1.
Appl Radiat Isot ; 202: 111064, 2023 Dec.
Artículo en Inglés | MEDLINE | ID: mdl-37866227

RESUMEN

An extrapolation chamber of type Böhm 23392, used for the dosimetry of RQR beam qualities at the Secondary Standard Dosimetry Laboratory of King Faisal Specialist Hospital, is presented. A computer-based motorized system consisting of a stepping motor coupled to the chamber's built-in micrometer screw was designed to expedite the measurements process, giving a linear relationship between the number of motor steps and the chamber depth with a Coefficient of Determination (COD) equal to 0.9990. The extrapolation chamber along with the motorized system was used to determine the extrapolation curves. The slope of the extrapolation curves, measured for the eight RQR beam qualities with a relative uncertainty between 0.17% and 0.58%, varies linearly with the beam quality expressed in terms of HVL, with a COD between 0.9995 and 1.0000.