IEEE Trans Nucl Sci ; 60(1): 9-15, 2013 Feb.
Artículo en Inglés | MEDLINE | ID: mdl-28250476

RESUMEN

We are currently investigating the feasibility of using highly pixelated Cadmium Zinc Telluride (CdZnTe) detectors for sub-500 µm resolution PET imaging applications. A 20 mm × 20 mm × 5 mm CdZnTe substrate was fabricated with 350 µm pitch pixels (250 µm anode pixels with 100 µm gap) and coplanar cathode. Charge sharing among the pixels of a 350 µm pitch detector was studied using collimated 122 keV and 511 keV gamma ray sources. For a 350 µm pitch CdZnTe detector, scatter plots of the charge signal of two neighboring pixels clearly show more charge sharing when the collimated beam hits the gap between adjacent pixels. Using collimated Co-57 and Ge-68 sources, we measured the count profiles and estimated the intrinsic spatial resolution of 350 µm pitch detector biased at -1000 V. Depth of interaction was analyzed based on two methods, i.e., cathode/anode ratio and electron drift time, in both 122 keV and 511 keV measurements. For single-pixel photopeak events, a linear correlation between cathode/anode ratio and electron drift time was shown, which would be useful for estimating the DOI information and preserving image resolution in CdZnTe PET imaging applications.