Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Más filtros











Base de datos
Intervalo de año de publicación
1.
Sci Rep ; 7(1): 3685, 2017 06 16.
Artículo en Inglés | MEDLINE | ID: mdl-28623362

RESUMEN

The limits of conventional light microscopy ("Abbe-Limit") depend critically on the numerical aperture (NA) of the objective lens. Imaging at large working distances or a large field-of-view typically requires low NA objectives, thereby reducing the optical resolution to the multi micrometer range. Based on numerical simulations of the intensity field distribution, we present an illumination concept for a super-resolution microscope which allows a three dimensional (3D) optical resolution around 150 nm for working distances up to the centimeter regime. In principle, the system allows great flexibility, because the illumination concept can be used to approximate the point-spread-function of conventional microscope optics, with the additional benefit of a customizable pupil function. Compared with the Abbe-limit using an objective lens with such a large working distance, a volume resolution enhancement potential in the order of 104 is estimated.

SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA