1.
Polymers (Basel) ; 16(5)2024 Feb 28.
Artículo en Inglés | MEDLINE | ID: mdl-38475334

RESUMEN

In this work, the development of a novel method for the detection of mercury (II) ions in wastewater using a mercury ion-imprinted polymer (IIP) combined with a quartz crystal microbalance (QCM) is described. The IIP was successfully synthesized via the polymerization of a of a novel fluorescein- and 2-aminophenol-functionalized methacrylic acid monomer, which was noted to have high binding affinity to mercury (II) ions. This polymer was subsequently coated on a QCM chip to create an IIP-QCM sensor. This sensor was established to have high selectivity and good sensitivity to mercury (II) ions, and had a limit of detection (LOD) of 14.17 ppb, a limit of quantification (LOQ) of 42.94 ppb, a signal-to-noise ratio (S/N) of 4.29, good repeatability, and a working range of 42.94 ppb to 2 ppm. The sensor was also able to analyze tap water and wastewater samples. The IIP-QCM is, therefore, promising as a highly selective, cost-effective, and rapid mercury ion sensor for applications involving the detection of mercury in wastewater.