1.
Rev Sci Instrum ; 92(3): 034708, 2021 Mar 01.
Artículo en Inglés | MEDLINE | ID: mdl-33820089

RESUMEN

We present a cryogenic microwave noise source with a characteristic impedance of 50 Ω, which can be installed in a coaxial line of a cryostat. The bath temperature of the noise source is continuously variable between 0.1 K and 5 K without causing significant back-action heating on the sample space. As a proof-of-concept experiment, we perform Y-factor measurements of an amplifier cascade that includes a traveling wave parametric amplifier and a commercial high electron mobility transistor amplifier. We observe system noise temperatures as low as 680-200 +20 mK at 5.7 GHz corresponding to 1.5-0.7 +0.1 excess photons. The system we present has immediate applications in the validation of solid-state qubit readout lines.