1.
Clin Chim Acta ; 411(11-12): 806-11, 2010 Jun 03.
Artículo en Inglés | MEDLINE | ID: mdl-20188087

RESUMEN

BACKGROUND: Cyclosporine (CsA) monitoring is essential for transplant success. We report a performance study of the recently released, fully automated Siemens ADVIA Centaur CsA assay. METHODS: Whole blood samples from 248 transplant patients were prepared using a new 1-step extraction method. Performance evaluations vs. HPLC-tandem MS (LC-MS/MS), Abbott TDx and AxSYM assays were conducted according to CLSI EP5-A2 and EP9-A2 guidelines. RESULTS: The correlation coefficient for LC-MS/MS and ADVIA Centaur was > or = 0.97 at each site, and for each transplant type. Regression analysis yielded y=0.94x+19 for all sites: 95% CI=0.91-0.96 (slope) and 10-28 (intercept). Absolute and relative bias was minimal for C0 and C2 sampling. Centaur vs. Abbott TDx and AxSYM assays: y = 0.72x+6, r = 0.98, 95% CI = 0.70-0.73 (slope), 3-9 (intercept); and y = 0.69x+18, r = 0.97, 95% CI = 0.67-0.71 (slope), 8-27(intercept). Within run CVs were 4.5%-7.1%, total CVs were 5.3%-7.7%. CONCLUSIONS: The ADVIA Centaur assay compared favorably with LC-MS/MS and Abbott assays, displaying good correlation for all transplant types and methods.


Asunto(s)
Ciclosporina/sangre , Monitoreo de Drogas/normas , Adolescente , Adulto , Anciano , Anciano de 80 o más Años , Unión Competitiva , Ciclosporina/aislamiento & purificación , Monitoreo de Drogas/métodos , Femenino , Inmunoensayo de Polarización Fluorescente/métodos , Inmunoensayo de Polarización Fluorescente/normas , Humanos , Inmunosupresores/sangre , Inmunosupresores/aislamiento & purificación , Masculino , Persona de Mediana Edad , Valores de Referencia , Trasplante/normas , Adulto Joven
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA
...