ABSTRACT
Since the inception of its proficiency test program to evaluate radionuclide measurement in hospitals and clinics, the National Metrology Laboratory of Ionizing Radiation-LNMRI, that represents Brazilian National Metrology Institute (NMI) for ionizing radiation has expanded its measurement and calibration capability. Requirements from the National Health Surveillance Agency from Ministry of Health (ANVISA), to producers of radiopharmaceuticals provided an opportunity to improve the full traceability chain to the highest level. Fluorodeoxyglucose (FDG-(18)F) is the only radiopharmaceutical simultaneously produced by all Brazilian radiopharmaceutical production centers (RPCs). By running this proficiency test, LNMRI began to provide them with the required traceability. For evaluation, the ratio of RPC to reference value results and ISO/IEC17043:2010 criteria were used. The reference value established as calibration factor on the secondary standard ionization chamber was obtained from three absolute measurements systems, and routinely confirmed in each round of proficiency test by CIEMAT/NIST liquid scintillation counting. The γ-emitting impurities were checked using a High-Purity Germanium (HPGe) detector. The results show that Brazilian RPCs are in accordance with (accuracy within ±10%) the Brazilian standard for evaluation of measurements with radionuclide calibrators (CNEN NN 3.05., 2013). Nevertheless, the RPCs should improve the methodology of uncertainty estimates, essential when using the statistical criteria of ISO/IEC 17043 standard, in addition to improving accuracy to levels consistent with their position in the national traceability chain.
Subject(s)
Fluorine Radioisotopes/analysis , Fluorine Radioisotopes/standards , Public Sector/standards , Radiometry/methods , Radiometry/standards , Brazil , Reference Standards , Reproducibility of Results , Sensitivity and SpecificityABSTRACT
Since 2003, radiopharmaceuticals for medical diagnostic purposes have been produced at the Instituto de Engenharia Nuclear, in Brazil, using two cyclotron accelerators - CV-28 and RDS111. As a result of the ever increasing production, a diagnostic assessment to reduce neutron dose rates received by occupationally exposed individuals during irradiation processes has been developed. The purpose of this work is to present this assessment, which is currently being applied to both the Fluorine and Iodine targets of CV-28 and RDS111 cyclotron accelerators.