Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 2 de 2
Filtrar
Más filtros










Base de datos
Intervalo de año de publicación
1.
Radiat Prot Dosimetry ; 156(3): 268-76, 2013 Sep.
Artículo en Inglés | MEDLINE | ID: mdl-23554424

RESUMEN

Photon and neutron soft tissue absorbed doses near the entrance door to the medical LINAC treatment room were measured with the use of thermoluminescent detectors LiF:Mg,Cu,P in the anthropomorphic phantom. Two different therapeutic beams (6 and 15 MV) and four treatment techniques were involved in the present study. This allowed one to investigate the contribution of scattered X rays, secondary neutrons and gamma rays to the radiation field. Photon absorbed dose rates 50 cm away from the LINAC room door during emission of 15-MV X rays varied between 4.1×10(-4) and 5.6×10(-4) Gy h(-1), depending on the gantry position and the irradiation field size, whereas in the case of 6-MV therapeutic irradiation these doses are ∼1.5 times lower. In the case of 15-MV beam emission, a mixed radiation field near the bunker door is observed with the photon radiation as the main component, which includes a 33.1 % contribution of the induced gamma radioactivity and ∼2.1 % contribution of the neutron radiation.


Asunto(s)
Aceleradores de Partículas , Radioterapia Conformacional/métodos , Dosimetría Termoluminiscente/instrumentación , Antropometría , Simulación por Computador , Rayos gamma , Humanos , Neutrones , Fantasmas de Imagen , Fotones , Dosis de Radiación , Medición de Riesgo , Rayos X
2.
Rep Pract Oncol Radiother ; 17(6): 339-46, 2012 Nov.
Artículo en Inglés | MEDLINE | ID: mdl-24669311

RESUMEN

BACKGROUND: High-energy photon and electron therapeutic beams generated in medical linear accelerators can cause the electronuclear and photonuclear reactions in which neutrons with a broad energy spectrum are produced. A low-energy component of this neutron radiation induces simple capture reactions from which various radioisotopes originate and in which the radioactivity of a linac head and various objects in the treatment room appear. AIM: The aim of this paper is to present the results of the thermal/resonance neutron fluence measurements during therapeutic beam emission and exemplary spectra of gamma radiation emitted by medical linac components activated in neutron reactions for four X-ray beams and for four electron beams generated by various manufacturers' accelerators installed in typical concrete bunkers in Polish oncological centers. MATERIALS AND METHODS: The measurements of neutron fluence were performed with the use of the induced activity method, whereas the spectra of gamma radiation from decays of the resulting radioisotopes were measured by means of a portable high-purity germanium detector set for field spectroscopy. RESULTS: The fluence of thermal neutrons as well as resonance neutrons connected with the emission of a 20 MV X-ray beam is ∼10(6) neutrons/cm(2) per 1 Gy of a dose in water at a reference depth. It is about one order of magnitude greater than that for the 15 MV X-ray beams and about two orders of magnitude greater than for the 18-22 MeV electron beams regardless of the type of an accelerator. CONCLUSION: The thermal as well as resonance neutron fluence depends strongly on the type and the nominal potential of a therapeutic beam. It is greater for X-ray beams than for electrons. The accelerator accessories and other large objects should not be stored in a treatment room during high-energy therapeutic beam emission to avoid their activation caused by thermal and resonance neutrons. Half-lives of the radioisotopes originating from the simple capture reaction (n,γ) (from minutes to hours) are long enough to accumulate radioactivity of components of the accelerator head. The radiation emitted by induced radioisotopes causes the additional doses to staff operating the accelerators.

SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA
...