J Ophthalmol ; 2015: 496382, 2015.
Artículo en Inglés | MEDLINE | ID: mdl-26075085

RESUMEN

Purpose. To thoroughly analyze corneal deformation responses curves obtained by Ocular Response Analyzer (ORA) testing in order to improve subclinical keratoconus detection. Methods. Observational case series of 87 control and 73 subclinical keratoconus eyes. Examination included corneal topography, tomography, and biomechanical testing with ORA. Factor analysis, logistic regression, and receiver operating characteristic curves were used to extract combinations of 45 corneal waveform descriptors. Main outcome measures were corneal-thickness-corrected corneal resistance factor (ccCRF), combinations of corneal descriptors, and their diagnostic performance. Results. Thirty-seven descriptors differed significantly in means between groups, and among them ccCRF afforded the highest individual diagnostic performance. Factor analysis identified first- and second-peak related descriptors as the most variable one. However, conventional biomechanical descriptors corneal resistance factor and hysteresis differed the most between control and keratoconic eyes. A combination of three factors including several corneal descriptors did not show better diagnostic performance than a combination of conventional indices. Conclusion. Multivariate analysis of ORA signals did not surpass simpler models in subclinical keratoconus detection, and there is considerable overlap between normal and ectatic eyes irrespective of the analysis model. Conventional biomechanical indices seem to already provide the best performance when appropriately considered.

2.
J Refract Surg ; 31(2): 116-23, 2015 Feb.
Artículo en Inglés | MEDLINE | ID: mdl-25735045

RESUMEN

PURPOSE: To evaluate intereye corneal asymmetry in Pentacam (Oculus Optikgeräte GmbH, Wetzlar, Germany) indices as a diagnostic method between normal patients and patients with keratoconus. METHODS: A retrospective, observational case series of 177 healthy, 44 indeterminate, and 121 patients with keratoconus classified by Pentacam ectasia detection indices, randomized to analysis and validation datasets. Intereye asymmetry in 20 Scheimpflug tomography corneal descriptors was calculated and compared to develop diagnostic models. RESULTS: Intereye asymmetry was not correlated with anisometropia in healthy controls but was correlated with the ectasia grade of the worse eye in patients with keratoconus. Patients with keratoconus had significantly greater intereye asymmetry in all descriptors except for relational thickness indices. Intereye asymmetry in front elevation at the thinnest corneal location afforded the single highest diagnostic performance (71% sensitivity and 85% specificity), whereas the best multivariate model combining intereye asymmetry in anterior and posterior keratometry, corneal thickness, and front and back elevation at the thinnest point provided 65% sensitivity and 97% specificity. Multivariate models upheld their performance in the validation dataset. Most (more than 90%) indeterminate patients, according to conventional Pentacam analysis, showed within-normal-range corneal asymmetry. CONCLUSIONS: Healthy corneas are markedly symmetric irrespective of anisometropia, but corneal asymmetry analysis does not provide sufficient sensitivity to be used alone for detecting keratoconus. However, its remarkable specificity suggests that it could be used combined with conventional single cornea Pentacam analysis to reduce the false-positive rate or in dubious cases.


Asunto(s)
Anisometropía/diagnóstico , Córnea/patología , Técnicas de Diagnóstico Oftalmológico , Queratocono/diagnóstico , Adulto , Femenino , Humanos , Masculino , Fotograbar , Curva ROC , Estudios Retrospectivos , Sensibilidad y Especificidad , Tomografía , Adulto Joven
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA