1.
Cureus ; 16(6): e61993, 2024 Jun.
Artículo en Inglés | MEDLINE | ID: mdl-38983996

RESUMEN

Objective The objective of this study was to compare K1, K2, Kmax, and pachymetry values from Pentacam and Galilei scans of corneal topography in order to assess their correlation and interchangeability in clinical practice. Methodology A total of 34 patients (68 eyes) were enrolled in the study. Corneal topography was performed using Pentacam and Galilei devices on the same day. K1, K2, Kmax, and pachymetry readings were obtained from the scans and analyzed using paired t-tests and Bland-Altman plots. Results There were minimal differences in clinical settings between Pentacam and Galilei for K1, K2, Kmax (>0.75 D), and pachymetry values (>15 um). However, there was a statistically significant difference found between Kmax and pachymetry, making their interchangeability questionable. Conclusion Pentacam and Galilei demonstrate a good correlation between corneal keratometric values (K1, K2, and Kmax) and pachymetry values in clinical settings, and they should be used interchangeably with caution.