Antimicrob Resist Infect Control ; 11(1): 81, 2022 06 03.
Artículo en Inglés | MEDLINE | ID: mdl-35659363

RESUMEN

BACKGROUND: One possible transmission route for nosocomial pathogens is contaminated medical devices. Formation of biofilms can exacerbate the problem. We report on a carbapenemase-producing Klebsiella pneumoniae that had caused an outbreak linked to contaminated duodenoscopes. To determine whether increased tolerance to disinfectants may have contributed to the outbreak, we investigated the susceptibility of the outbreak strain to disinfectants commonly used for duodenoscope reprocessing. Disinfection efficacy was tested on planktonic bacteria and on biofilm. METHODS: Disinfectant efficacy testing was performed for planktonic bacteria according to EN standards 13727 and 14561 and for biofilm using the Bead Assay for Biofilms. Disinfection was defined as ≥ 5log10 reduction in recoverable colony forming units (CFU). RESULTS: The outbreak strain was an OXA-48 carbapenemase-producing K. pneumoniae of sequence type 101. We found a slightly increased tolerance of the outbreak strain in planktonic form to peracetic acid (PAA), but not to other disinfectants tested. Since PAA was the disinfectant used for duodenoscope reprocessing, we investigated the effect of PAA on biofilm of the outbreak strain. Remarkably, disinfection of biofilm of the outbreak strain could not be achieved by the standard PAA concentration used for duodenoscope reprocessing at the time of outbreak. An increased tolerance to PAA was not observed in a K. pneumoniae type strain tested in parallel. CONCLUSIONS: Biofilm of the K. pneumoniae outbreak strain was tolerant to standard disinfection during duodenoscope reprocessing. This study establishes for the first time a direct link between biofilm formation, increased tolerance to disinfectants, reprocessing failure of duodenoscopes and nosocomial transmission of carbapenem-resistant K. pneumoniae.


Asunto(s)
Enterobacteriaceae Resistentes a los Carbapenémicos , Infección Hospitalaria , Desinfectantes , Bacterias , Biopelículas , Carbapenémicos/farmacología , Infección Hospitalaria/epidemiología , Infección Hospitalaria/prevención & control , Brotes de Enfermedades , Desinfectantes/farmacología , Duodenoscopía , Humanos , Klebsiella pneumoniae , Ácido Peracético/farmacología
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA