1.
Micromachines (Basel)
; 12(8)2021 Aug 16.
Article
in English
| MEDLINE
| ID: mdl-34442593
ABSTRACT
The majority of soft pneumatic actuators for rehabilitation exercises have been designed for adult users. Specifically, there is a paucity of soft rehabilitative devices designed for infants with upper and lower limb motor disabilities. We present a low-profile vacuum-powered artificial muscle (LP-VPAM) with dimensions suitable for infants. The actuator produced a maximum force of 26 N at vacuum pressures of -40 kPa. When implemented in an experimental model of an infant leg in an antagonistic-agonist configuration to measure resultant knee flexion, the actuator generated knee flexion angles of 43° and 61° in the prone and side-lying position, respectively.