AJR Am J Roentgenol ; 212(3): 607-613, 2019 03.
Artículo en Inglés | MEDLINE | ID: mdl-30645158

RESUMEN

OBJECTIVE: The eye lens is one of the most radiosensitive organs, and medical radiation is one of the main causes of cataracts. To protect the lens during head CT examinations, protectors have been developed; however, they can lead to image artifacts, which is a major disadvantage of their use. This study retrospectively evaluates the frequency and extent of artifacts caused by these protectors related to three anatomic regions (eye, brain, and bone) and their dependence on protector positioning. MATERIALS AND METHODS: Datasets from 261 consecutive head CT examinations obtained during 3.5 months of routine clinical imaging were assessed. Diagnostic quality of the images was evaluated by objective measuring and subjective scoring on a 5-point Likert scale. Furthermore, the position of the lens protector in correlation to the eye lens and the intensity and frequency of artifacts were analyzed. RESULTS: Only 4.6% of all analyzed examinations were completely free from artifacts; 95.4% showed artifacts at least in the orbital cavity. Although the brain was affected in 27.8% of cases, in only 5.8% of cases was there a risk of misinterpretation, such as suspected intracranial bleeding. In 24.9% of cases, the lens was not properly covered by the protector. A too cranial position of the protector was identified as the main risk factor for cerebral artifacts. CONCLUSION: Eye shielding for brain CT examinations often leads to artifacts. However, in only a small percentage of cases do these artifacts affect tissue depiction in regions beyond the eye (i.e., brain or bones). Correct positioning is mandatory to minimize artifacts.


Asunto(s)
Artefactos , Dispositivos de Protección de los Ojos , Cristalino/efectos de la radiación , Protección Radiológica/instrumentación , Tomografía Computarizada por Rayos X , Adolescente , Adulto , Anciano , Anciano de 80 o más Años , Niño , Preescolar , Femenino , Humanos , Lactante , Masculino , Persona de Mediana Edad , Posicionamiento del Paciente , Estudios Retrospectivos
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA