1.
Phys Med Biol ; 52(15): 4541-52, 2007 Aug 07.
Article in English | MEDLINE | ID: mdl-17634649

ABSTRACT

Most IMRT techniques have been designed to treat targets smaller than the field size of conventional linac accelerators. In order to overcome the field size restrictions in applying IMRT, we developed a two isocenter IMRT technique to treat long volume targets. The technique exploits an extended dose gradient throughout a junction region of 4-6 cm to minimize the impact of field match errors on a junction dose and manipulates the inverse planning and IMRT segments to fill in the dose gradient and achieve dose uniformity. Techniques for abutting both conventional fields with IMRT ('Static + IMRT') and IMRT fields ('IMRT + IMRT') using two separate isocenters have been developed. Five long volume sarcoma cases have been planned in Pinnacle (Philips, Madison, USA) using Elekta Synergy and Varian 2100EX linacs; two of the cases were clinically treated with this technique. Advantages were demonstrated with well-controlled junction target uniformity and tolerance to setup uncertainties. The junction target dose heterogeneity was controlled at a level of +/-5%; for 3 mm setup errors at the field edges, the junction target dose changed less than 5% and the dose sparing to organs at risk (OARs) was maintained. Film measurements confirmed the treatment planning results.


Subject(s)
Models, Biological , Neoplasms/radiotherapy , Radiometry/methods , Radiotherapy Planning, Computer-Assisted/methods , Radiotherapy, Conformal/methods , Computer Simulation , Humans , Organ Size , Radiotherapy Dosage
2.
Phys Med Biol ; 50(6): 1119-29, 2005 Mar 21.
Article in English | MEDLINE | ID: mdl-15798312

ABSTRACT

The energy dependence of alanine/EPR dosimetry for 8, 12, 18 and 22 MeV clinical electron beams was investigated by experiment and by Monte Carlo simulations. Alanine pellets in a waterproof holder were irradiated in a water phantom using an Elekta Precise linear accelerator. The dose rates at the reference point were determined following the TG-51 protocol using an NACP-02 parallel-plate chamber calibrated in a (60)Co beam. The EPR spectra of irradiated pellets were measured using a Bruker EMX 081 EPR spectrometer. Experimentally, we found no significant change in alanine/EPR response to absorbed dose-to-water over the energy range 8-22 MeV at an uncertainty level of 0.6%. However, the response for high-energy electrons is about 1.3 (+/-1.1)% lower than for (60)Co. The EGSnrc Monte Carlo system was used to calculate the ratio of absorbed dose-to-alanine to absorbed dose-to-water and it was shown that there is 1.3 (+/-0.2)% reduction in this ratio from the (60)Co beam to the electron beams, which confirms the experimental results. Alanine/EPR response per unit absorbed dose-to-alanine was also investigated and it is the same for high-energy electrons and (60)Co gamma-rays.


Subject(s)
Alanine/radiation effects , Brachytherapy/instrumentation , Cobalt Radioisotopes/analysis , Cobalt Radioisotopes/therapeutic use , Electron Spin Resonance Spectroscopy/instrumentation , Radiometry/instrumentation , Radiotherapy Planning, Computer-Assisted/instrumentation , Brachytherapy/methods , Computer Simulation , Electron Spin Resonance Spectroscopy/methods , Equipment Design , Equipment Failure Analysis , Linear Energy Transfer , Models, Biological , Models, Statistical , Monte Carlo Method , Radiometry/methods , Radiopharmaceuticals/analysis , Radiopharmaceuticals/therapeutic use , Radiotherapy Dosage , Radiotherapy Planning, Computer-Assisted/methods
3.
Phys Med Biol ; 49(2): 257-70, 2004 Jan 21.
Article in English | MEDLINE | ID: mdl-15083670

ABSTRACT

The energy dependence of alanine/EPR dosimetry, in terms of absorbed dose-to-water for clinical 6, 10, 25 MV x-rays and 60Co gamma-rays was investigated by measurements and Monte Carlo (MC) calculations. The dose rates were traceable to the NRC primary standard for absorbed dose, a sealed water calorimetry. The electron paramagnetic resonance (EPR) spectra of irradiated pellets were measured using a Bruker EMX 081 EPR spectrometer. The DOSRZnrc Monte Carlo code of the EGSnrc system was used to simulate the experimental conditions with BEAM code calculated input spectra of x-rays and gamma-rays. Within the experimental uncertainty of 0.5%, the alanine EPR response to absorbed dose-to-water for x-rays was not dependent on beam quality from 6 MV to 25 MV, but on average, it was about 0.6% lower than its response to 60Co gamma-rays. Combining experimental data with Monte Carlo calculations, it is found that the alanine/EPR response per unit absorbed dose-to-alanine is the same for clinical x-rays and 60Co gamma-rays within the uncertainty of 0.6%. Monte Carlo simulations showed that neither the presence of PMMA holder nor varying the dosimeter thickness between 1 mm and 5 mm has significant effect of the energy dependence of alanine/EPR dosimetry within the calculation uncertainty of 0.3%.


Subject(s)
Alanine/chemistry , Electron Spin Resonance Spectroscopy/methods , Radiometry/methods , Calibration , Calorimetry , Cobalt Radioisotopes/therapeutic use , Dose-Response Relationship, Radiation , Gamma Rays , Monte Carlo Method , Photons , Radiation Dosage , X-Rays