1.
Anal Biochem ; 672: 115159, 2023 07 01.
Artículo en Inglés | MEDLINE | ID: mdl-37072098

RESUMEN

Objective To develop a kit for detecting human epidermal growth factor receptor 2 (HER-2) in the human body. Methods The HER-2 kit was evaluated based on an automated magnetic particle chemiluminescence platform. The kit was developed using the double antibody sandwich-complexation method. Results The kit showed a linear range of 0.01-800 ng/mL, with a linear R2 of >0.999. The limit of the blank was 0.0039 ng/mL, and the precision at 1.00 ng/mL was 9.4%. The recovery rate at 10.00 ng/mL was 97.81-101.81%. The negative serum reference range was 0-8.23 ng/mL. Conclusions The kit had a wide linear range, high accuracy, good precision, and high sensitivity, indicating that it has good application prospects.


Asunto(s)
Juego de Reactivos para Diagnóstico , Receptor ErbB-2 , Humanos , Anticuerpos , Inmunoensayo/métodos , Magnetismo , Receptor ErbB-2/sangre