dose reduction in abdominal CT in comparison with filtered back-projection (FBP) and a partial model-based iterative reconstruction (ASiR-V). METHODS: We scanned a phantom at three dose levels: regular (7 mGy), low (3 mGy) and ultra-low (1 mGy). Images were reconstructed using DLIR (low, medium and high levels) and ASiR-V (0% = FBP, 50% and 100%). Noise and contrast-dependent spatial resolution were characterized by computing noise power spectra and target transfer functions, respectively. Detectability indexes of simulated acute appendicitis or colonic diverticulitis (low contrast), and calcium-containing urinary stones (high contrast) (|ΔHU| = 50 and 500, respectively) were calculated using the nonprewhitening with eye filter model observer. RESULTS: At all dose levels, increasing DLIR and ASiR-V levels both markedly decreased noise magnitude compared with FBP, with DLIR low and medium maintaining noise texture overall. For both low- and high-contrast spatial resolution, DLIR not only maintained, but even slightly enhanced spatial resolution in comparison with FBP across all dose levels. Conversely, increasing ASiR-V impaired low-contrast spatial resolution compared with FBP. Overall, DLIR outperformed ASiR-V in all simulated clinical scenarios. For both low- and high-contrast diagnostic tasks, increasing DLIR substantially enhanced detectability at any dose and contrast levels for any simulated lesion size. CONCLUSIONS: Unlike ASiR-V, DLIR substantially reduces noise while maintaining noise texture and slightly enhancing spatial resolution overall. DLIR outperforms ASiR-V by enabling higher detectability of both low- and high-contrast simulated abdominal lesions across all investigated dose levels.


Subject(s)
Deep Learning , Radiographic Image Interpretation, Computer-Assisted , Algorithms , Image Processing, Computer-Assisted , Radiation Dosage , Tomography, X-Ray Computed
5.
Dentomaxillofac Radiol ; 49(6): 20190468, 2020 Sep 01.
Article in English | MEDLINE | ID: mdl-32267774

ABSTRACT

OBJECTIVES: The aim of this study was to establish diagnostic reference levels (DRLs) in the field of dental maxillofacial and ear-nose-throat (ENT) practices using cone beam CT (CBCT) in Switzerland. METHODS: A questionnaire was sent to owners of CBCTs in Switzerland; to a total of 612 institutions. The answers were analyzed for each indication, provided that enough data were available. The DRLs were defined as the 75th percentile of air kerma product distribution (PKA). RESULTS: 227 answers were collected (38% of all centers). Third quartile of PKA values were obtained for five dental indications: 662 mGy cm² for wisdom tooth, 683 mGy cm² for single tooth implant treatment, 542 mGy cm² for tooth position anomalies, 569 mGy cm² for pathological dentoalveolar modifications, and 639 mGy cm² for endodontics. The standard field of view (FOV) size of 5 cm in diameter x 5 cm in height was proposed. CONCLUSIONS: Large ranges of FOV and PKA were found for a given indication, demonstrating the importance of establishing DRLs as well as FOV recommendations in view of optimizing the present practice. For now, only DRLs for dental and maxillofacial could be defined; because of a lack of ENT data, no DRL values for ENT practices could be derived from this survey.


Subject(s)
Cone-Beam Computed Tomography , Radiation Dosage , Reference Values , Surveys and Questionnaires , Switzerland
6.
EJNMMI Phys ; 7(1): 1, 2020 Jan 06.
Article in English | MEDLINE | ID: mdl-31907664

ABSTRACT

BACKGROUND: We assessed and compared image quality obtained with clinical 18F-FDG whole-body oncologic PET protocols used in three different, state-of-the-art digital PET/CT and two conventional PMT-based PET/CT devices. Our goal was to evaluate an  improved trade-off between administered activity (patient dose exposure/signal-to-noise ratio) and acquisition time (patient comfort) while preserving diagnostic information achievable with the recently introduced digital detector technology compared to previous analogue PET technology. METHODS: We performed list-mode (LM) PET acquisitions using a NEMA/IEC NU2 phantom, with activity concentrations of 5 kBq/mL and 25 kBq/mL for the background (9.5 L) and sphere inserts, respectively. For each device, reconstructions were obtained varying the image statistics (10, 30, 60, 90, 120, 180, and 300 s from LM data) and the number of iterations (range 1 to 10) in addition to the employed local clinical protocol setup. We measured for each reconstructed dataset: the quantitative cross-calibration, the image noise on the uniform background assessed by the coefficient of variation (COV), and the recovery coefficients (RCs) evaluated in the hot spheres. Additionally, we compared the characteristic time-activity-product (TAP) that is the product of scan time per bed position × mass-activity administered (in min·MBq/kg) across datasets. RESULTS: Good system cross-calibration was obtained for all tested datasets with < 6% deviation from the expected value was observed. For all clinical protocol settings, image noise was compatible with clinical interpretation (COV < 15%). Digital PET showed an improved background signal-to-noise ratio as compared to conventional PMT-based PET. RCs were comparable between digital and PMT-based PET datasets. Compared to PMT-based PET, digital systems provided comparable image quality with lower TAP (from ~ 40% less and up to 70% less). CONCLUSIONS: This study compared the achievable clinical image quality in three state-of-the-art digital PET/CT devices (from different vendors) as well as in two conventional PMT-based PET. Reported results show that a comparable image quality is achievable with a TAP reduction of ~ 40% in digital PET. This could lead to a significant reduction of the administered mass-activity and/or scan time with direct benefits in terms of dose exposure and patient comfort.

7.
J Med Imaging (Bellingham) ; 6(2): 025501, 2019 Apr.
Article in English | MEDLINE | ID: mdl-31131292

ABSTRACT

Task-based image quality procedures in CT that substitute a human observer with a model observer usually use single-slice images with uniform backgrounds from homogeneous phantoms. However, anatomical structures and inhomogeneities in organs generate noise that can affect the detection performance of human observers. The purpose of this work was to assess the impact of background type, uniform or liver, and the viewing modality, single- or multislice, on the detection performance of human and model observers. We collected abdominal CT scans from patients and homogeneous phantom scans in which we digitally inserted low-contrast signals that mimicked a liver lesion. We ran a rating experiment with the two background conditions with three signal sizes and three human observers presenting images in two reading modalities: single- and multislice. In addition, channelized Hotelling observers (CHO) for single- and multislice detection were implemented and evaluated according to their degree of correlation with the human observer performance. For human observers, there was a small but significant improvement in performance with multislice compared to the single-slice viewing mode. Our data did not reveal a significant difference between uniform and anatomical backgrounds. Model observers demonstrated a good correlation with human observers for both viewing modalities. Human observers have very similar performances in both multi- and single-slice viewing mode. It is therefore preferable to use single-slice CHO as this model is computationally more tractable than multislice CHO. However, using images from a homogeneous phantom can result in overestimating image quality as CHO performance tends to be higher in uniform than anatomical backgrounds, while human observers have similar detection performances.

8.
J Cardiovasc Magn Reson ; 21(1): 11, 2019 02 07.
Article in English | MEDLINE | ID: mdl-30728035

ABSTRACT

BACKGROUND: Our objectives were first to determine the optimal coronary computed tomography angiography (CTA) protocol for the quantification and detection of simulated coronary artery cross-sectional area (CSA) differences in vitro, and secondly to quantitatively compare the performance of the optimized CTA protocol with a previously validated radial coronary cardiovascular magnetic resonance (CMR) technique. METHODS: 256-multidetector CTA and radial coronary CMR were used to obtain images of a custom in vitro resolution phantom simulating a range of physiological responses of coronary arteries to stress. CSAs were automatically quantified and compared with known nominal values to determine the accuracy, precision, signal-to-noise ratio (SNR), and circularity of CSA measurements, as well as the limit of detection (LOD) of CSA differences. Various iodine concentrations, radiation dose levels, tube potentials, and iterative image reconstruction algorithms (ASiR-V) were investigated to determine the optimal CTA protocol. The performance of the optimized CTA protocol was then compared with a radial coronary CMR method previously developed for endothelial function assessment under both static and moving conditions. RESULTS: The iodine concentration, dose level, tube potential, and reconstruction algorithm all had significant effects (all p <  0.001) on the accuracy, precision, LOD, SNR, and circularity of CSA measurements with CTA. The best precision, LOD, SNR, and circularity with CTA were achieved with 6% iodine, 20 mGy, 100 kVp, and 90% ASiR-V. Compared with the optimized CTA protocol under static conditions, radial coronary CMR was less accurate (- 0.91 ± 0.13 mm2 vs. -0.35 ± 0.04 mm2, p <  0.001), but more precise (0.08 ± 0.02 mm2 vs. 0.21 ± 0.02 mm2, p <  0.001), and enabled the detection of significantly smaller CSA differences (0.16 ± 0.06 mm2 vs. 0.52 ± 0.04 mm2; p <  0.001; corresponding to CSA percentage differences of 2.3 ± 0.8% vs. 7.4 ± 0.6% for a 3-mm baseline diameter). The same results held true under moving conditions as CSA measurements with CMR were less affected by motion. CONCLUSIONS: Radial coronary CMR was more precise and outperformed CTA for the specific task of detecting small CSA differences in vitro, and was able to reliably identify CSA changes an order of magnitude smaller than those reported for healthy physiological vasomotor responses of proximal coronary arteries. However, CTA yielded more accurate CSA measurements, which may prove useful in other clinical scenarios, such as coronary artery stenosis assessment.


Subject(s)
Computed Tomography Angiography , Coronary Angiography/methods , Coronary Vessels/diagnostic imaging , Endothelium, Vascular/diagnostic imaging , Magnetic Resonance Imaging , Multidetector Computed Tomography , Computed Tomography Angiography/instrumentation , Contrast Media , Coronary Angiography/instrumentation , Coronary Circulation , Humans , Limit of Detection , Magnetic Resonance Imaging/instrumentation , Meglumine , Multidetector Computed Tomography/instrumentation , Organometallic Compounds , Phantoms, Imaging , Predictive Value of Tests , Reproducibility of Results , Vasodilation
9.
Sci Rep ; 8(1): 17734, 2018 12 07.
Article in English | MEDLINE | ID: mdl-30531988

ABSTRACT

To investigate the impact of a partial model-based iterative reconstruction (ASiR-V) on image quality in thoracic oncologic multidetector computed tomography (MDCT), using human and mathematical model observers. Twenty cancer patients examined with regular-dose thoracic-abdominal-pelvic MDCT were retrospectively included. Thoracic images reconstructed using a sharp kernel and filtered back-projection (reference) or ASiR-V (0-100%, 20% increments; follow-up) were analysed by three thoracic radiologists. Advanced quantitative physical metrics, including detectability indexes of simulated 4-mm-diameter solid non-calcified nodules and ground-glass opacities, were computed at regular and reduced doses using a custom-designed phantom. All three radiologists preferred higher ASiR-V levels (best = 80%). Increasing ASiR-V substantially decreased noise magnitude, with slight changes in noise texture. For high-contrast objects, changing the ASiR-V level had no major effect on spatial resolution; whereas for lower-contrast objects, increasing ASiR-V substantially decreased spatial resolution, more markedly at reduced dose. For both high- and lower-contrast pulmonary lesions, detectability remained excellent, regardless of ASiR-V and dose levels, and increased significantly with increasing ASiR-V levels (all p < 0.001). While high ASiR-V levels (80%) are recommended to detect solid non-calcified nodules and ground-glass opacities in regular-dose thoracic oncologic MDCT, care must be taken because, for lower-contrast pulmonary lesions, high ASiR-V levels slightly change noise texture and substantially decrease spatial resolution, more markedly at reduced dose.


Subject(s)
Multidetector Computed Tomography/methods , Radiographic Image Interpretation, Computer-Assisted/methods , Radiography, Thoracic/methods , Aged , Aged, 80 and over , Algorithms , Artifacts , Female , Humans , Male , Middle Aged , Pelvis/pathology , Phantoms, Imaging , Radiation Dosage , Radionuclide Imaging/methods , Retrospective Studies , Signal-To-Noise Ratio
10.
Eur Radiol ; 28(12): 5203-5210, 2018 Dec.
Article in English | MEDLINE | ID: mdl-29858638

ABSTRACT

OBJECTIVE: We investigated the variability in diagnostic information inherent in computed tomography (CT) images acquired at 68 different CT units, with the selected acquisition protocols aiming to answer the same clinical question. METHODS: An anthropomorphic abdominal phantom with two optional rings was scanned on 68 CT systems from 62 centres using the local clinical acquisition parameters of the portal venous phase for the detection of focal liver lesions. Low-contrast detectability (LCD) was assessed objectively with channelised Hotelling observer (CHO) using the receiver operating characteristic (ROC) paradigm. For each lesion size, the area under the ROC curve (AUC) was calculated and considered as a figure of merit. The volume computed tomography dose index (CTDIvol) was used to indicate radiation dose exposure. RESULTS: The median CTDIvol used was 5.8 mGy, 10.5 mGy and 16.3 mGy for the small, medium and large phantoms, respectively. The median AUC obtained from clinical CT protocols was 0.96, 0.90 and 0.83 for the small, medium and large phantoms, respectively. CONCLUSIONS: Our study used a model observer to highlight the difference in image quality levels when dealing with the same clinical question. This difference was important and increased with growing phantom size, which generated large variations in patient exposure. In the end, a standardisation initiative may be launched to ensure comparable diagnostic information for well-defined clinical questions. The image quality requirements, related to the clinical question to be answered, should be the starting point of patient dose optimisation. KEY POINTS: • Model observers enable to assess image quality objectively based on clinical tasks. • Objective image quality assessment should always include several patient sizes. • Clinical diagnostic image quality should be the starting point for patient dose optimisation. • Dose optimisation by applying DRLs only is insufficient for ensuring clinical requirements.


Subject(s)
Abdomen/diagnostic imaging , Phantoms, Imaging , Radiation Exposure/analysis , Tomography, X-Ray Computed/methods , Humans , ROC Curve , Radiation Dosage
11.
J Radiol Prot ; 38(3): 1013-1036, 2018 Sep.
Article in English | MEDLINE | ID: mdl-29786616

ABSTRACT

PURPOSE: Definition of new national diagnostic reference levels (DRLs) for volume computed tomography dose index (CTDIvol) and dose length product (DLP) for neuro-paediatric CT examinations depending on the medical indication. METHODS: Paediatric cranial CT data sets acquired between January 2013 and December 2016 were retrospectively collected between July 2016 and March 2017 from eight of the largest university and cantonal hospitals that perform most of the neuro-paediatric CTs in Switzerland. A consensus review of CTDIvol and DLP was undertaken for three defined anatomical regions: brain, facial bone, and petrous bone, each with and without contrast medium application. All indications for cranial CT imaging in paediatrics were assigned to one of these three regions. Descriptive statistical analysis of the distribution of the median values for CTDIvol and DLP yielded values in the minimum, maximum, 25th percentile (1st quartile), median (2nd quartile), and 75th percentile (3rd quartile). New DRLs for neuro-paediatric CT examinations in Switzerland were based on the 75th percentiles of the distributions of the median values of all eight centres. Where appropriate, values were rounded such that the DRLs increase or at least remain constant as the age of the patient increases. RESULTS: Our results revealed DRLs for CTDIvol and DLP up to 20% lower than the DRLs used so far in Switzerland and elsewhere in Europe. CONCLUSIONS: This study provides Swiss neuro-paediatric CT DRL values to establish optimum conditions for paediatric cranial CT examinations. Periodic national updates of DRLs, following international comparisons, are essential.


Subject(s)
Brain/diagnostic imaging , Tomography, X-Ray Computed/methods , Adolescent , Child , Child, Preschool , Humans , Infant , Infant, Newborn , Radiation Dosage , Reference Values , Retrospective Studies
12.
Phys Med ; 48: 111-118, 2018 Apr.
Article in English | MEDLINE | ID: mdl-29728223

ABSTRACT

PURPOSE: To investigate how various generations of iterative reconstruction (IR) algorithms impact low-contrast detectability (LCD) in abdominal computed tomography (CT) for different patient effective diameters, using a quantitative task-based approach. METHODS: Investigations were performed using an anthropomorphic abdominal phantom with two optional additional rings to simulate varying patient effective diameters (25, 30, and 35 cm), and containing multiple spherical targets (5, 6, and 8 mm in diameter) with a 20-HU contrast difference. The phantom was scanned using routine abdominal protocols (CTDIvol, 5.9-16 mGy) on four CT systems from two manufacturers. Images were reconstructed using both filtered back-projection (FBP) and various IR algorithms: ASiR 50%, SAFIRE 3 (both statistical IRs), ASiR-V 50%, ADMIRE 3 (both partial model-based IRs), or Veo (full model-based IR). Section thickness/interval was 2/1 mm or 2.5/1.25 mm, except 0.625/0.625 mm for Veo. We assessed LCD using a channelized Hotelling observer with 10 dense differences of Gaussian channels, with the area under the receiver operating characteristic curve (AUC) as a figure of merit. RESULTS: For the smallest phantom (25-cm diameter) and smallest lesion size (5-mm diameter), AUC for FBP and the various IR algorithms did not significantly differ for any of the tested CT systems. For the largest phantom (35-cm diameter), Veo yielded the highest AUC improvement (8.5%). Statistical and partial model-based IR algorithms did not significantly improve LCD. CONCLUSION: In abdominal CT, switching from FBP to IR algorithms offers limited possibilities for achieving significant dose reductions while ensuring a constant objective LCD.


Subject(s)
Abdomen/anatomy & histology , Abdomen/diagnostic imaging , Algorithms , Image Processing, Computer-Assisted/methods , Phantoms, Imaging , Signal-To-Noise Ratio , Tomography, X-Ray Computed/instrumentation , Humans
13.
Phys Med ; 48: 156-161, 2018 Apr.
Article in English | MEDLINE | ID: mdl-29631867

ABSTRACT

When performing CT examinations on pregnant patients, great effort should be dedicated towards optimising the exposure of the mother and the conceptus. For this purpose, many radiology departments use high-Z garments to be wrapped around the patient's lower abdomen for out-of-plane organ shielding to protect the fetus. To assess their current protection efficiency, we performed a literature review and compared the efficiencies mentioned in the literature to Monte-Carlo calculations of CT protocols for which the overall scan length was reduced. We found 11 relevant articles, all of them reporting uterus exposure due to CT imaging performed for exclusion of pulmonary embolism, one of the leading causes of peripartum deaths in western countries. Uterus doses ranged between 60 and 660 µGy per examination, and relative dose reductions to the uterus due to high-Z garments were between 20 and 56%. Calculations showed that reducing the scan length by one to three centimetres could potentially reduce uterus dose up to 24% for chest imaging, and even 47% for upper abdominal imaging. These dose reductions were in the order of those achieved by high-Z garments. However, using the latter may negatively influence the diagnostic image quality and even interfere with the automatic exposure control system thus increasing patient dose if positioned in the primary beam, for example in the overranging length in helical acquisition. We conclude that efforts should be concentrated on positioning the patient correctly in the gantry and optimising protocol parameters, rather than using high-Z garments for out-of-plane uterus shielding.


Subject(s)
Fetus/radiation effects , Monte Carlo Method , Radiation Dosage , Radiation Protection , Tomography, X-Ray Computed/adverse effects , Female , Humans , Phantoms, Imaging , Pregnancy , Radiation Exposure/prevention & control
14.
Med Phys ; 44(9): e153-e163, 2017 Sep.
Article in English | MEDLINE | ID: mdl-28901621

ABSTRACT

PURPOSE: We sought to compare objectively computed tomography (CT) scanner performance for three clinically relevant protocols using a task-based image quality assessment method in order to assess the potential for radiation dose reduction. METHODS: Four CT scanners released between 2003 and 2007 by different manufacturers were compared with four CT scanners released between 2012 and 2014 by the same manufacturers using ideal linear model observers (MO): prewhitening (PW) MO and channelized Hotelling (CHO) MO with Laguerre-Gauss channels for high-contrast spatial resolution and low-contrast detectability (LCD) performance, respectively. High-contrast spatial resolution was assessed using a custom-made phantom that enabled the computation of the target transfer function (TTF) and noise power spectrum (NPS). Low-contrast detectability was assessed using a commercially available anthropomorphic abdominal phantom providing equivalent diameters of 24, 29.6, and 34.6 cm. Three protocols were reviewed: a head (trauma) and an abdominal (urinary stones) protocol were applied to assess high-contrast spatial resolution performance; and another abdominal (focal liver lesions) protocol was applied for LCD. The liver protocol was tested using fixed and modulated tube currents. The PW MO was proposed for assessing high-contrast detectability performance of the various CT scanners. RESULTS: Compared with older generation CT scanners, three newer systems displayed significant improvements in high-contrast detectability over that of their predecessors. A fourth, newer system had lower performance. The CHO MO was appropriate for assessing LCD performance and revealed that an excellent level of image quality could be obtained with newer scanners at significantly lower dose levels. CONCLUSIONS: This study shows that MO can objectively benchmark CT scanners using a task-based image quality method, thus helping to estimate the potential for further dose reductions offered by the latest systems. Such an approach may be useful for adequately and quantitatively comparing clinically relevant image quality among various scanners.


Subject(s)
Radiation Dosage , Tomography Scanners, X-Ray Computed , Tomography, X-Ray Computed , Clinical Protocols , Humans , Phantoms, Imaging
15.
Acad Radiol ; 24(9): 1114-1124, 2017 09.
Article in English | MEDLINE | ID: mdl-28365232

ABSTRACT

RATIONALE AND OBJECTIVES: To compare adaptive statistical iterative reconstruction (ASIR) and model-based iterative reconstruction (MBIR) algorithms for reduced-dose computed tomography (CT). MATERIALS AND METHODS: Forty-four young oncology patients (mean age 30 ± 9 years) were included. After routine thoraco-abdominal CT (dose 100%, average CTDIvol 9.1 ± 2.4 mGy, range 4.4-16.9 mGy), follow-up CT was acquired at 50% (average CTDIvol 4.5 ± 1.2 mGy, range 2.2-8.4 mGy) in 29 patients additionally at 20% dose (average CTDIvol 1.9 ± 0.5 mGy, range 0.9-3.4 mGy). Each reduced-dose CT was reconstructed using both ASIR and MBIR. Four radiologists (two juniors and two seniors) blinded to dose and technique read each set of CT images regarding objective and subjective image qualities (high- or low-contrast structures), subjective noise or pixilated appearance, diagnostic confidence, and lesion detection. RESULTS: At all dose levels, objective image noise was significantly lower with MBIR than with ASIR (P < 0.001). The subjective image quality for low-contrast structures was significantly higher with MBIR than with ASIR (P < 0.001). Reduced-dose abdominal CT images of patients with higher body mass index (BMI) were read with significantly higher diagnostic confidence than images of slimmer patients (P < 0.001) and had higher subjective image quality, regardless of technique. Although MBIR images appeared significantly more pixilated than ASIR images, they were read with higher diagnostic confidence, especially by juniors (P < 0.001). CONCLUSIONS: Reduced-dose CT during the follow-up of young oncology patients should be reconstructed with MBIR to ensure diagnostic quality. Elevated body mass index does not hamper the quality of reduced-dose CT.


Subject(s)
Algorithms , Image Processing, Computer-Assisted/methods , Neoplasms/diagnostic imaging , Radiation Dosage , Radiographic Image Interpretation, Computer-Assisted/methods , Tomography, X-Ray Computed/methods , Adolescent , Adult , Body Mass Index , Female , Humans , Male , Radiation Exposure , Signal-To-Noise Ratio , Single-Blind Method , Young Adult
16.
Z Med Phys ; 27(2): 86-97, 2017 Jun.
Article in English | MEDLINE | ID: mdl-27156923

ABSTRACT

PURPOSE: This study aims to assess CT image quality in a way that would meet specific requirements of clinical practice. Physics metrics like Fourier transform derived metrics were traditionally employed for that. However, assessment methods through a detection task have also developed quite extensively lately, and we chose here to rely on this modality for image quality assessment. Our goal was to develop a tool adapted for a fast and reliable CT image quality assessment in order to pave the way for new CT benchmarking techniques in a clinical context. Additionally, we also used this method to estimate the benefits brought by some IR algorithms. MATERIALS AND METHODS: A modified QRM chest phantom containing spheres of 5 and 8mm at contrast levels of 10 and 20HU at 120kVp was used. Images of the phantom were acquired at CTDIvol of 0.8, 3.6, 8.2 and 14.5mGy, before being reconstructed using FBP, ASIR 40 and MBIR on a GE HD 750 CT scanner. They were then assessed by eight human observers undergoing a 4-AFC test. After that, these data were compared with the results obtained from two different model observers (NPWE and CHO with DDoG channels). The study investigated the effects of the acquisition conditions as well as reconstruction methods. RESULTS: NPWE and CHO models both gave coherent results and approximated human observer results well. Moreover, the reconstruction technique used to retrieve the images had a clear impact on the PC values. Both models suggest that switching from FBP to ASIR 40 and particularly to MBIR produces an increase of the low contrast detection, provided a minimum level of exposure is reached. CONCLUSION: Our work shows that both CHO with DDoG channels and NPWE models both approximate the trend of humans performing a detection task. Both models also suggest that the use of MBIR goes along with an increase of the PCs, indicating that further dose reduction is still possible when using those techniques. Eventually, the CHO model associated to the protocol we described in this study happened to be an efficient way to assess CT images in a clinical environment. In the future, this simple method could represent a sound basis to benchmark clinical practice and CT devices.


Subject(s)
Benchmarking , Phantoms, Imaging , Tomography, X-Ray Computed/standards , Algorithms , Humans , Observer Variation , Radiation Dosage , Radiographic Image Interpretation, Computer-Assisted , Tomography, X-Ray Computed/methods
17.
Am J Cardiol ; 118(2): 188-94, 2016 07 15.
Article in English | MEDLINE | ID: mdl-27239022

ABSTRACT

We sought to compare operator radiation exposure during procedures using right femoral access (RFA), right radial access (RRA), and left radial access (LRA) during coronary angiography (CA) and percutaneous coronary intervention (PCI). Because of an increased incidence of long-term malignancy in interventional cardiologists, operator radiation exposure is of rising concern. This prospective study included all consecutive patients who underwent elective or emergency CA ± PCI from September 2014 to March 2015. The primary end point was operator radiation exposure, quantified as the ratio of operator cumulative dose (CD) and patient radiation reported as dose-area product (DAP) (CD/DAP). Secondary end points included CD, DAP, and fluoroscopy time (FT). Overall 830 procedures (457 CA [55%] and 373 PCI [45%]) were performed, 455 (55%) through RFA, 272 (33%) through RRA, and 103 (12%) through LRA. The CD/DAP was lower in RFA (0.09 µSv/Gycm(2) [0.02 to 0.20]) compared with RRA (0.47 µSv/Gycm(2) [0.25 to 0.75], p <0.001). The LRA showed lower CD/DAP compared with RRA (p <0.001). CD was significantly lower in RFA (3 µSv [1 to 7]) compared with RRA (12 µSv [6 to 29], p <0.001). The LRA showed lower CD compared with RRA (p <0.001). There were no significant differences in DAP among the 3 access sites. FT was similar for the 3 groups (RFA 7 ± 7, RRA 5 ± 5, LRA 6 ± 5 minutes, RFA vs RRA: p = 1, RFA vs LRA: p = 0.16, RRA vs LRA: p = 0.52). In conclusion, the use of RFA during CA ± PCI is associated with significantly lower operator radiation exposure compared with RRA. LRA is associated with significantly lower operator radiation exposure compared with RRA.


Subject(s)
Cardiac Catheterization , Cardiologists , Coronary Angiography , Coronary Vessels/diagnostic imaging , Occupational Exposure/statistics & numerical data , Percutaneous Coronary Intervention , Radiation Dosage , Radiation Exposure/statistics & numerical data , Aged , Aged, 80 and over , Coronary Vessels/surgery , Female , Femoral Artery , Fluoroscopy , Humans , Male , Middle Aged , Operative Time , Prospective Studies , Radial Artery , Radiometry , Time Factors
18.
Radiat Prot Dosimetry ; 169(1-4): 68-72, 2016 06.
Article in English | MEDLINE | ID: mdl-26962148

ABSTRACT

The goal of the present work was to report and investigate the performances of a new iterative reconstruction algorithm, using a model observer. For that, a dedicated low-contrast phantom containing different targets was scanned at four volume computed tomography dose index (CTDIvol) levels on a Siemens SOMATOM Force computed tomography (CT). The acquired images were reconstructed using the ADMIRE algorithm and were then assessed by three human observers who performed alternative forced choice experiments. Next, a channelised hotelling observer model was applied on the same set of images. The comparison between the two was performed using the percentage correct as a figure of merit. The results indicated a strong agreement between human and model observer as well as an improvement in the low-contrast detection when switching from an ADMIRE strength of 1-3. Good results were also observed even in situations where the target was hard to detect, suggesting that patient dose could be further reduced and optimised.

19.
Radiat Prot Dosimetry ; 169(1-4): 78-83, 2016 Jun.
Article in English | MEDLINE | ID: mdl-26940439

ABSTRACT

Patient dose optimisation in computed tomography (CT) should be done using clinically relevant tasks when dealing with image quality assessments. In the present work, low-contrast detectability for an average patient morphology was assessed on 56 CT units, using a model observer applied on images acquired with two specific protocols of an anthropomorphic phantom containing spheres. Images were assessed using the channelised Hotelling observer (CHO) with dense difference of Gaussian channels. The results were computed by performing receiver operating characteristics analysis (ROC) and using the area under the ROC curve (AUC) as a figure of merit. The results showed a small disparity at a volume computed tomography dose index (CTDIvol) of 15 mGy depending on the CT units for the chosen image quality criterion. For 8-mm targets, AUCs were 0.999 ± 0.018 at 20 Hounsfield units (HU) and 0.927 ± 0.054 at 10 HU. For 5-mm targets, AUCs were 0.947 ± 0.059 and 0.702 ± 0.068 at 20 and 10 HU, respectively. The robustness of the CHO opens the way for CT protocol benchmarking and optimisation processes.


Subject(s)
Benchmarking/standards , Radiation Exposure/analysis , Radiation Monitoring/standards , Radiation Protection/standards , Radiographic Image Enhancement/standards , Tomography, X-Ray Computed/standards , Practice Guidelines as Topic , Radiation Exposure/prevention & control , Reproducibility of Results , Sensitivity and Specificity , Switzerland
20.
Radiat Prot Dosimetry ; 169(1-4): 73-7, 2016 Jun.
Article in English | MEDLINE | ID: mdl-26922787

ABSTRACT

Evaluating image quality by using receiver operating characteristic studies is time consuming and difficult to implement. This work assesses a new iterative algorithm using a channelised Hotelling observer (CHO). For this purpose, an anthropomorphic abdomen phantom with spheres of various sizes and contrasts was scanned at 3 volume computed tomography dose index (CTDIvol) levels on a GE Revolution CT. Images were reconstructed using the iterative reconstruction method adaptive statistical iterative reconstruction-V (ASIR-V) at ASIR-V 0, 50 and 70 % and assessed by applying a CHO with dense difference of Gaussian and internal noise. Both CHO and human observers (HO) were compared based on a four-alternative forced-choice experiment, using the percentage correct as a figure of merit. The results showed accordance between CHO and HO. Moreover, an improvement in the low-contrast detection was observed when switching from ASIR-V 0 to 50 %. The results underpin the finding that ASIR-V allows dose reduction.


Subject(s)
Algorithms , Imaging, Three-Dimensional/methods , Radiation Exposure/prevention & control , Radiographic Image Enhancement/methods , Radiography, Abdominal/methods , Tomography, X-Ray Computed/methods , Humans , Image Enhancement/methods , Observer Variation , Radiation Exposure/analysis , Radiation Protection/methods , Reproducibility of Results , Sensitivity and Specificity