Artículo en Inglés | MEDLINE | ID: mdl-38082610

RESUMEN

Traditionally, the functional development, testing and verification of higher level automation, control and safety functions in extracorporeal circulations systems relies very much on the conduction of animal experiments. This especially applies to functions requiring interaction with the patients physiology. Besides ethical aspects, animal experiments have the disadvantage of higher complexity, limited reproducibility and certain pathological conditions can only be represented with difficulty. An alternative to this can be Hardware-in-the-loop (HIL) simulators.In this paper we present a HIL solution to simulate the hydraulic interaction between the cannula and the connected blood vessel. We derive a hydraulic model for an elastic vessel with particular emphasis on negative transmural pressure. In this case the resistance is mainly determined by collapse phenomena. The next section describes the design of an universal resistance actuator that can simulate vascular resistances in the expected range. Combined in the HIL simulator, the simulation model then generates the setpoint for the actuator while simultaneously receiving the resulting internal states of the hydraulic interface. This creates a truly interactive HIL simulator where the device under test interacts in the same way as with a physiological system.


Asunto(s)
Cánula , Computadores , Humanos , Reproducibilidad de los Resultados , Simulación por Computador , Hemodinámica
2.
J Cardiovasc Transl Res ; 16(4): 842-851, 2023 08.
Artículo en Inglés | MEDLINE | ID: mdl-36662482

RESUMEN

When returning blood to the pulmonary artery (PA), the inflow jet interferes with local hemodynamics. We investigated the consequences for several connection scenarios using transient computational fluid dynamics simulations. The PA was derived from CT data. Three aspects were varied: graft flow rate, anastomosis location, and inflow jet path length from anastomosis site to impingement on the PA wall. Lateral anastomosis locations caused abnormal flow distribution between the left and right PA. The central location provided near-physiological distribution but induced higher wall shear stress (WSS). All effects were most pronounced at high graft flows. A central location is beneficial regarding flow distribution, but the resulting high WSS might promote detachment of local thromboembolisms or influence the autonomic nervous innervation. Lateral locations, depending on jet path length, result in lower WSS at the cost of an unfavorable flow distribution that could promote pulmonary vasculature changes. Case-specific decisions and further research are necessary.


Asunto(s)
Hidrodinámica , Arteria Pulmonar , Arteria Pulmonar/cirugía , Hemodinámica , Modelos Cardiovasculares , Estrés Mecánico , Simulación por Computador
3.
Artif Organs ; 47(4): 695-704, 2023 Apr.
Artículo en Inglés | MEDLINE | ID: mdl-36420613

RESUMEN

BACKGROUND: Extracorporeal membrane oxygenation (ECMO) became an accepted therapy for the treatment of severe acute respiratory distress syndrome and chronic obstructive pulmonary disease. However, ECMO systems are still prone to thrombus formation and decrease of gas exchange over time. Therefore, it is necessary to conduct qualified studies to identify parameters for optimization of ECMO systems, and especially the oxygenator. However, commercially marketed oxygenators are not always appropriate and available for certain research use cases. Therefore, we aimed to design an oxygenator, which is suitable for various test conditions such as blood tests, numerical simulation, and membrane studies, and can be modified in membrane area size and manufactured in laboratory. METHODS: Main design criteria are a homogeneous blood flow without stagnation zones, low pressure drop, manufacturability in the lab, size variability with one set of housing parts and cost-efficiency. Our newly designed oxygenator was tested comparatively regarding blood cell damage, gas transfer performance and pressure drop to prove the validity of the design in accordance with a commercial device. RESULTS: No statistically significant difference between the tested oxygenators was detected and our new oxygenator demonstrated sufficient hemocompatibility. Furthermore, our variable oxygenator has proven that it can be easily manufactured in the laboratory, allows to use various membrane fiber configurations and can be reopened easily and non-destructively for analysis after use, and the original geometry is available for numerical simulations. CONCLUSION: Therefore, we consider this newly developed device as a valuable tool for basic experimental and numerical research on the optimization of oxygenators.


Asunto(s)
Oxigenación por Membrana Extracorpórea , Enfermedad Pulmonar Obstructiva Crónica , Síndrome de Dificultad Respiratoria , Trombosis , Humanos , Oxigenadores , Oxigenación por Membrana Extracorpórea/métodos , Síndrome de Dificultad Respiratoria/diagnóstico , Síndrome de Dificultad Respiratoria/terapia , Diseño de Equipo , Oxigenadores de Membrana
4.
Braz J Anesthesiol ; 72(3): 398-406, 2022.
Artículo en Inglés | MEDLINE | ID: mdl-35644204

RESUMEN

BACKGROUND: The effect of mild changes in CO2 levels to organ perfusion and tissue inflammation are well known, whereas an influence of hypercapnia under general anesthesia on adverse events as nausea and vomiting, or length of hospital stay is barely examined. The goal of our meta-analysis was to identify possibly positive effects of hypercapnia versus normocapnia in general anesthesia in adult patients. METHODS: We conducted a systematic review of parallel-arm randomised controlled trials comparing hypercapnia versus normocapnia in adult patients undergoing general anesthesia. In July 2018 and September 2019 we searched "CENTRAL‿, "MEDLINE‿, and "Embase‿, checked reference lists of all included studies and relevant systematic reviews for additional references to trials. Two review authors independently assessed trials for inclusion, extracted data, and completed a "Risk of bias‿ assessment for all included studies. RESULTS: Our search identified 297 records after abstract screening 30 full-text papers remained for further examination. Ten publications met our inclusion criteria and were used for narrative description of this systematic review. Three studies were eligible for the meta-analysis normocapnia versus hypercapnia with the outcomes: time to extubation and adverse events. On average, time to extubation was significantly reduced in the hypercapnia group with a mean difference 3.78 (95% CI 0.85 to 6.71). No difference was found regarding adverse events. CONCLUSIONS: The findings of our study do not enable us to produce evidence of a positive influence of increased CO2 partial pressure levels during general anesthesia. A well-planned, adequately powered randomized controlled trial would be desirable in the future.


Asunto(s)
Dióxido de Carbono , Hipercapnia , Adulto , Anestesia General/efectos adversos , Humanos
5.
Membranes (Basel) ; 12(2)2022 Jan 22.
Artículo en Inglés | MEDLINE | ID: mdl-35207055

RESUMEN

Extracorporeal membrane oxygenation (ECMO) is an established rescue therapy for patients with chronic respiratory failure waiting for lung transplantation (LTx). The therapy inherent immobilization may result in fatigue, consecutive deteriorated prognosis, and even lost eligibility for transplantation. We conducted a feasibility study on a novel system designed for the deployment of a portable ECMO device, enabling the physical exercise of awake patients prior to LTx. The system comprises a novel oxygenator with a directly connected blood pump, a double-lumen cannula, gas blender and supply, as well as control and energy management. In vitro experiments included tests regarding performance, efficiency, and blood damage. A reduced system was tested in vivo for feasibility using a novel large animal model. Six anesthetized pigs were first positioned in supine position, followed by a 45° angle, simulating an upright position of the patients. We monitored performance and vital parameters. All in vitro experiments showed good performance for the respective subsystems and the integrated system. The acute in vivo trials of 8 h duration confirmed the results. The novel portable ECMO-system enables adequate oxygenation and decarboxylation sufficient for, e.g., the physical exercise of designated LTx-recipients. These results are promising and suggest further preclinical studies on safety and efficacy to facilitate translation into clinical application.

6.
Anaesthesist ; 71(5): 377-383, 2022 05.
Artículo en Alemán | MEDLINE | ID: mdl-35147751

RESUMEN

The current S2k guidelines on treatment of thermal injuries in adults are summarized in this article from the perspective of anesthesiology, emergency medicine and intensive care medicine. The guidelines were prepared under the auspices of the German Society for Burn Medicine with the participation of other professional societies and interest groups and were published last year in revised form by the AWMF.


Asunto(s)
Anestesiología , Quemaduras , Adulto , Quemaduras/terapia , Cuidados Críticos , Alemania , Humanos
7.
Minerva Pediatr (Torino) ; 74(2): 116-120, 2022 04.
Artículo en Inglés | MEDLINE | ID: mdl-32274909

RESUMEN

BACKGROUND: Hyperbaric oxygenation therapy (HBOT) is used as emergency treatment for decompression sickness, gas embolism, carbon monoxide intoxication, and necrotizing fasciitis. There is low evidence and little clinical knowledge about the treatment of children with HBOT. METHODS: We sent an internet-based questionnaire to HBO centers in Europe to gain information about their experience with children and HBOT. RESULTS: Out of all HBO-centers who participated in the questionnaire 90% treat children analogue to adults about indication and HBOT protocol. Most treated children had life-threatening indications or the risk of organ loss. The reported rate of side effects was: 6.8% anxiety, 2.4% barotrauma, 0.9% seizure, 0.2% retinopathy and no case of pulmonary barotrauma or oxygen toxicity. CONCLUSIONS: HBO therapy for children is present in European HBO centers. The rate of severe side effects is as low to the rates in adults; apart from this, oxygen-related seizures and anxiety are more frequent. A special focus seems necessary on the psychological management of the children, because anxiety is common depending on the age of the children. Especially for smaller children, an adequate psychological support seems essential. Prospective observational or controlled studies in children seem necessary to create relevant clinical evidence for HBOT and to observe the rate of side-effects.


Asunto(s)
Barotrauma , Oxigenoterapia Hiperbárica , Barotrauma/etiología , Niño , Europa (Continente)/epidemiología , Humanos , Oxigenoterapia Hiperbárica/efectos adversos , Oxigenoterapia Hiperbárica/métodos , Oxígeno , Encuestas y Cuestionarios
8.
Membranes (Basel) ; 12(1)2021 Dec 31.
Artículo en Inglés | MEDLINE | ID: mdl-35054581

RESUMEN

Carbon monoxide (CO) poisoning is the leading cause of poisoning-related deaths globally. The currently available therapy options are normobaric oxygen (NBO) and hyperbaric oxygen (HBO). While NBO lacks in efficacy, HBO is not available in all areas and countries. We present a novel method, extracorporeal hyperoxygenation therapy (EHT), for the treatment of CO poisoning that eliminates the CO by treating blood extracorporeally at elevated oxygen partial pressure. In this study, we proof the principle of the method in vitro using procine blood: Firstly, we investigated the difference in the CO elimination of a hollow fibre membrane oxygenator and a specifically designed batch oxygenator based on the bubble oxygenator principle at elevated pressures (1, 3 bar). Secondly, the batch oxygenator was redesigned and tested for a broader range of pressures (1, 3, 5, 7 bar) and temperatures (23, 30, 37 °C). So far, the shortest measured carboxyhemoglobin half-life in the blood was 21.32 min. In conclusion, EHT has the potential to provide an easily available and effective method for the treatment of CO poisoning.

9.
Handchir Mikrochir Plast Chir ; 52(6): 497-504, 2020 Dec.
Artículo en Alemán | MEDLINE | ID: mdl-32968995

RESUMEN

INTRODUCTION: More than half a million patients suffer from minor burns in Germany per year. In 2018, almost 2000 patients needed intensive care for their burn injuries. Despite high standards of burn care, mortality remains high. Burn injuries may lead to long-term sequelae. In order to provide up-to-date burn care, guidelines are available online with public access. METHODS AND RESULTS: This overview presents a summary of the German AWMF guideline for the treatment of thermal injuries in adults (https://www.awmf.org/leitlinien/detail/ll/044-001.html). Experts of eleven different medical organisations and specialties have contributed to this S2k guideline with their expertise. The focus of the article is on acute burn wound assessment, the indication for specialised care in burn centres, the management of the burn wound at the trauma scene and in hospitals as well as scar management and rehabilitation. CONCLUSION: This overview reports on the consensus-based treatment of acute burn wounds in adults in Germany. The article is intended to guide doctors and professional caretakers to perform state-of-the-art burn care. The current guideline aims to improve burn outcome.


Asunto(s)
Quemaduras , Adulto , Unidades de Quemados , Quemaduras/terapia , Alemania , Humanos
10.
Artif Organs ; 42(4): 410-419, 2018 Apr.
Artículo en Inglés | MEDLINE | ID: mdl-29436001

RESUMEN

Extracorporeal membrane oxygenation (ECMO) is mainly used for the therapy of acute respiratory distress syndrome and chronic obstructive lung disease. In the last years, the development of these systems underwent huge steps in optimization, but there are still problems with thrombus formation, clogging, and thus insufficient gas exchange. One idea of ECMO optimization is a pulsatile blood flow through the oxygenator, but this is still a controversy discussion. Analyzing available publications, it was not possible to identify a general statement about the effect of pulsatile blood flow on the gas exchange performance. The variety of parameters and circuit components have such a high influence on the outcome that a direct comparison of the studies is difficult. For this reason, we performed a structured study to evaluate the effects of pulsatile blood flow on the gas exchange performance of oxygenator. In in vitro tests according to DIN EN ISO 7199, we tested a small oxygenator (0.25 m2 exchange surface, polymethylpentene fibers, 33 mL priming volume) with constant and pulsatile blood flow in comparison. Therefore, we varied the mean blood flow from 250 to 1200 mL/min, the amplitude of 0, 20, and 50%, and the frequency of 30, 60, and 90 bpm. The results demonstrate that the gas transfer for pulsatile and constant blood flow was similar (oxygen: 36-64 mLO2 /LBlood ; carbon dioxide 35-80 mLCO2 /LBlood ) for the same mean blood flow ranges. Over all, the results and analyses showed a statistically nonsignificant difference between pulsatile and nonpulsatile flow. Consequently, we conclude that the implementation of pulsatile blood flow has only a small to no effect on the gas exchange performance in an oxygenator. As the results were obtained using an oxygenator with a coiled fiber bundle, the test must be verified for a stacked fiber oxygenator.


Asunto(s)
Oxigenación por Membrana Extracorpórea/métodos , Modelos Cardiovasculares , Oxigenadores de Membrana/efectos adversos , Flujo Pulsátil , Oxigenación por Membrana Extracorpórea/instrumentación , Humanos , Enfermedad Pulmonar Obstructiva Crónica/terapia , Síndrome de Dificultad Respiratoria/terapia
11.
Biomed Tech (Berl) ; 62(2): 213-223, 2017 Apr 01.
Artículo en Inglés | MEDLINE | ID: mdl-28306515

RESUMEN

This paper presents a decentralized safety concept for networked intensive care setups, for which a decentralized network of sensors and actuators is realized by embedded microcontroller nodes. It is evaluated for up to eleven medical devices in a setup for automated acute respiratory distress syndrome (ARDS) therapy. In this contribution we highlight a blood pump supervision as exemplary safety measure, which allows a reliable bubble detection in an extracorporeal blood circulation. The approach is validated with data of animal experiments including 35 bubbles with a size between 0.05 and 0.3 ml. All 18 bubbles with a size down to 0.15 ml are successfully detected. By using hidden Markov models (HMMs) as statistical method the number of necessary sensors can be reduced by two pressure sensors.


Asunto(s)
Alarmas Clínicas , Cuidados Críticos/métodos , Circulación Extracorporea/métodos , Monitoreo Fisiológico/métodos , Síndrome de Dificultad Respiratoria/diagnóstico , Síndrome de Dificultad Respiratoria/terapia , Terapia Asistida por Computador/métodos , Animales , Retroalimentación Fisiológica , Humanos , Monitoreo Fisiológico/instrumentación , Seguridad del Paciente , Reconocimiento de Normas Patrones Automatizadas/métodos , Resultado del Tratamiento
13.
Biomed Tech (Berl) ; 62(2): 199-212, 2017 Apr 01.
Artículo en Inglés | MEDLINE | ID: mdl-28121615

RESUMEN

A new concept is presented for cooperative automation of mechanical ventilation and extracorporeal membrane oxygenation (ECMO) therapy for treatment of acute respiratory distress syndrome (ARDS). While mechanical ventilation is continuously optimized to promote lung protection, extracorporeal gas transfer rates are simultaneously adjusted to control oxygen supply and carbon dioxide removal using a robust patient-in-the-loop control system. In addition, the cooperative therapy management uses higher-level algorithms to adjust both therapeutic approaches. The controller synthesis is derived based on the introduced objectives, the experimental setup and the uncertain models. Finally, the autonomous ARDS therapy system capabilities are demonstrated and discussed based on in vivo data from animal experiments.


Asunto(s)
Oxigenación por Membrana Extracorpórea/métodos , Pulmón/fisiopatología , Modelos Biológicos , Respiración Artificial/métodos , Síndrome de Dificultad Respiratoria/fisiopatología , Síndrome de Dificultad Respiratoria/terapia , Mecánica Respiratoria , Animales , Terapia Combinada/instrumentación , Terapia Combinada/métodos , Simulación por Computador , Diseño de Equipo , Análisis de Falla de Equipo , Oxigenación por Membrana Extracorpórea/instrumentación , Retroalimentación Fisiológica , Humanos , Respiración Artificial/instrumentación , Síndrome de Dificultad Respiratoria/diagnóstico , Integración de Sistemas , Terapia Asistida por Computador/instrumentación , Terapia Asistida por Computador/métodos , Resultado del Tratamiento
14.
Eur J Cardiothorac Surg ; 51(3): 449-456, 2017 03 01.
Artículo en Inglés | MEDLINE | ID: mdl-27806995

RESUMEN

Objectives: For patients with a severe acute respiratory distress syndrome (ARDS), extracorporeal membrane oxygenation (ECMO) represents a life-saving measure. Frequently, patients with severe ARDS also show signs of severe sepsis. As blood contact with the membrane oxygenator surface leads to adverse effects due to insufficient biocompatibility partly caused by activation of platelets, coagulation factors and leucocytes, we hypothesized that these adverse effects would be amplified if septic blood in a preactivated state came into contact with the membrane oxygenator. Methods: In a previously established in vitro 12-h ECMO test system (mock loop), we used septic or healthy domestic pig blood to analyse coagulation and inflammatory parameters. Sepsis was induced by a caecal ligation and puncture model in pigs. Results: At the beginning of the mock loop experiments, the septic blood showed significantly increased thrombin-antithrombin complexes (76.9 vs 27.7 µg/l), D-dimers (1.2 vs 0.3 mg/l) and fibrinogen concentration (1.8 vs 1.5 g/l), as well as elevated extrinsic coagulation activity (shorter EXTEM-CT: 44.2 vs 57 s) and higher lactate (3.4 vs 1.5 mmol/l) and cytokine levels (interleukin-6: 827 vs 31 pg/ml) when compared with the blood from healthy animals. Despite the preactivated status of the septic blood, no further increase of coagulation activity, inflammatory response or increased oxygenator resistance was observed in comparison to the control experiments. Conclusion: Septic porcine blood was not further activated due to the contact with an oxygenator, and no increased clot formation or biocompatibility problems were observed.


Asunto(s)
Coagulación Sanguínea/fisiología , Oxigenación por Membrana Extracorpórea/métodos , Sepsis/sangre , Animales , Materiales Biocompatibles , Recolección de Muestras de Sangre/métodos , Citocinas/sangre , Modelos Animales de Enfermedad , Hemodinámica/fisiología , Mediadores de Inflamación/metabolismo , Oxigenadores de Membrana , Activación Plaquetaria , Sepsis/diagnóstico , Sus scrofa
15.
J Cell Mol Med ; 21(1): 35-45, 2017 01.
Artículo en Inglés | MEDLINE | ID: mdl-27605340

RESUMEN

D-dopachrome tautomerase (D-DT/MIF-2) is a member of the macrophage migration inhibitory factor (MIF) cytokine superfamily, and a close structural homolog of MIF. MIF and D-DT have been reported to be involved in obesity, but there is little known about the regulation of D-DT in adipose tissue inflammation and wound healing. Subcutaneous adipose tissue was collected from 54 healthy donors and 28 donors with acutely inflamed wounds undergoing wound debridement. In addition, epididymal fat pads of mice were injected with lipopolysaccharide to study receptor expression and cell migration in vivo. D-DT protein levels and mRNA expression were significantly decreased in subcutaneous adipose tissue adjacent to acutely inflamed wounds. D-DT improved fibroblast viability and increased proliferation in vitro. While D-DT alone did not have a significant effect on in vitro fibroblast wound healing, simultaneous addition of neutralizing MIF antibody resulted in a significant improvement of fibroblast wound healing. Interestingly, expression of the MIF and D-DT receptor CD74 was down-regulated while the MIF receptors CXCR2 and CXCR4 were up-regulated primarily on macrophages indicating that the MIF-CXCR2/4 axis may promote recruitment of inflammatory cells into adipose tissue. Our results describe a reciprocal role of D-DT to MIF in inflamed adipose tissue, and indicate that D-DT may be beneficial in wound repair by improving fibroblast survival and proliferation.


Asunto(s)
Tejido Adiposo/metabolismo , Inflamación/metabolismo , Oxidorreductasas Intramoleculares/metabolismo , Cicatrización de Heridas/fisiología , Tejido Adiposo/patología , Animales , Antígenos de Diferenciación de Linfocitos B/metabolismo , Movimiento Celular/fisiología , Proliferación Celular/fisiología , Regulación hacia Abajo/fisiología , Fibroblastos/metabolismo , Fibroblastos/patología , Antígenos de Histocompatibilidad Clase II/metabolismo , Humanos , Inflamación/patología , Macrófagos/metabolismo , Masculino , Ratones , Ratones Endogámicos C57BL , Obesidad/metabolismo , Obesidad/patología , Receptores CXCR4/metabolismo , Receptores de Interleucina-8B/metabolismo , Regulación hacia Arriba/fisiología
16.
Artículo en Alemán | MEDLINE | ID: mdl-27022695

RESUMEN

Skin and soft tissue infections may progress rapidly and take a fatal ending unless not treated in time. A 44-year old male patient without any pre-existing conditions got hospitalized with a bursitis ofthe right olecranon and unspecific general symptoms. Within a short period of time he became critically ill due this seemingly harmless infection. We describe our approach leading to the right diagnoses and the treatment of this unexpected progress.


Asunto(s)
Bursitis/diagnóstico por imagen , Terapia Combinada/métodos , Choque Séptico/diagnóstico , Choque Séptico/microbiología , Choque Séptico/terapia , Infecciones de los Tejidos Blandos/diagnóstico por imagen , Adulto , Animales , Bursitis/microbiología , Bursitis/prevención & control , Cuidados Críticos/métodos , Diagnóstico Diferencial , Progresión de la Enfermedad , Articulación del Codo , Fiebre de Origen Desconocido/diagnóstico , Fiebre de Origen Desconocido/prevención & control , Humanos , Masculino , Infecciones de los Tejidos Blandos/microbiología , Infecciones de los Tejidos Blandos/prevención & control , Resultado del Tratamiento
17.
BMC Anesthesiol ; 15: 158, 2015 Oct 30.
Artículo en Inglés | MEDLINE | ID: mdl-26518485

RESUMEN

BACKGROUND: In this observational study near infrared spectroscopy (NIRS) was evaluated as a non-invasive monitor of impaired tissue oxygenation (StO2) after cardiac surgery. StO2, cardiac output, mixed venous oxygen saturation and mean arterial pressure were compared with lactate clearance as established measure for sufficient tissue perfusion and oxygen metabolism. METHODS: Forty patients after cardiac surgery (24 aortocoronary bypass grafting, 5 heart valve, 3 ascending aorta and 8 combined procedures) were monitored until postoperative day 1 with NIRS of the thenar muscle (InSpectra™ StO2-monitor, Hutchinson Technology), a pulmonary-artery catheter and intermittent blood gas analyses for the assessment of lactate clearance. RESULTS: StO2 was reduced 4 h after surgery (75 ± 6 %), but recovered at day 1 (84 ± 5 %), while lactate concentration remained increased. Using uni- and multivariate regression analysis, minimum StO2 (r = 0.46, p <0.01) and cardiac index (r = 0.40, p <0.05) correlated with lactate clearance at day 1, while minimum mixed venous saturation and mean arterial pressure did not. In a receiver-operating characteristics (ROC) analysis, minimum StO2 (with a threshold of 75 %) predicted a lactate clearance <10 % at day 1 with an area under the ROC-curve of 0.83, a sensitivity of 78 % and a specificity of 88 %. In the subgroup with StO2 <75 %, troponin and creatine kinase MB were significantly increased at day 1. CONCLUSIONS: StO2 below 75 % in the first hours after surgery was a better early indicator of persistent impaired lactate clearance at day 1 than cardiac index, mixed venous oxygen saturation or mean arterial pressure.


Asunto(s)
Procedimientos Quirúrgicos Cardíacos/métodos , Ácido Láctico/metabolismo , Oxígeno/metabolismo , Espectroscopía Infrarroja Corta/métodos , Anciano , Anciano de 80 o más Años , Presión Arterial/fisiología , Análisis de los Gases de la Sangre , Gasto Cardíaco/fisiología , Cateterismo de Swan-Ganz/métodos , Femenino , Humanos , Masculino , Persona de Mediana Edad , Análisis Multivariante , Estudios Prospectivos , Análisis de Regresión , Sensibilidad y Especificidad , Factores de Tiempo
18.
Biomed Tech (Berl) ; 59(2): 135-45, 2014 Apr 01.
Artículo en Inglés | MEDLINE | ID: mdl-24523303

RESUMEN

The ultimate objective of university research and development projects is usually to create knowledge, but also to successfully transfer results to industry for subsequent marketing. We hypothesized that the university technology transfer requires efficient measures to improve this important step. Besides good scientific practice, foresighted and industry-specific adapted documentation of research processes in terms of a quality management system might improve the technology transfer. In order to bridge the gap between research institute and cooperating industry, a model project has been accompanied by a project specific amount of quality management. However, such a system had to remain manageable and must not constrain the researchers' creativity. Moreover, topics and research team are strongly interdisciplinary, which entails difficulties regarding communication because of different perspectives and terminology. In parallel to the technical work of the model project, an adaptable quality management system with a quality manual, defined procedures, and forms and documents accompanying the research, development and validation was implemented. After process acquisition and analysis the appropriate amount of management for the model project was identified by a self-developed rating system considering project characteristics like size, innovation, stakeholders, interdisciplinarity, etc. Employees were trained according to their needs. The management was supported and the technical documentation was optimized. Finally, the quality management system has been transferred successfully to further projects.


Asunto(s)
Equipos y Suministros/normas , Guías como Asunto , Investigación/organización & administración , Transferencia de Tecnología , Gestión de la Calidad Total/organización & administración , Investigación Biomédica Traslacional/organización & administración , Universidades/organización & administración , Alemania
19.
Diving Hyperb Med ; 44(4): 241-2, 2014 Dec.
Artículo en Inglés | MEDLINE | ID: mdl-25596838

RESUMEN

INTRODUCTION: Lead is a toxic element which is known to accumulate in the body. Nevertheless, it is very widely used as a diving weight. METHODS: Blood samples were taken from 20 recreational scuba divers to assess blood lead concentrations. RESULTS: The last dive before blood sampling was an average of 4.8 weeks previously (range 1-18 weeks). All the samples were within the normal background range, the highest lead concentration being 44.8 µg∙L⁻¹ with an average concentration of 26.5 µg∙L⁻¹ (range 11.7-44.8 µg∙L⁻¹). CONCLUSIONS: The results show no elevated blood lead concentrations in this group of divers compared to background levels. However, owing to the small number of divers studied and the variable, often long interval between the last dive and blood sampling, the results cannot be generalized.


Asunto(s)
Buceo , Plomo/sangre , Adulto , Buceo/estadística & datos numéricos , Femenino , Humanos , Masculino , Persona de Mediana Edad , Proyectos Piloto , Factores de Tiempo
20.
Biomed Tech (Berl) ; 59(2): 125-33, 2014 Apr 01.
Artículo en Inglés | MEDLINE | ID: mdl-24327525

RESUMEN

Extracorporeal life support (ECLS) is a well-established technique for the treatment of different cardiac and pulmonary diseases, e.g., congenital heart disease and acute respiratory distress syndrome. Additionally, severely ill patients who cannot be weaned from the heart-lung machine directly after surgery have to be put on ECLS for further therapy. Although both systems include identical components, a seamless transition is not possible yet. The adaption of the circuit to the patients' size and demand is limited owing to the components available. The project I³-Assist aims at a novel concept for extracorporeal circulation. To better match the patient's therapeutic demand of support, an individual number of one-size oxygenators and heat exchangers will be combined. A seamless transition between cardiopulmonary bypass and ECLS will be possible as well as the exchange of components during therapy to enhance circuit maintenance throughout long-term support. Until today, a novel oxygenator and heat exchanger along with a simplified manufacturing protocol have been established. The first layouts of the unit to allow the spill- and bubble-free connection and disconnection of modules as well as improved cannulas and a rotational pump are investigated using computational fluid dynamics. Tests were performed according to current guidelines in vitro and in vivo. The test results show the feasibility and potential of the concept.


Asunto(s)
Circulación Extracorporea/instrumentación , Máquina Corazón-Pulmón , Terapia Asistida por Computador/instrumentación , Interfaz Usuario-Computador , Diseño Asistido por Computadora , Diseño de Equipo , Análisis de Falla de Equipo , Circulación Extracorporea/métodos , Estudios de Factibilidad , Humanos , Integración de Sistemas , Terapia Asistida por Computador/métodos