Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Más filtros










Base de datos
Intervalo de año de publicación
1.
J Clin Monit Comput ; 34(6): 1361-1367, 2020 Dec.
Artículo en Inglés | MEDLINE | ID: mdl-31773375

RESUMEN

An accurate determination of body core temperature is crucial during surgery in order to avoid and treat hypothermia, which is associated with poor outcome. In a prospective observational study, we evaluated the suitability of the Tcore™ device (Drägerwerk AG & Co. KGaA, Lübeck, Germany)-a non-invasive thermometer-to accurately determine core body temperature. In patients undergoing surgery for ovarian cancer, core body temperature (CBT) was determined with the Tcore™ sensor attached to the forehead and compared with blood temperature (Tblood) as measured within the femoro-iliacal artery. Both temperatures were recorded every 10 s and the measurement error was calculated. 57,302 data pairs of CBT and Tblood were obtained in 22 patients. In a repeated-measurements version of the Bland and Altman test, a bias of - 0.02 °C and 95% limits of agreement of - 0.48 to 0.44 °C were calculated. In a population analysis, a median absolute error of 0 [- 0.1; + 0.1] °C, a bias of 0 [- 0.276; 0.271] % and an inaccuracy of 0.276 [0.274; 0.354] % was determined. Although the Tcore™ sensor was attached to the frontal skin, it provided an accurate measurement of core body temperature in the investigated intraoperative setting.


Asunto(s)
Monitoreo Intraoperatorio , Termómetros , Temperatura Corporal , Humanos , Monitoreo Fisiológico , Temperatura
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA
...