Results 1 - 20 de 2.485
Filter
1.
Radiat Environ Biophys ; 63(2): 297-306, 2024 May.
Article in English | MEDLINE | ID: mdl-38722389

ABSTRACT

For locally advanced cervical cancer, the standard therapeutic approach involves concomitant chemoradiation therapy, supplemented by a brachytherapy boost. Moreover, an external beam radiotherapy (RT) boost should be considered for treating gross lymph node (LN) volumes. Two boost approaches exist with Volumetric Intensity Modulated Arc Therapy (VMAT): Sequential (SEQ) and Simultaneous Integrated Boost (SIB). This study undertakes a comprehensive dosimetric and radiobiological comparison between these two boost strategies. The study encompassed ten patients who underwent RT for cervical cancer with node-positive disease. Two sets of treatment plans were generated for each patient: SIB-VMAT and SEQ-VMAT. Dosimetric as well as radiobiological parameters including tumour control probability (TCP) and normal tissue complication probability (NTCP) were compared. Both techniques were analyzed for two different levels of LN involvement - only pelvic LNs and pelvic with para-aortic LNs. Statistical analysis was performed using SPSS software version 25.0. SIB-VMAT exhibited superior target coverage, yielding improved doses to the planning target volume (PTV) and gross tumour volume (GTV). Notably, SIB-VMAT plans displayed markedly superior dose conformity. While SEQ-VMAT displayed favorable organ sparing for femoral heads, SIB-VMAT appeared as the more efficient approach for mitigating bladder and bowel doses. TCP was significantly higher with SIB-VMAT, suggesting a higher likelihood of successful tumour control. Conversely, no statistically significant difference in NTCP was observed between the two techniques. This study's findings underscore the advantages of SIB-VMAT over SEQ-VMAT in terms of improved target coverage, dose conformity, and tumour control probability. In particular, SIB-VMAT demonstrated potential benefits for cases involving para-aortic nodes. It is concluded that SIB-VMAT should be the preferred approach in all cases of locally advanced cervical cancer.


Subject(s)
Radiotherapy Dosage , Radiotherapy, Intensity-Modulated , Uterine Cervical Neoplasms , Humans , Uterine Cervical Neoplasms/radiotherapy , Uterine Cervical Neoplasms/pathology , Female , Radiotherapy, Intensity-Modulated/methods , Radiotherapy Planning, Computer-Assisted/methods , Radiometry , Middle Aged , Organs at Risk/radiation effects , Lymphatic Metastasis/radiotherapy
2.
Biomed Phys Eng Express ; 10(4)2024 May 10.
Article in English | MEDLINE | ID: mdl-38697044

ABSTRACT

Objective.The aim of this work was to develop a Phase I control chart framework for the recently proposed multivariate risk-adjusted Hotelling'sT2chart. Although this control chart alone can identify most patients receiving extreme organ-at-risk (OAR) dose, it is restricted by underlying distributional assumptions, making it sensitive to extreme observations in the sample, as is typically found in radiotherapy plan quality data such as dose-volume histogram (DVH) points. This can lead to slightly poor-quality plans that should have been identified as out-of-control (OC) to be signaled in-control (IC).Approach. We develop a robust iterative control chart framework to identify all OC patients with abnormally high OAR dose and improve them via re-optimization to achieve an IC sample prior to establishing the Phase I control chart, which can be used to monitor future treatment plans.Main Results. Eighty head-and-neck patients were used in this study. After the first iteration, P14, P67, and P68 were detected as OC for high brainstem dose, warranting re-optimization aimed to reduce brainstem dose without worsening other planning criteria. The DVH and control chart were updated after re-optimization. On the second iteration, P14, P67, and P68 were IC, but P40 was identified as OC. After re-optimizing P40's plan and updating the DVH and control chart, P40 was IC, but P14* (P14's re-optimized plan) and P62 were flagged as OC. P14* could not be re-optimized without worsening target coverage, so only P62 was re-optimized. Ultimately, a fully IC sample was achieved. Multiple iterations were needed to identify and improve all OC patients, and to establish a more robust control limit to monitor future treatment plans.Significance. The iterative procedure resulted in a fully IC sample of patients. With this sample, a more robust Phase I control chart that can monitor OAR doses of new plans was established.


Subject(s)
Organs at Risk , Quality Control , Radiotherapy Dosage , Radiotherapy Planning, Computer-Assisted , Humans , Organs at Risk/radiation effects , Radiotherapy Planning, Computer-Assisted/methods , Head and Neck Neoplasms/radiotherapy , Algorithms
3.
Radiat Oncol ; 19(1): 54, 2024 May 03.
Article in English | MEDLINE | ID: mdl-38702761

ABSTRACT

BACKGROUND: Stereotactic ablative body radiotherapy (SABR) is an emerging treatment alternative for patients with localized low and intermediate risk prostate cancer patients. As already explored by some authors in the context of conventional moderate hypofractionated radiotherapy, focal boost of the index lesion defined by magnetic resonance imaging (MRI) is associated with an improved biochemical outcome. The objective of this phase II trial is to determine the effectiveness (in terms of biochemical, morphological and functional control), the safety and impact on quality of life, of prostate SABR with MRI guided focal dose intensification in males with intermediate and high-risk localized prostate cancer. METHODS: Patients with intermediate and high-risk prostate cancer according to NCCN definition will be treated with SABR 36.25 Gy in 5 fractions to the whole prostate gland with MRI guided simultaneous integrated focal boost (SIB) to the index lesion (IL) up to 50 Gy in 5 fractions, using a protocol of bladder trigone and urethra sparing. Intra-fractional motion will be monitored with daily cone beam computed tomography (CBCT) and intra-fractional tracking with intraprostatic gold fiducials. Androgen deprivation therapy (ADT) will be allowed. The primary endpoint will be efficacy in terms of biochemical and local control assessed by Phoenix criteria and post-treatment MRI respectively. The secondary endpoints will encompass acute and late toxicity, quality of life (QoL) and progression-free survival. Finally, the subgroup of high-risk patients will be involved in a prospective study focused on immuno-phenotyping. DISCUSSION: To the best of our knowledge, this is the first trial to evaluate the impact of post-treatment MRI on local control among patients with intermediate and high-risk prostate cancer undergoing SABR and MRI guided focal intensification. The results of this trial will enhance our understanding of treatment focal intensification through the employment of the SABR technique within this specific patient subgroup, particularly among those with high-risk disease, and will help to clarify the significance of MRI in monitoring local responses. Hopefully will also help to design more personalized biomarker-based phase III trials in this specific context. Additionally, this trial is expected to be incorporated into a prospective radiomics study focused on localized prostate cancer treated with radiotherapy. TRIAL REGISTRATION: Clinicaltrials.gov identifier: NCT05919524; Registered 17 July 2023. TRIAL SPONSOR: IRAD/SEOR (Instituto de Investigación de Oncología Radioterápica / Sociedad Española de Oncología Radioterápica). STUDY SETTING: Clinicaltrials.gov identifier: NCT05919524; Registered 17 July 2023. TRIAL STATUS: Protocol version number and date: v. 5/ 17 May-2023. Date of recruitment start: August 8, 2023. Date of recruitment completion: July 1, 2024.


Subject(s)
Prostatic Neoplasms , Radiosurgery , Radiotherapy, Image-Guided , Humans , Male , Prostatic Neoplasms/radiotherapy , Prostatic Neoplasms/diagnostic imaging , Prostatic Neoplasms/surgery , Prostatic Neoplasms/pathology , Radiosurgery/methods , Radiotherapy, Image-Guided/methods , Urinary Bladder/radiation effects , Organs at Risk/radiation effects , Organ Sparing Treatments/methods , Magnetic Resonance Imaging/methods , Prospective Studies , Quality of Life , Radiotherapy Dosage , Aged , Radiotherapy Planning, Computer-Assisted/methods , Middle Aged
4.
J Appl Clin Med Phys ; 25(5): e14361, 2024 May.
Article in English | MEDLINE | ID: mdl-38642406

ABSTRACT

PURPOSES: This study aimed to develop and validate algorithms for automating intensity modulated radiation therapy (IMRT) planning in breast cancer patients, with a focus on patient anatomical characteristics. MATERIAL AND METHODS: We retrospectively selected 400 breast cancer patients without lymph node involvement for automated treatment planning. Automation was achieved using the Eclipse Scripting Application Programming Interface (ESAPI) integrated into the Eclipse Treatment Planning System. We employed three beam insertion geometries and three optimization strategies, resulting in 3600 plans, each delivering a 40.05 Gy dose in 15 fractions. Gantry angles in the tangent fields were selected based on a criterion involving the minimum intersection area between the Planning Target Volume (PTV) and the ipsilateral lung in the Beam's Eye View projection. ESAPI was also used to gather patient anatomical data, serving as input for Random Forest models to select the optimal plan. The Random Forest classification considered both beam insertion geometry and optimization strategy. Dosimetric data were evaluated in accordance with the Radiation Therapy Oncology Group (RTOG) 1005 protocol. RESULTS: Overall, all approaches generated high-quality plans, with approximately 94% meeting the acceptable dose criteria for organs at risk and/or target coverage as defined by RTOG guidelines. Average automated plan generation time ranged from 6 min and 37 s to 9 min and 22 s, with the mean time increasing with additional fields. The Random Forest approach did not successfully enable automatic planning strategy selection. Instead, our automated planning system allows users to choose from the tested geometry and strategy options. CONCLUSIONS: Although our attempt to correlate patient anatomical features with planning strategy using machine learning tools was unsuccessful, the resulting dosimetric outcomes proved satisfactory. Our algorithm consistently produced high-quality plans, offering significant time and efficiency advantages.


Subject(s)
Algorithms , Breast Neoplasms , Organs at Risk , Radiotherapy Dosage , Radiotherapy Planning, Computer-Assisted , Radiotherapy, Intensity-Modulated , Humans , Radiotherapy Planning, Computer-Assisted/methods , Radiotherapy, Intensity-Modulated/methods , Female , Breast Neoplasms/radiotherapy , Organs at Risk/radiation effects , Retrospective Studies , Automation , Prognosis
5.
J Appl Clin Med Phys ; 25(5): e14345, 2024 May.
Article in English | MEDLINE | ID: mdl-38664894

ABSTRACT

PURPOSE: To establish the clinical applicability of deep-learning organ-at-risk autocontouring models (DL-AC) for brain radiotherapy. The dosimetric impact of contour editing, prior to model training, on performance was evaluated for both CT and MRI-based models. The correlation between geometric and dosimetric measures was also investigated to establish whether dosimetric assessment is required for clinical validation. METHOD: CT and MRI-based deep learning autosegmentation models were trained using edited and unedited clinical contours. Autosegmentations were dosimetrically compared to gold standard contours for a test cohort. D1%, D5%, D50%, and maximum dose were used as clinically relevant dosimetric measures. The statistical significance of dosimetric differences between the gold standard and autocontours was established using paired Student's t-tests. Clinically significant cases were identified via dosimetric headroom to the OAR tolerance. Pearson's Correlations were used to investigate the relationship between geometric measures and absolute percentage dose changes for each autosegmentation model. RESULTS: Except for the right orbit, when delineated using MRI models, the dosimetric statistical analysis revealed no superior model in terms of the dosimetric accuracy between the CT DL-AC models or between the MRI DL-AC for any investigated brain OARs. The number of patients where the clinical significance threshold was exceeded was higher for the optic chiasm D1% than other OARs, for all autosegmentation models. A weak correlation was consistently observed between the outcomes of dosimetric and geometric evaluations. CONCLUSIONS: Editing contours before training the DL-AC model had no significant impact on dosimetry. The geometric test metrics were inadequate to estimate the impact of contour inaccuracies on dose. Accordingly, dosimetric analysis is needed to evaluate the clinical applicability of DL-AC models in the brain.


Subject(s)
Brain Neoplasms , Deep Learning , Magnetic Resonance Imaging , Organs at Risk , Radiotherapy Dosage , Radiotherapy Planning, Computer-Assisted , Tomography, X-Ray Computed , Humans , Organs at Risk/radiation effects , Magnetic Resonance Imaging/methods , Tomography, X-Ray Computed/methods , Brain Neoplasms/radiotherapy , Brain Neoplasms/diagnostic imaging , Radiotherapy Planning, Computer-Assisted/methods , Radiotherapy, Intensity-Modulated/methods , Radiometry/methods , Image Processing, Computer-Assisted/methods
6.
J Appl Clin Med Phys ; 25(5): e14366, 2024 May.
Article in English | MEDLINE | ID: mdl-38669190

ABSTRACT

PURPOSE: Skin collimation is a useful tool in electron beam therapy (EBT) to decrease the penumbra at the field edge and minimize dose to nearby superficial organs at risk (OARs), but manually fabricating these collimation devices in the clinic to conform to the patient's anatomy can be a difficult and time intensive process. This work compares two types of patient-specific skin collimation (in-house 3D printed and vendor-provided machined brass) using clinically relevant metrics. METHODS: Attenuation measurements were performed to determine the thickness of each material needed to adequately shield both 6 and 9 MeV electron beams. Relative and absolute dose planes at various depths were measured using radiochromic film to compare the surface dose, flatness, and penumbra of the different skin collimation materials. RESULTS: Clinically acceptable thicknesses of each material were determined for both 6 and 9 MeV electron beams. Field width, flatness, and penumbra results between the two systems were very similar and significantly improved compared to measurements performed with no surface collimation. CONCLUSION: Both skin collimation methods investigated in this work generate sharp penumbras at the field edge and can minimize dose to superficial OARs compared to treatment fields with no surface collimation. The benefits of skin collimation are greatest for lower energy electron beams, and the benefits decrease as the measurement depth increases. Using bolus with skin collimation is recommended to avoid surface dose enhancement seen with collimators placed on the skin surface. Ultimately, the appropriate choice of material will depend on the desire to create these devices in-house or outsource the fabrication to a vendor.


Subject(s)
Electrons , Organs at Risk , Printing, Three-Dimensional , Radiotherapy Dosage , Radiotherapy Planning, Computer-Assisted , Skin , Humans , Electrons/therapeutic use , Radiotherapy Planning, Computer-Assisted/methods , Organs at Risk/radiation effects , Skin/radiation effects , Phantoms, Imaging , Neoplasms/radiotherapy , Particle Accelerators/instrumentation
7.
Asian Pac J Cancer Prev ; 25(4): 1383-1390, 2024 04 01.
Article in English | MEDLINE | ID: mdl-38680000

ABSTRACT

BACKGROUND: The study aims to investigate potential dosimetric benefits between non-coplanar and coplanar beam arrangements of Volumetric-Modulated Arc Therapy (VMAT) plans for liver stereotactic body radiotherapy (SBRT). METHODS: Thirteen patients who had undergone liver SBRT treatment in our department were chosen retrospectively for the study. Two sets of SBRT-VMAT plans namely, non-coplanar (NC-VMAT) and Coplanar (C-VMAT) were generated in Monaco(v5.11) planning system for Elekta Versa HD Linac using unflatten 6MV photon. The NC-VMAT plans were created by two/three non-coplanar partial arcs with couch rotation of ±150 and had an arc span of 1300 to 1600 whereas the C-VMAT plans consisted of a full arc. Both plans were compared by statistically analyzing various dosimetric and technical parameters. RESULTS: There is no statistically significant difference observed between the C-VMAT and NC-VMAT plans for planning target volume (PTV) coverage. However, the spine dose (D1cc) was much less in the NC-VMAT plan compared to the C-VMAT plan, with mean values of 6.127 ± 3.08Gy and 9.058 ± 4.76Gy, respectively (p-value=0.002). The low dose spillage to the healthy tissue was compared by the volume receiving 5Gy (V5Gy) and 10Gy (V10Gy). V5Gy of the NC-VMAT plan was 2399.23±1870.76cc while that of C-VMAT plans was 2835.36±1930.20cc with the p-value <0.001. Moreover, the monitor units(MU) were less with NC-VMAT than with C-VMAT SBRT plans (p=0.015). CONCLUSION: The plan quality of NC-VMAT plans was favorable compared to C-VMAT plans for liver SBRT especially in reducing spine dose, low dose spillage to healthy tissue, and MU.


Subject(s)
Liver Neoplasms , Organs at Risk , Radiosurgery , Radiotherapy Dosage , Radiotherapy Planning, Computer-Assisted , Radiotherapy, Intensity-Modulated , Humans , Radiosurgery/methods , Radiotherapy Planning, Computer-Assisted/methods , Radiotherapy, Intensity-Modulated/methods , Liver Neoplasms/radiotherapy , Liver Neoplasms/surgery , Retrospective Studies , Organs at Risk/radiation effects , Prognosis , Male , Female , Follow-Up Studies , Aged , Middle Aged
8.
Technol Cancer Res Treat ; 23: 15330338241241898, 2024.
Article in English | MEDLINE | ID: mdl-38557213

ABSTRACT

Introduction: In this study, we sought to develop a thermoplastic patient-specific helmet bolus that could deliver a uniform therapeutic dose to the target and minimize the dose to the normal brain during whole-scalp treatment with a humanoid head phantom. Methods: The bolus material was a commercial thermoplastic used for patient immobilization, and the holes in the netting were filled with melted paraffin. We compared volumetric-modulated arc therapy treatment plans with and without the bolus for quantitative dose distribution analysis. We analyzed the dose distribution in the region of interest to compare dose differences between target and normal organs. For quantitative analysis of treatment dose, OSLD chips were attached at the vertex (VX), posterior occipital (PO), right (RT), and left temporal (LT) locations. Results: The average dose in the clinical target volume was 6553.8 cGy (99.3%) with bolus and 5874 cGy (89%) without bolus, differing by more than 10% from the prescribed dose (6600 cGy) to the scalp target. For the normal brain, it was 3747.8 cGy (56.8%) with bolus and 5484.6 cGy (83.1%) without bolus. These results show that while the dose to the treatment target decreased, the average dose to the normal brain, which is mostly inside the treatment target, increased by more than 25%. With the bolus, the OSLD measured dose was 102.5 ± 1.2% for VX and 101.5 ± 1.9%, 95.9 ± 1.9%, and 81.8 ± 2.1% for PO, RT, and LT, respectively. In addition, the average dose in the treatment plan was 102%, 101%, 93.6%, and 80.7% for VX, PO, RT, and LT. When no bolus was administered, 59.6 ± 2.4%, 112.6 ± 1.8%, 47.1 ± 1.6%, and 53.1 ± 2.3% were assessed as OSLD doses for VX, PO, RT, and LT, respectively. Conclusion: This study proposed a method to fabricate patient-specific boluses that are highly reproducible, accessible, and easy to fabricate for radiotherapy to the entire scalp and can effectively spare normal tissue while delivering sufficient surface dose.


Subject(s)
Organothiophosphorus Compounds , Radiotherapy, Intensity-Modulated , Humans , Radiotherapy, Intensity-Modulated/methods , Scalp , Radiotherapy Dosage , Radiotherapy Planning, Computer-Assisted/methods , Feasibility Studies , Head Protective Devices , Organs at Risk/radiation effects
9.
Acta Oncol ; 63: 240-247, 2024 Apr 29.
Article in English | MEDLINE | ID: mdl-38682458

ABSTRACT

BACKGROUND AND PURPOSE: Doses to the coronary arteries in breast cancer (BC) radiotherapy (RT) have been suggested to be a risk predictor of long-term cardiac toxicity after BC treatment. We investigated the dose-risk relationships between near maximum doses (Dmax) to the right coronary artery (RCA) and left anterior descending coronary artery (LAD) and ischemic heart disease (IHD) mortality after BC RT. PATIENTS AND METHODS: In a cohort of 2,813 women diagnosed with BC between 1958 and 1992 with a follow-up of at least 10 years, we identified 134 cases of death due to IHD 10-19 years after BC diagnosis. For each case, one control was selected within the cohort matched for age at diagnosis. 3D-volume and 3D-dose reconstructions were obtained from individual RT charts. We estimated the Dmax to the RCA and the LAD and the mean heart dose (MHD). We performed conditional logistic regression analysis comparing piecewise spline transformation and simple linear modeling for best fit. RESULTS: There was a linear dose-risk relationship for both the Dmax to the RCA (odds ratio [OR]/Gray [Gy] 1.03 [1.01-1.05]) and the LAD (OR/Gy 1.04 [1.02-1.06]) in a multivariable model. For MHD there was a linear dose-risk relationship (1,14 OR/Gy [1.08-1.19]. For all relationships, simple linear modelling was superior to spline transformations. INTERPRETATION: Doses to both the RCA and LAD are independent risk predictors of long-term cardiotoxicity after RT for BC In addition to the LAD, the RCA should be regarded as an organ at risk in RT planning.


Subject(s)
Breast Neoplasms , Coronary Vessels , Myocardial Ischemia , Humans , Female , Breast Neoplasms/radiotherapy , Breast Neoplasms/mortality , Case-Control Studies , Middle Aged , Coronary Vessels/radiation effects , Coronary Vessels/pathology , Myocardial Ischemia/etiology , Myocardial Ischemia/mortality , Aged , Adult , Radiation Injuries/etiology , Radiation Injuries/epidemiology , Radiation Injuries/mortality , Radiotherapy Dosage , Dose-Response Relationship, Radiation , Organs at Risk/radiation effects , Follow-Up Studies , Cohort Studies
10.
Phys Med ; 121: 103346, 2024 May.
Article in English | MEDLINE | ID: mdl-38608421

ABSTRACT

Partial breast irradiation for the treatment of early-stage breast cancer patients can be performed by means of Intra Operative electron Radiation Therapy (IOeRT). One of the main limitations of this technique is the absence of a treatment planning system (TPS) that could greatly help in ensuring a proper coverage of the target volume during irradiation. An IOeRT TPS has been developed using a fast Monte Carlo (MC) and an ultrasound imaging system to provide the best irradiation strategy (electron beam energy, applicator position and bevel angle) and to facilitate the optimisation of dose prescription and delivery to the target volume while maximising the organs at risk sparing. The study has been performed in silico, exploiting MC simulations of a breast cancer treatment. Ultrasound-based input has been used to compute the absorbed dose maps in different irradiation strategies and a quantitative comparison between the different options was carried out using Dose Volume Histograms. The system was capable of exploring different beam energies and applicator positions in few minutes, identifying the best strategy with an overall computation time that was found to be completely compatible with clinical implementation. The systematic uncertainty related to tissue deformation during treatment delivery with respect to imaging acquisition was taken into account. The potential and feasibility of a GPU based full MC TPS implementation of IOeRT breast cancer treatments has been demonstrated in-silico. This long awaited tool will greatly improve the treatment safety and efficacy, overcoming the limits identified within the clinical trials carried out so far.


Subject(s)
Breast Neoplasms , Monte Carlo Method , Radiotherapy Planning, Computer-Assisted , Breast Neoplasms/radiotherapy , Breast Neoplasms/diagnostic imaging , Humans , Radiotherapy Planning, Computer-Assisted/methods , Radiotherapy Dosage , Electrons/therapeutic use , Time Factors , Computer Graphics , Female , Organs at Risk/radiation effects
11.
Phys Med ; 121: 103369, 2024 May.
Article in English | MEDLINE | ID: mdl-38669811

ABSTRACT

PURPOSE: In radiotherapy it is often necessary to transfer a patient's DICOM (Digital Imaging and COmmunications in Medicine) dataset from one system to another for re-treatment, plan-summation or registration purposes. The aim of the study is to evaluate effects of dataset transfer between treatment planning systems. MATERIALS AND METHODS: Twenty-five patients treated in a 0.35T MR-Linac (MRidian, ViewRay) for locally-advanced pancreatic cancer were enrolled. For each patient, a nominal dose distribution was optimized on the planning MRI. Each plan was daily re-optimized if needed to match the anatomy and exported from MRIdian-TPS (ViewRay Inc.) to Eclipse-TPS (Siemens-Varian). A comparison between the two TPSs was performed considering the PTV and OARs volumes (cc), as well as dose coverages and clinical constraints. RESULTS: From the twenty-five enrolled patients, 139 plans were included in the data comparison. The median values of percentage PTV volume variation are 10.8 % for each fraction, while percentage differences of PTV coverage have a mean value of -1.4 %. The median values of the percentage OARs volume variation are 16.0 %, 7.0 %, 10.4 % and 8.5 % for duodenum, stomach, small and large bowel, respectively. The percentage variations of the dose constraints are 41.0 %, 52.7 % and 49.8 % for duodenum, stomach and small bowel, respectively. CONCLUSIONS: This study has demonstrated a non-negligible variation in size and dosimetric parameters when datasets are transferred between TPSs. Such variations should be clinically considered. Investigations are focused on DICOM structure algorithm employed by the TPSs during the transfer to understand the cause of such variations.


Subject(s)
Pancreatic Neoplasms , Radiometry , Radiotherapy Dosage , Radiotherapy Planning, Computer-Assisted , Radiotherapy Planning, Computer-Assisted/methods , Humans , Pancreatic Neoplasms/radiotherapy , Pancreatic Neoplasms/diagnostic imaging , Organs at Risk/radiation effects , Magnetic Resonance Imaging
12.
Radiat Oncol ; 19(1): 53, 2024 Apr 30.
Article in English | MEDLINE | ID: mdl-38689338

ABSTRACT

PURPOSE: The number of older adults with head and neck squamous cell carcinoma (HNSCC) is continuously increasing. Older HNSCC patients may be more vulnerable to radiotherapy-related toxicities, so that extrapolation of available normal tissue complication probability (NTCP) models to this population may not be appropriate. Hence, we aimed to investigate the correlation between organ at risk (OAR) doses and chronic toxicities in older patients with HNSCC undergoing definitive radiotherapy. METHODS: Patients treated with definitive radiotherapy, either alone or with concomitant systemic treatment, between 2009 and 2019 in a large tertiary cancer center were eligible for this analysis. OARs were contoured based on international consensus guidelines, and EQD2 doses using α/ß values of 3 Gy for late effects were calculated based on the radiation treatment plans. Treatment-related toxicities were graded according to Common Terminology Criteria for Adverse Events version 5.0. Logistic regression analyses were carried out, and NTCP models were developed and internally validated using the bootstrapping method. RESULTS: A total of 180 patients with a median age of 73 years fulfilled the inclusion criteria and were analyzed. Seventy-three patients developed chronic moderate xerostomia (grade 2), 34 moderate dysgeusia (grade 2), and 59 moderate-to-severe (grade 2-3) dysphagia after definitive radiotherapy. The soft palate dose was significantly associated with all analyzed toxicities (xerostomia: OR = 1.028, dysgeusia: OR = 1.022, dysphagia: OR = 1.027) in the multivariable regression. The superior pharyngeal constrictor muscle was also significantly related to chronic dysphagia (OR = 1.030). Consecutively developed and internally validated NTCP models were predictive for the analyzed toxicities (optimism-corrected AUCs after bootstrapping: AUCxerostomia=0.64, AUCdysgeusia=0.60, AUCdysphagia=0.64). CONCLUSIONS: Our data suggest that the dose to the soft palate is associated with chronic moderate xerostomia, moderate dysgeusia and moderate-to-severe dysphagia in older HNSCC patients undergoing definitive radiotherapy. If validated in external studies, efforts should be undertaken to reduce the soft palate dose in these patients.


Subject(s)
Head and Neck Neoplasms , Organs at Risk , Palate, Soft , Radiation Injuries , Radiotherapy Dosage , Squamous Cell Carcinoma of Head and Neck , Humans , Aged , Female , Male , Head and Neck Neoplasms/radiotherapy , Organs at Risk/radiation effects , Palate, Soft/radiation effects , Radiation Injuries/etiology , Aged, 80 and over , Middle Aged , Squamous Cell Carcinoma of Head and Neck/radiotherapy , Retrospective Studies , Radiotherapy Planning, Computer-Assisted/methods
13.
J Appl Clin Med Phys ; 25(5): e14318, 2024 May.
Article in English | MEDLINE | ID: mdl-38427776

ABSTRACT

PURPOSE: To quantify the impact of treatment planning system beam model parameters, based on the actual spread in radiotherapy community data, on clinical treatment plans and determine which complexity metrics best describe the impact beam modeling errors have on dose accuracy. METHODS: Ten beam modeling parameters for a Varian accelerator were modified in RayStation to match radiotherapy community data at the 2.5, 25, 50, 75, and 97.5 percentile levels. These modifications were evaluated on 25 patient cases, including prostate, non-small cell lung, H&N, brain, and mesothelioma, generating 1,000 plan perturbations. Differences in the mean planned dose to clinical target volumes (CTV) and organs at risk (OAR) were evaluated with respect to the planned dose using the reference (50th-percentile) parameter values. Correlation between CTV dose differences, and 18 different complexity metrics were evaluated using linear regression; R-squared values were used to determine the best metric. RESULTS: Perturbations to MLC offset and transmission parameters demonstrated the greatest changes in dose: up to 5.7% in CTVs and 16.7% for OARs. More complex clinical plans showed greater dose perturbation with atypical beam model parameters. The mean MLC Gap and Tongue & Groove index (TGi) complexity metrics best described the impact of TPS beam modeling variations on clinical dose delivery across all anatomical sites; similar, though not identical, trends between complexity and dose perturbation were observed among all sites. CONCLUSION: Extreme values for MLC offset and MLC transmission beam modeling parameters were found to most substantially impact the dose distribution of clinical plans and careful attention should be given to these beam modeling parameters. The mean MLC Gap and TGi complexity metrics were best suited to identifying clinical plans most sensitive to beam modeling errors; this could help provide focus for clinical QA in identifying unacceptable plans.


Subject(s)
Neoplasms , Organs at Risk , Particle Accelerators , Radiotherapy Dosage , Radiotherapy Planning, Computer-Assisted , Radiotherapy, Intensity-Modulated , Humans , Radiotherapy Planning, Computer-Assisted/methods , Radiotherapy, Intensity-Modulated/methods , Organs at Risk/radiation effects , Neoplasms/radiotherapy , Particle Accelerators/instrumentation , Algorithms
14.
Radiat Prot Dosimetry ; 200(6): 538-543, 2024 Apr 20.
Article in English | MEDLINE | ID: mdl-38441907

ABSTRACT

The objective of this study is dosimetric comparison between the O-ring Halcyon and C-arm Clinac iX linac for volumetric modulated arc therapy (VMAT) plans for head & neck (H&N) cancer and carcinoma cervix patients. Total 60 patients of H&N cancer and carcinoma cervix were enrolled prospectively from March 2021 to March 2023. VMAT plans with 6 MV photons for Halcyon and Clinac iX were generated and compared for each patient by dose volume histogram for planning target volume coverage and organ at risk (OAR) sparing. There were no differences in between both the linacs for PTV D2% and D98%, homogeneity index, conformity index, Dmax (maximum dose) and Dmean (mean dose) of OAR. Halcyon had significantly shorter treatment time compared to Clinac iX. Halcyon delivered higher integral dose and monitor units. O-ring Halcyon produces VMAT plans comparable to other C-arm linacs for H&N and carcinoma cervix patients.


Subject(s)
Head and Neck Neoplasms , Organs at Risk , Particle Accelerators , Radiotherapy Dosage , Radiotherapy Planning, Computer-Assisted , Radiotherapy, Intensity-Modulated , Uterine Cervical Neoplasms , Humans , Radiotherapy, Intensity-Modulated/methods , Female , Uterine Cervical Neoplasms/radiotherapy , Radiotherapy Planning, Computer-Assisted/methods , Particle Accelerators/instrumentation , Head and Neck Neoplasms/radiotherapy , Organs at Risk/radiation effects , Radiometry/methods , Prospective Studies , Middle Aged , Adult
15.
Radiother Oncol ; 194: 110199, 2024 May.
Article in English | MEDLINE | ID: mdl-38438017

ABSTRACT

INTRODUCTION: Current consensus guidelines for definitive cervical cancer intensity modulated radiation therapy (IMRT) recommend inclusion of the entire uterus within the clinical target volume, however this is debated. We aimed to evaluate outcomes of patients with cervical cancer who were treated with less than whole uterus irradiation. METHODS: We identified 109 patients with FIGO Stage IB-IVA cervical cancer treated definitively with concurrent chemoradiation, including IMRT and brachytherapy, from 2010 to 2022 at a single institution where the practice was to include the gross cervix tumor with an internal target volume with differences in bladder filing accounted for, plus additional 5 mm planning target volume (PTV) margin. Local, regional, and distant recurrences were analyzed using competing risk methods, and a Wilcoxon rank sum test was performed to assess differences in dose to organs at risk based on the proportion of the uterus included in the PTV, with the median proportion of the uterus included (75 %) used as the cut-point. RESULTS: The median follow-up time was 65 months (range 3-352 months). The 2-year cumulative incidence of LR for the entire cohort was 4.2 % (95 % confidence interval [CI] 1.3-9.7). Compared with patients who had ≥ 75 % of the uterus included in the PTV, patients who had < 75 % of the uterus included in the PTV had significantly lower bowel D200cc (p = 0.02). The cumulative incidence of local failure (LR) was not significantly different between the two groups. CONCLUSIONS: Including less than the whole uterus for definitive cervix cancer IMRT does not seem to compromise local control. Less than whole uterus irradiation could be considered for carefully selected cervix cancer patients to decrease bowel dose and possible treatment-related toxicity.


Subject(s)
Brachytherapy , Chemoradiotherapy , Radiotherapy, Intensity-Modulated , Uterine Cervical Neoplasms , Uterus , Humans , Female , Uterine Cervical Neoplasms/radiotherapy , Uterine Cervical Neoplasms/pathology , Middle Aged , Radiotherapy, Intensity-Modulated/methods , Adult , Aged , Brachytherapy/methods , Brachytherapy/adverse effects , Uterus/radiation effects , Uterus/pathology , Chemoradiotherapy/methods , Organs at Risk/radiation effects , Radiotherapy Dosage , Aged, 80 and over , Neoplasm Recurrence, Local/pathology , Neoplasm Staging , Retrospective Studies
16.
Radiother Oncol ; 194: 110195, 2024 May.
Article in English | MEDLINE | ID: mdl-38442840

ABSTRACT

BACKGROUND AND PURPOSE: Partial breast irradiation (PBI)has beenthe Danish Breast Cancer Group(DBCG) standard for selected breast cancer patients since 2016 based onearlyresults from the DBCG PBI trial.During trial accrual, respiratory-gated radiotherapy was introduced in Denmark. This study aims to investigate the effect of respiratory-gating on mean heart dose (MHD). PATIENTS AND METHODS: From 2009 to 2016 the DBCG PBI trial included 230 patientswith left-sided breast cancer receiving external beam PBI, 40 Gy/15 fractions/3 weeks.Localization of the tumor bed on the planning CT scan, the use of respiratory-gating, coverage of the clinical target volume (CTV), and doses to organs at risk were collected. RESULTS: Respiratory-gating was used in 123 patients (53 %). In 176 patients (77 %) the tumor bed was in the upper and in 54 patients (23 %) in the lower breast quadrants. The median MHD was 0.37 Gy (interquartile range 0.26-0.57 Gy), 0.33 Gy (0.23-0.49 Gy) for respiratory-gating, and 0.49 Gy (0.31-0.70 Gy) for free breathing, p < 0.0001. MHD was < 1 Gy in 206 patients (90 %) and < 2 Gy in 221 patients (96 %). Respiratory-gating led to significantly lower MHD for upper-located, but not for lower-located tumor beds, however, all MHD were low irrespective of respiratory-gating. Respiratory-gating did not improve CTV coverage or lower lung doses. CONCLUSIONS: PBI ensured a low MHD for most patients. Adding respiratory-gating further reduced MHD for upper-located but not for lower-located tumor beds but did not influence target coverage or lung doses. Respiratory-gating is no longer DBCG standard for left-sided PBI.


Subject(s)
Organs at Risk , Humans , Female , Middle Aged , Organs at Risk/radiation effects , Denmark , Aged , Breast Neoplasms/radiotherapy , Breast Neoplasms/pathology , Unilateral Breast Neoplasms/radiotherapy , Radiotherapy Dosage , Heart/radiation effects , Radiotherapy Planning, Computer-Assisted/methods , Respiratory-Gated Imaging Techniques/methods , Adult
17.
Phys Med ; 120: 103331, 2024 Apr.
Article in English | MEDLINE | ID: mdl-38484461

ABSTRACT

PURPOSE: Within a multi-institutional project, we aimed to assess the transferability of knowledge-based (KB) plan prediction models in the case of whole breast irradiation (WBI) for left-side breast irradiation with tangential fields (TF). METHODS: Eight institutions set KB models, following previously shared common criteria. Plan prediction performance was tested on 16 new patients (2 pts per centre) extracting dose-volume-histogram (DVH) prediction bands of heart, ipsilateral lung, contralateral lung and breast. The inter-institutional variability was quantified by the standard deviations (SDint) of predicted DVHs and mean-dose (Dmean). The transferability of models, for the heart and the ipsilateral lung, was evaluated by the range of geometric Principal Component (PC1) applicability of a model to test patients of the other 7 institutions. RESULTS: SDint of the DVH was 1.8 % and 1.6 % for the ipsilateral lung and the heart, respectively (20 %-80 % dose range); concerning Dmean, SDint was 0.9 Gy and 0.6 Gy for the ipsilateral lung and the heart, respectively (<0.2 Gy for contralateral organs). Mean predicted doses ranged between 4.3 and 5.9 Gy for the ipsilateral lung and 1.1-2.3 Gy for the heart. PC1 analysis suggested no relevant differences among models, except for one centre showing a systematic larger sparing of the heart, concomitant to a worse PTV coverage, due to high priority in sparing the left anterior descending coronary artery. CONCLUSIONS: Results showed high transferability among models and low inter-institutional variability of 2% for plan prediction. These findings encourage the building of benchmark models in the case of TF-WBI.


Subject(s)
Breast Neoplasms , Radiotherapy, Intensity-Modulated , Thoracic Wall , Humans , Female , Radiotherapy, Intensity-Modulated/methods , Radiotherapy Dosage , Radiotherapy Planning, Computer-Assisted/methods , Breast , Organs at Risk/radiation effects
18.
Clin Oncol (R Coll Radiol) ; 36(6): e154-e162, 2024 Jun.
Article in English | MEDLINE | ID: mdl-38553363

ABSTRACT

BACKGROUND AND PURPOSE: For high-risk neuroblastoma, planning target volume coverage is often compromised to respect adjacent kidney tolerance. This trial investigated whether intensity-modulated arc radiotherapy techniques (IMAT) could facilitate dose escalation better than conventional techniques. MATERIALS AND METHODS: Children with high-risk abdominal neuroblastoma referred for radiotherapy to the primary tumour site and involved regional lymph nodes were randomised to receive either standard dose (21 Gy in 14 fractions) or escalated dose (36 Gy in 24 fractions) radiotherapy. Dual planning with both a conventional anterior-posterior parallel opposed pair radiotherapy technique and an IMAT technique was performed. The quality of target volume and organ-at-risk delineation, and dosimetric plans, were externally reviewed. Dosimetric parameters were used to judge the superior technique for treatment. This feasibility trial was not powered to detect improvement in outcome with dose escalation. RESULTS: Between 2017 and 2020, 50 patients were randomised and dual-planned. The IMAT technique was judged more favourable in 48 patients. In all patients randomised to receive 36 Gy, IMAT would have permitted delivery of the full dose (median D50% 36.0 Gy, inter-quartile range 36.0-36.1 Gy) to the target volume, whereas dose compromise would have been required with conventional planning (median D50% 35.6 Gy, inter-quartile range 28.7-35.9 Gy). CONCLUSION: IMAT facilitates safe dose escalation to 36 Gy in patients receiving radiotherapy for neuroblastoma. The value of dose escalation is now being evaluated in a current prospective phase III randomised trial.


Subject(s)
Feasibility Studies , Neuroblastoma , Radiotherapy Dosage , Radiotherapy, Intensity-Modulated , Humans , Neuroblastoma/radiotherapy , Radiotherapy, Intensity-Modulated/methods , Male , Female , Child, Preschool , Child , Infant , Radiotherapy Planning, Computer-Assisted/methods , Organs at Risk/radiation effects
19.
Radiother Oncol ; 194: 110184, 2024 May.
Article in English | MEDLINE | ID: mdl-38453055

ABSTRACT

BACKGROUND AND PURPOSE: Safe reirradiation relies on assessment of cumulative doses to organs at risk (OARs) across multiple treatments. Different clinical pathways can result in inconsistent estimates. Here, we quantified the consistency of cumulative dose to OARs across multi-centre clinical pathways. MATERIAL AND METHODS: We provided DICOM planning CT, structures and doses for two reirradiation cases: head & neck (HN) and lung. Participants followed their standard pathway to assess the cumulative physical and EQD2 doses (with provided α/ß values), and submitted DVH metrics and a description of their pathways. Participants could also submit physical dose distributions from Course 1 mapped onto the CT of Course 2 using their best available tools. To assess isolated impact of image registrations, a single observer accumulated each submitted spatially mapped physical dose for every participating centre. RESULTS: Cumulative dose assessment was performed by 24 participants. Pathways included rigid (n = 15), or deformable (n = 5) image registration-based 3D dose summation, visual inspection of isodose line contours (n = 1), or summation of dose metrics extracted from each course (n = 3). Largest variations were observed in near-maximum cumulative doses (25.4 - 41.8 Gy for HN, 2.4 - 33.8 Gy for lung OARs), with lower variations in volume/dose metrics to large organs. A standardised process involving spatial mapping of the first course dose to the second course CT followed by summation improved consistency for most near-maximum dose metrics in both cases. CONCLUSION: Large variations highlight the uncertainty in reporting cumulative doses in reirradiation scenarios, with implications for outcome analysis and understanding of published doses. Using a standardised workflow potentially including spatially mapped doses improves consistency in determination of accumulated dose in reirradiation scenarios.


Subject(s)
Head and Neck Neoplasms , Lung Neoplasms , Organs at Risk , Radiotherapy Dosage , Radiotherapy Planning, Computer-Assisted , Re-Irradiation , Humans , Re-Irradiation/methods , Head and Neck Neoplasms/radiotherapy , Radiotherapy Planning, Computer-Assisted/methods , Organs at Risk/radiation effects , Lung Neoplasms/radiotherapy , Lung Neoplasms/diagnostic imaging , Tomography, X-Ray Computed
20.
J Appl Clin Med Phys ; 25(5): e14350, 2024 May.
Article in English | MEDLINE | ID: mdl-38546277

ABSTRACT

OBJECTIVE: Adaptive planning to accommodate anatomic changes during treatment often requires repeated segmentation. In this study, prior patient-specific data was integrateda into a registration-guided multi-channel multi-path (Rg-MCMP) segmentation framework to improve the accuracy of repeated clinical target volume (CTV) segmentation. METHODS: This study was based on CT image datasets for a total of 90 cervical cancer patients who received two courses of radiotherapy. A total of 15 patients were selected randomly as the test set. In the Rg-MCMP segmentation framework, the first-course CT images (CT1) were registered to second-course CT images (CT2) to yield aligned CT images (aCT1), and the CTV in the first course (CTV1) was propagated to yield aligned CTV contours (aCTV1). Then, aCT1, aCTV1, and CT2 were combined as the inputs for 3D U-Net consisting of a channel-based multi-path feature extraction network. The performance of the Rg-MCMP segmentation framework was evaluated and compared with the single-channel single-path model (SCSP), the standalone registration methods, and the registration-guided multi-channel single-path (Rg-MCSP) model. The Dice similarity coefficient (DSC), 95% Hausdorff distance (HD95), and average surface distance (ASD) were used as the metrics. RESULTS: The average DSC of CTV for the deformable image DIR-MCMP model was found to be 0.892, greater than that of the standalone DIR (0.856), SCSP (0.837), and DIR-MCSP (0.877), which were improvements of 4.2%, 6.6%, and 1.7%, respectively. Similarly, the rigid body DIR-MCMP model yielded an average DSC of 0.875, which exceeded standalone RB (0.787), SCSP (0.837), and registration-guided multi-channel single-path (0.848), which were improvements of 11.2%, 4.5%, and 3.2%, respectively. These improvements in DSC were statistically significant (p < 0.05). CONCLUSION: The proposed Rg-MCMP framework achieved excellent accuracy in CTV segmentation as part of the adaptive radiotherapy workflow.


Subject(s)
Radiotherapy Dosage , Radiotherapy Planning, Computer-Assisted , Radiotherapy, Intensity-Modulated , Tomography, X-Ray Computed , Uterine Cervical Neoplasms , Humans , Uterine Cervical Neoplasms/radiotherapy , Uterine Cervical Neoplasms/diagnostic imaging , Female , Radiotherapy Planning, Computer-Assisted/methods , Tomography, X-Ray Computed/methods , Radiotherapy, Intensity-Modulated/methods , Algorithms , Organs at Risk/radiation effects , Image Processing, Computer-Assisted/methods , Prognosis