Phys Med ; 113: 102657, 2023 Sep.
Artículo en Inglés | MEDLINE | ID: mdl-37567068

RESUMEN

PURPOSE: Different methods are available to identify haematopoietically active bone marrow (ActBM). However, their use can be challenging for radiotherapy routine treatments, since they require specific equipment and dedicated time. A machine learning (ML) approach, based on radiomic features as inputs to three different classifiers, was applied to computed tomography (CT) images to identify haematopoietically active bone marrow in anal cancer patients. METHODS: A total of 40 patients was assigned to the construction set (training set + test set). Fluorine-18-Fluorodeoxyglucose Positron Emission Tomography (18FDG-PET) images were used to detect the active part of the pelvic bone marrow (ActPBM) and stored as ground-truth for three subregions: iliac, lower pelvis and lumbosacral bone marrow (ActIBM, ActLPBM, ActLSBM). Three parameters were used for the correspondence analyses between 18FDG-PET and ML classifiers: DICE index, Precision and Recall. RESULTS: For the 40-patient cohort, median values [min; max] of the Dice index were 0.69 [0.20; 0.84], 0.76 [0.25; 0.89], and 0.36 [0.15; 0.67] for ActIBM, ActLSBM, and ActLPBM, respectively. The Precision/Recall (P/R) ratio median value for the ActLPBM structure was 0.59 [0.20; 1.84] (over segmentation), while for the other two subregions the P/R ratio median has values of 1.249 [0.43; 4.15] for ActIBM and 1.093 [0.24; 1.91] for ActLSBM (under segmentation). CONCLUSION: A satisfactory degree of overlap compared to 18FDG-PET was found for 2 out of the 3 subregions within pelvic bones. Further optimization and generalization of the process is required before clinical implementation.


Asunto(s)
Neoplasias del Ano , Médula Ósea , Humanos , Médula Ósea/diagnóstico por imagen , Fluorodesoxiglucosa F18 , Tomografía de Emisión de Positrones/métodos , Tomografía Computarizada por Rayos X , Neoplasias del Ano/diagnóstico por imagen , Neoplasias del Ano/terapia , Aprendizaje Automático , Tomografía Computarizada por Tomografía de Emisión de Positrones/métodos , Radiofármacos , Estudios Retrospectivos
2.
Med Phys ; 42(9): 5287-300, 2015 Sep.
Artículo en Inglés | MEDLINE | ID: mdl-26328978

RESUMEN

PURPOSE: To describe the dosimetric commissioning and quality assurance (QA) of the actively scanned proton and carbon ion beams at the Italian National Center for Oncological Hadrontherapy. METHODS: The laterally integrated depth-dose-distributions (IDDs) were acquired with the PTW Peakfinder, a variable depth water column, equipped with two Bragg peak ionization chambers. fluka Monte Carlo code was used to generate the energy libraries, the IDDs in water, and the fragment spectra for carbon beams. EBT3 films were used for spot size measurements, beam position over the scan field, and homogeneity in 2D-fields. Beam monitor calibration was performed in terms of number of particles per monitor unit using both a Farmer-type and an Advanced Markus ionization chamber. The beam position at the isocenter, beam monitor calibration curve, dose constancy in the center of the spread-out-Bragg-peak, dose homogeneity in 2D-fields, beam energy, spot size, and spot position over the scan field are all checked on a daily basis for both protons and carbon ions and on all beam lines. RESULTS: The simulated IDDs showed an excellent agreement with the measured experimental curves. The measured full width at half maximum (FWHM) of the pencil beam in air at the isocenter was energy-dependent for both particle species: in particular, for protons, the spot size ranged from 0.7 to 2.2 cm. For carbon ions, two sets of spot size are available: FWHM ranged from 0.4 to 0.8 cm (for the smaller spot size) and from 0.8 to 1.1 cm (for the larger one). The spot position was accurate to within ± 1 mm over the whole 20 × 20 cm(2) scan field; homogeneity in a uniform squared field was within ± 5% for both particle types at any energy. QA results exceeding tolerance levels were rarely found. In the reporting period, the machine downtime was around 6%, of which 4.5% was due to planned maintenance shutdowns. CONCLUSIONS: After successful dosimetric beam commissioning, quality assurance measurements performed during a 24-month period show very stable beam characteristics, which are therefore suitable for performing safe and accurate patient treatments.


Asunto(s)
Radioterapia de Iones Pesados/métodos , Neoplasias/radioterapia , Terapia de Protones/métodos , Garantía de la Calidad de Atención de Salud , Calibración , Radioterapia de Iones Pesados/instrumentación , Radioterapia de Iones Pesados/normas , Método de Montecarlo , Fantasmas de Imagen , Terapia de Protones/instrumentación , Terapia de Protones/normas , Radiometría , Dosificación Radioterapéutica