J Virol Methods ; 162(1-2): 288-90, 2009 Dec.
Article in English | MEDLINE | ID: mdl-19699237

ABSTRACT

Tetramethylrhodamine (TAMRA) and black hole quencher 1 (BHQ1) quenched probes and five one-step RT-PCR kits were evaluated in TaqMan real-time RT-PCR assays for detection of respiratory pathogens. The intra-assay variability of the BHQ1 probes were 1.2-2.8-fold lower than those of the TAMRA probes. All kits amplified the specific targets, but differed in their sensitivity by up to 3 orders of magnitude. The AgPath-ID kit provided the best overall performance for all assay targets.


Subject(s)
DNA Probes , Fluorescent Dyes , Reagent Kits, Diagnostic , Respiratory System , Reverse Transcriptase Polymerase Chain Reaction/methods , Rhodamines , Bacteria/classification , Bacteria/genetics , Bacteria/isolation & purification , Bacterial Infections/virology , Fluorescent Dyes/classification , Fluorescent Dyes/metabolism , Humans , Respiratory System/microbiology , Respiratory System/virology , Rhodamines/metabolism , Taq Polymerase , Virus Diseases/virology , Viruses/classification , Viruses/genetics , Viruses/isolation & purification