J Chromatogr A ; 828(1-2): 219-27, 1998 Dec 18.
Article in English | MEDLINE | ID: mdl-9916308

ABSTRACT

An ultra-sensitive method for the determination of fluspirilene in plasma was established, using high-performance liquid chromatographic separation with tandem mass spectrometric detection. The samples were extracted with hexane/isoamyl alcohol, separated on a Phenomenex Luna C18 5 mu 150 x 2.1 mm column with a mobile phase consisting of methanol-water-acetic acid (600:400:1) at a flow-rate of 0.3 ml/min. Detection was achieved by a Finnigan Matt mass spectrometer (LCQ) at unit resolution in full scan mode scanning the product ion spectrum from m/z 130-500 and monitoring the transition of the protonated molecular ion at m/z 476.2, to the sum of the largest product ions m/z 371, 342 and 274 (MS-MS). Electrospray ionisation was used for ion production. The mean recovery for fluspirilene was 90% with a lower limit of quantification of 21.50 pg/ml using 1 ml plasma for extraction. This is the first chromatographic method described for the determination of fluspirilene in plasma that is accurate and sensitive enough to be used in pharmacokinetic studies.


Subject(s)
Antipsychotic Agents/blood , Chromatography, High Pressure Liquid/methods , Fluspirilene/blood , Mass Spectrometry/methods , Humans , Reference Standards , Reference Values , Reproducibility of Results , Sensitivity and Specificity