Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Más filtros










Base de datos
Intervalo de año de publicación
1.
Respir Care ; 56(8): 1115-21, 2011 Aug.
Artículo en Inglés | MEDLINE | ID: mdl-21801578

RESUMEN

BACKGROUND: Diffusing capacity of the lung for carbon monoxide (D(LCO)) can be affected by abnormal hemoglobin (Hb) or carboxyhemoglobin (COHb) levels. Predicted D(LCO) can be adjusted to reflect abnormal Hb or COHb levels. Until recently, blood sampling was required to determine Hb and COHb levels, but a new pulse oximeter, the Masimo RAD-57, can measure Hb and COHb noninvasively. We hypothesized that there would be no significant difference between the invasive and noninvasive Hb and COHb measurements for adjusting D(LCO). METHODS: In patients referred to our university hospital for D(LCO) testing, we simultaneously took arterial blood gas samples and measured Hb and COHb with the RAD-57 (SpHb and SpCOHb, respectively). We analyzed the paired values and the Hb-adjusted and COHb-adjusted predicted D(LCO) values with t tests and Bland-Altman plots. We compared the differences in predicted D(LCO) to a clinical threshold of 3 mL/min/mm Hg. RESULTS: SpHb differed from Hb measured via arterial blood analysis (12.1 ± 2.4 g/dL vs 13.3 ± 2.1 g/dL, P < .001). SpCOHb did not differ significantly from COHb (ie, measured via arterial blood analysis) (2.1 ± 4.0 vs 2.5 ± 2.3, P = .25), but there was wide variability. There were small but statistically significant differences in the adjusted predicted D(LCO), depending on whether blood or pulse oximetry values were used. Predicted D(LCO) adjusted for both Hb and COHb was 22.5 ± 4.8 mL/min/mm Hg measured with the RAD-57 and 23.5 ± 4.5 mL/min/mm Hg via arterial blood analysis (P < .001). The limits of agreement for pulse oximetry adjusted D(LCO) exceeded the clinical threshold of 3 mL/min/mm Hg for Hb adjustments and combined Hb + COHb. Predicted D(LCO) values differed by > 3 mL/min/mm Hg in 17% of patients. CONCLUSIONS: Pulse oximetry may be of limited usefulness for adjusting either predicted or measured D(LCO) values, but might be useful to screen patients for invasive testing, particularly if the D(LCO) is close to the lower limit of normal.


Asunto(s)
Monóxido de Carbono/análisis , Carboxihemoglobina/análisis , Pulmón/metabolismo , Oximetría/instrumentación , Capacidad de Difusión Pulmonar/instrumentación , Diseño de Equipo , Estudios de Seguimiento , Humanos , Masculino , Persona de Mediana Edad , Reproducibilidad de los Resultados
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA
...