1.
Appl Radiat Isot ; 123: 54-59, 2017 May.
Artículo en Inglés | MEDLINE | ID: mdl-28242294

RESUMEN

The BiPo-3 detector is a low radioactive detector dedicated to measuring ultra-low natural contaminations of 208Tl and 214Bi in thin materials, initially developed to measure the radiopurity of the double ß decay source foils of the SuperNEMO experiment at the µBq/kg level. The BiPo-3 technique consists in installing the foil of interest between two thin ultra-radiopure scintillators coupled to low radioactive photomultipliers. The design and performances of the detector are presented. In this paper, the final results of the 208Tl and 214Bi activity measurements of the first enriched 82Se foils are reported for the first time, showing the capability of the detector to reach sensitivities in the range of some µBq/kg.

2.
Phys Rev Lett ; 119(4): 041801, 2017 Jul 28.
Artículo en Inglés | MEDLINE | ID: mdl-29341770

RESUMEN

We report the results of a first experimental search for lepton number violation by four units in the neutrinoless quadruple-ß decay of ^{150}Nd using a total exposure of 0.19 kg yr recorded with the NEMO-3 detector at the Modane Underground Laboratory. We find no evidence of this decay and set lower limits on the half-life in the range T_{1/2}>(1.1-3.2)×10^{21} yr at the 90% C.L., depending on the model used for the kinematic distributions of the emitted electrons.