1.
Appl Radiat Isot ; 163: 109200, 2020 Sep.
Article in English | MEDLINE | ID: mdl-32561041

ABSTRACT

This study validated a model of the Siemens Symbia T16 dual-head SPECT/CT gamma camera created using the Monte Carlo program SIMIND for 177Lu. The validation was done by comparing experimental and simulated gamma camera performance criteria tests for the 177Lu 208 keV photopeak with a medium-energy collimator. Results showed good agreement between the experimental and simulated values. These results illustrated that SIMIND could emulate the Symbia T16 successfully and therefore, can be used with confidence to model 177Lu images.

2.
Phys Med ; 27(1): 11-20, 2011 Jan.
Article in English | MEDLINE | ID: mdl-20206563

ABSTRACT

Transmission sources used for image attenuation correction, allowing image quantification, are collimated to reduce scatter. We propose the same effect can be achieved for an uncollimated source by increasing source to patient distance. The aim was to compare planar image performance characteristics and absorbed doses of uncollimated and collimated radioactive printed paper transmission sources. The scatter contribution to the uncollimated (99m)Tc source data was evaluated for different combinations of detector phantom distance, detector source distance and phantom source distance. Measurements were performed by increasing the Lucite phantom thickness in 1cm steps to 20 cm. Spatial resolution, detection efficiency and entrance absorbed dose rate were measured for the uncollimated and collimated transmission source images. Results derived from the energy spectra, obtained with the uncollimated transmission source indicate that scatter contribution increases with decreasing detector source distance. The scatter component in the uncollimated transmission images (detector source distances ≥ 60 cm; phantom source distances ≥ 40 cm) was comparable to that obtained with collimated transmission images. Attenuation coefficients obtained compared well (0.168 cm⁻¹ vs. 0.171 cm⁻¹). The full widths at half maxima differed by less than 0.9 mm. The detection efficiency of the uncollimated source was 2.5 times higher than obtained with the collimated source. The entrance absorbed dose obtained from an uncollimated source was 3.75 times larger than that obtained from the collimated source. An uncollimated transmission source (detector source distance ≥ 60 cm) results in acceptable image characteristics and presents a low cost, low dose, high efficiency option for transmission imaging.


Subject(s)
Image Processing, Computer-Assisted/methods , Phantoms, Imaging , Tomography, Emission-Computed/methods , Humans , Image Processing, Computer-Assisted/instrumentation , Radiation Dosage , Reproducibility of Results , Scattering, Radiation , Sensitivity and Specificity , Technetium , Tomography, Emission-Computed/instrumentation
3.
Phys Med Biol ; 52(15): N329-37, 2007 Aug 07.
Article in English | MEDLINE | ID: mdl-17634634

ABSTRACT

This note proposes the use of a standard inkjet printer to produce radioactive (99m)Tc phantoms that can be used for routine quality control of gamma cameras. The amount of activity that will be deposited on paper per unit area was predicted and compared to the measured activity. The uniformity of the printouts was compared to the uniformity obtained with a standard (57)Co flood source. A scintillation detector connected to a multi-channel analyzer was used to evaluate the uniformity of the printout independently from the gamma camera. Joining two A4 size printed phantoms to create larger sources was evaluated. The spatial resolution obtained with printed sources was compared to that obtained using standard line source techniques. The results indicated that the uniformity of the printed phantoms compared well with those obtained with the (57)Co flood source (integral uniformity 2.29% (printed source) and 2.10% ((57)Co flood source)). There was no difference in the resolution measurements obtained with the printed sources and those obtained with the standard methods. This study demonstrates that affordable phantoms can easily be created to evaluate system uniformity and resolution in any department where a standard PC and inkjet printer are available.


Subject(s)
Computer Peripherals , Phantoms, Imaging , Positron-Emission Tomography/instrumentation , Printing/instrumentation , Quality Assurance, Health Care/methods , Technetium/chemistry , Equipment Design , Equipment Failure Analysis , Positron-Emission Tomography/methods , Radiopharmaceuticals/chemistry , Reproducibility of Results , Sensitivity and Specificity
4.
Cardiovasc J S Afr ; 12(4): 196-200, 2001.
Article in English | MEDLINE | ID: mdl-11717696

ABSTRACT

BACKGROUND: The accuracy of the left ventricular ejection fraction (LVEF) calculated from gated single photon emission computed tomography (GSPECT) studies is dependent on the accuracy of the determination of the end-diastolic volume (EDV) and end-systolic volume (ESV) of the left ventricle (LV). In this study we evaluated the feasibility of calculating the EDV, ESV and LVEF from the area under the polar graph (APC) of the edges of the LV image determined by a first derivative edge-detection method. METHODS AND RESULTS: Technetium-99m ((99m)Tc) sestamibi GSPECT studies and planar equilibrium radionuclide ventriculography (ERNV) were performed on 15 male subjects in whom the LVEF ranged from 19% to 75%. Images were reconstructed to obtain short axis slices of the LV spanning the cardiac cycle. On each slice the LV edge points were determined at 10 degrees intervals using the APC method. The area of each short axis slice was determined by conversion to polar co-ordinates, interpolation and numerical integration of the graphs and multiplication by a pre-determined conversion factor. RESULTS: Edges were successfully determined in all 15 patients using the APC method. The LVEF results correlate well with conventional planar ERNV studies (r = 0.96, LVEF(GSPECT) = 8.80 + 0.66 LVEF( ERNV)). The absolute difference between the LVEF for ERNV and for the APC method was 6.1% with a standard deviation of 7.6%. The reproducibility of SPECT LVEF using the APC method was good (intra-observer r = 0.99, inter-observer r = 0.99). CONCLUSIONS: The APC method provides for easy and accurate ejection fraction determination with limited underlying mathematical assumptions. The ability to interpolate the edge points provides for stable edge detection even in hypoperfused myocardium.


Subject(s)
Stroke Volume , Technetium Tc 99m Sestamibi , Tomography, Emission-Computed, Single-Photon/methods , Ventricular Dysfunction, Left/diagnostic imaging , Ventricular Function, Left , Adult , Humans , Male , Middle Aged , Observer Variation , Radiopharmaceuticals , Reproducibility of Results , Tomography, Emission-Computed, Single-Photon/standards , Tomography, Emission-Computed, Single-Photon/statistics & numerical data
5.
J Nucl Med ; 40(9): 1532-5, 1999 Sep.
Article in English | MEDLINE | ID: mdl-10492376

ABSTRACT

UNLABELLED: The use of 99mTc-labeled red blood cells (RBC) for the evaluation of left ventricular function using equilibrium-gated blood-pool imaging suffers from several problems and potential risks. In this study, we estimated the absorbed radiation dose of 99mTc-labeled dimercaptopropionyl human serum albumin (DMP-HSA) as a potential alternative to 99mTc-RBC. METHODS: After the administration of 99mTc-DMP-HSA, whole-body imaging was performed up to 48 h after injection in five volunteers. The heart contents, liver and remainder of the body were used as source organs. Multicompartment modeling of the biodistribution was performed and absorbed radiation dose estimates for 99mTc-DMP-HSA were obtained using the Medical Internal Radiation Dose (MIRD) calculation. RESULTS: Residence times of 0.62 and 0.43 h were obtained for the heart contents and liver, respectively. Radiation dose estimates yielded an effective dose of 0.0055 mSv/MBq. CONCLUSION: 99MTC-DMP-HSA yielded absorbed radiation doses comparable with those of 99mTc-RBC. Therefore, the radiation properties of 99mTc-DMP-HSA are such that it can be used for clinical diagnostic studies.


Subject(s)
Organotechnetium Compounds/pharmacokinetics , Radiopharmaceuticals/pharmacokinetics , Absorption , Adult , Biological Transport , Computer Simulation , Humans , Male , Radiation Dosage , Tissue Distribution