1.
Int J Lab Hematol ; 43(6): 1593-1598, 2021 Dec.
Artículo en Inglés | MEDLINE | ID: mdl-34252265

RESUMEN

BACKGROUND: The Sysmex CN-6500 is a new haemostasis analyser with an integrated immunoassay module that performs chemiluminescence enzyme assay (CLEIA) in addition to coagulation, turbidimetric, chromogenic and platelet aggregation tests. AIMS: To evaluate the analytical performance of the CN-6500 against the predicate device (Sysmex HISCL-800) for soluble thrombomodulin (TM), thrombin-antithrombin (TAT), tissue plasminogen activator/plasminogen activator inhibitor 1 complex (tPAI-C) and plasmin α2 plasmin inhibitor complex (PIC) assays. METHODS: Imprecision was assessed by testing two levels of quality control plasmas 10 times on 5 separate days. Comparability was studied in 230 plasmas from normal donors (n = 30), patients with suspected disseminated intravascular coagulation (DIC, n = 100), sepsis (n = 20) or liver disease (n = 20), lipaemic (n = 20), haemolysed (n = 20) and icteric samples (n = 20). Limit of detection, limit of quantitation and linearity were determined by testing serial dilutions of normal plasma. Sample carryover was assessed by testing samples with high and low normal levels of the analytes concerned. RESULTS: The CN-6500 performed 21 CLEIA tests per hour, while simultaneously performing coagulation tests. Acceptable between-run imprecision was obtained using commercial controls with normal and high activity for each analyte (%CV <4%), for all four assays. Excellent linearity was observed (slope 0.89-1.03; r2 >0.99) across the measurement range. The lower limits of detection and quantitation were as follows: TM <0.3/0.6 TU/ml, TAT >0.1/<0.2 ng/ml, PIC <0.004/<0.008 µg/ml and tPAI-C < 0.01/<0.1 ng/ml, respectively. All four assays showed excellent correlation between analysers and were unaffected by haemolysis, icterus or lipaemia. No carryover was observed. CONCLUSIONS: Our data demonstrate that the performance of the CLEIA assays on the CN-6500 is comparable to that of a stand-alone immunoassay analyser.


Asunto(s)
Pruebas de Coagulación Sanguínea/normas , Técnicas para Inmunoenzimas/métodos , Técnicas para Inmunoenzimas/normas , Mediciones Luminiscentes/métodos , Mediciones Luminiscentes/normas , Automatización de Laboratorios , Coagulación Sanguínea , Pruebas de Coagulación Sanguínea/instrumentación , Pruebas de Coagulación Sanguínea/métodos , Humanos , Técnicas para Inmunoenzimas/instrumentación , Mediciones Luminiscentes/instrumentación , Reproducibilidad de los Resultados , Sensibilidad y Especificidad
2.
Int J Lab Hematol ; 42(5): 643-649, 2020 Oct.
Artículo en Inglés | MEDLINE | ID: mdl-32885901

RESUMEN

BACKGROUND: The CN-6000 (Sysmex Corp.) is a new haemostasis analyser with blood coagulation, amidolytic, immuno-turbidometric and light transmission aggregometry (LTA) capabilities. Transmitted light is monitored at multiple wavelengths (340, 405, 575, 660, 800 nm), from an LED light source. AIMS: To evaluate the performance of the CN-6000 against a predicate device. METHODS: The CN-6000 was evaluated against the CS-5100 (Sysmex) for 14 different tests, using 880 samples from normal subjects, anticoagulated patients, critically ill patients, plasmas with high or low fibrinogen content or abnormal levels of interfering substances. Between-day assay imprecision was assessed using commercial QC materials (n = 10 replicates on each of 5 days). RESULTS: Acceptable levels of imprecision were obtained for all assays. Agreement between the two analysers was excellent for all assays. Throughput was 35% higher using the CN-6000 (337 vs 250 tests per hour for PT, aPTT and fibrinogen). The CN-6000 also demonstrated improved clot detection in plasmas with high levels of interfering substances as demonstrated by a 29% reduction in "vote-outs" due to low light transmission (24 vs 34). CONCLUSIONS: The CN-6000 demonstrated excellent comparability with the predicate instrument and acceptable levels of imprecision in all assays. Improvements in throughput and clot detection in the presence of interfering substances were also shown.


Asunto(s)
Pruebas de Coagulación Sanguínea/instrumentación , Pruebas de Coagulación Sanguínea/métodos , Coagulación Sanguínea , Pruebas de Coagulación Sanguínea/normas , Humanos , Reproducibilidad de los Resultados , Sensibilidad y Especificidad
3.
Blood Adv ; 1(16): 1254-1258, 2017 Jul 11.
Artículo en Inglés | MEDLINE | ID: mdl-29296765

RESUMEN

Finding an inherited complement abnormality in HSCT-associated TMA provides a rationale for the use of a complement inhibitor.Alternative complement inhibitors such as Coversin should be considered in patients who are resistant to eculizumab.

4.
Lancet Haematol ; 3(9): e426-36, 2016 Sep.
Artículo en Inglés | MEDLINE | ID: mdl-27570089

RESUMEN

BACKGROUND: Rivaroxaban is established for the treatment and secondary prevention of venous thromboembolism, but whether it is useful in patients with antiphospholipid syndrome is uncertain. METHODS: This randomised, controlled, open-label, phase 2/3, non-inferiority trial, done in two UK hospitals, included patients with antiphospholipid syndrome who were taking warfarin for previous venous thromboembolism, with a target international normalised ratio of 2·5. Patients were randomly assigned 1:1 to continue with warfarin or receive 20 mg oral rivaroxaban daily. Randomisation was done centrally, stratified by centre and patient type (with vs without systemic lupus erythematosus). The primary outcome was percentage change in endogenous thrombin potential (ETP) from randomisation to day 42, with non-inferiority set at less than 20% difference from warfarin in mean percentage change. Analysis was by modified intention to treat. Other thrombin generation parameters, thrombosis, and bleeding were also assessed. Treatment effect was measured as the ratio of rivaroxaban to warfarin for thrombin generation. This trial is registered with the ISRCTN registry, number ISRCTN68222801. FINDINGS: Of 116 patients randomised between June 5, 2013, and Nov 11, 2014, 54 who received rivaroxaban and 56 who received warfarin were assessed. At day 42, ETP was higher in the rivaroxaban than in the warfarin group (geometric mean 1086 nmol/L per min, 95% CI 957-1233 vs 548, 484-621, treatment effect 2·0, 95% CI 1·7-2·4, p<0·0001). Peak thrombin generation was lower in the rivaroxaban group (56 nmol/L, 95% CI 47-66 vs 86 nmol/L, 72-102, treatment effect 0·6, 95% CI 0·5-0·8, p=0·0006). No thrombosis or major bleeding were seen. Serious adverse events occurred in four patients in each group. INTERPRETATION: ETP for rivaroxaban did not reach the non-inferiority threshold, but as there was no increase in thrombotic risk compared with standard-intensity warfarin, this drug could be an effective and safe alternative in patients with antiphospholipid syndrome and previous venous thromboembolism. FUNDING: Arthritis Research UK, Comprehensive Clinical Trials Unit at UCL, LUPUS UK, Bayer, National Institute for Health Research Biomedical Research Centre.


Asunto(s)
Síndrome Antifosfolípido/tratamiento farmacológico , Lupus Eritematoso Sistémico/tratamiento farmacológico , Rivaroxabán/uso terapéutico , Trombosis/tratamiento farmacológico , Warfarina/uso terapéutico , Anticoagulantes/uso terapéutico , Síndrome Antifosfolípido/complicaciones , Síndrome Antifosfolípido/epidemiología , Estudios de Equivalencia como Asunto , Inhibidores del Factor Xa/uso terapéutico , Femenino , Estudios de Seguimiento , Humanos , Lupus Eritematoso Sistémico/complicaciones , Lupus Eritematoso Sistémico/epidemiología , Masculino , Persona de Mediana Edad , Prevalencia , Tasa de Supervivencia , Trombosis/complicaciones , Trombosis/epidemiología , Resultado del Tratamiento , Reino Unido/epidemiología
5.
ERJ Open Res ; 2(4)2016 Oct.
Artículo en Inglés | MEDLINE | ID: mdl-28053974

RESUMEN

Rise in sputum complement (C3a, C5a) levels during COPD exacerbation is associated with recovery time http://ow.ly/ZaPj303xxPf.

7.
Clin Lab Med ; 35(1): 123-31, 2015 Mar.
Artículo en Inglés | MEDLINE | ID: mdl-25676376

RESUMEN

Many modern automated cell counters in high-volume clinical hematology laboratories use new, improved technologies for routine platelet analysis. The latest progress includes the use of state-of-the art information technology, specific fluorescent dyes, and monoclonal antibodies to obtain more reliable platelet counts. This information allows the accurate and precise enumeration of platelets even in thrombocytopenic patients and the reporting of novel platelet parameters. In the near future, digital image analysis may permit even better platelet analysis.


Asunto(s)
Recuento de Plaquetas/métodos , Automatización , Historia del Siglo XX , Historia del Siglo XXI , Recuento de Plaquetas/historia , Recuento de Plaquetas/instrumentación , Recuento de Plaquetas/tendencias
8.
J Clin Pathol ; 68(2): 161-5, 2015 Feb.
Artículo en Inglés | MEDLINE | ID: mdl-25425659

RESUMEN

BACKGROUND: Historically, haematology analyser flags for abnormal white blood cells (WBCs) show good sensitivity but lower specificity, causing unnecessary blood film reviews. While the WBC differential channel on Sysmex XE and XN instruments reports a combined flag for blasts/abnormal lymphocytes, the new white precursor cell channel (WPC) on the XN series has been introduced to separate this into a specific flag for either cell type or, if no abnormality, remove the flag entirely. AIMS: To compare the efficiency of abnormal WBC flags from the XN WPC to our existing analyser and determine whether WPC can reduce false positive flags and blood films required. METHODS: Abnormal WBC flags from the Sysmex XE-5000 and XN-1000 were compared to manual differential and blood film morphology on 300 K2EDTA samples from infants and children. RESULTS: The XN WPC flag for blasts was more sensitive and specific than flags indicating blasts on the XE-5000, with a reduction in false positives from 64% (XE) to 36% (XN). Overall efficiency of the WPC flag for abnormal lymphocytes was 94% vs 79% on the XE. WPC reduced false positive flags for blasts and abnormal lymphocytes on neonatal samples by 50%. Automatic reflex analysis by WPC correctly removed a false positive flag from the white cell differential channel on 46% of samples. Total abnormal WBC flags from XN WPC were less (73) than the XE-5000 (92). CONCLUSIONS: XN WPC demonstrated superior efficiency of abnormal WBC flags on paediatric samples, compared to the XE-5000, with greater sensitivity and specificity of flagging, reducing blood films for review.


Asunto(s)
Citometría de Flujo/instrumentación , Hospitales Pediátricos , Recuento de Leucocitos/instrumentación , Linfocitos/patología , Adolescente , Algoritmos , Automatización de Laboratorios , Niño , Preescolar , Diseño de Equipo , Reacciones Falso Positivas , Humanos , Lactante , Recién Nacido , Londres , Valor Predictivo de las Pruebas , Reproducibilidad de los Resultados , Procesamiento de Señales Asistido por Computador , Centros de Atención Terciaria , Flujo de Trabajo
9.
Thromb Haemost ; 113(1): 13-9, 2015 Jan.
Artículo en Inglés | MEDLINE | ID: mdl-25318976

RESUMEN

Accurate diagnosis of obstetric antiphospholipid syndrome (APS) is a prerequisite for optimal clinical management. The international consensus (revised Sapporo) criteria for obstetric APS do not include low positive anticardiolipin (aCL) and anti ß2 glycoprotein I (aß2GPI) antibodies (< 99th centile) and/or certain clinical criteria such as two unexplained miscarriages, three non-consecutive miscarriages, late pre-eclampsia, placental abruption, late premature birth, or two or more unexplained in vitro fertilisation failures. In this review we examine the available evidence to address the question of whether patients who exhibit non-criteria clinical and/or laboratory manifestations should be included within the spectrum of obstetric APS. Prospective and retrospective cohort studies of women with pregnancy morbidity, particularly recurrent pregnancy loss, suggest that elimination of aCL and/or IgM aß2GPI, or low positive positive aCL or aß2GPI from APS laboratory diagnostic criteria may result in missing the diagnosis in a sizeable number of women who could be regarded to have obstetric APS. Such prospective and retrospective studies also suggest that women with non-criteria obstetric APS may benefit from standard treatment for obstetric APS with low-molecular-weight heparin plus low-dose aspirin, with good pregnancy outcomes. Thus, non-criteria manifestations of obstetric APS may be clinically relevant, and merit investigation of therapeutic approaches. Women with obstetric APS appear to be at a higher risk than other women of pre-eclampsia, placenta-mediated complications and neonatal mortality, and also at increased long-term risk of thrombotic events. The applicability of these observations to outcomes in women with non-criteria obstetric APS remains to be determined.


Asunto(s)
Anticuerpos Antifosfolípidos/sangre , Anticoagulantes/administración & dosificación , Síndrome Antifosfolípido/diagnóstico , Síndrome Antifosfolípido/tratamiento farmacológico , Complicaciones del Embarazo/diagnóstico , Complicaciones del Embarazo/tratamiento farmacológico , Anticoagulantes/efectos adversos , Síndrome Antifosfolípido/sangre , Síndrome Antifosfolípido/mortalidad , Aspirina/administración & dosificación , Biomarcadores/sangre , Quimioterapia Combinada , Femenino , Heparina de Bajo-Peso-Molecular/administración & dosificación , Humanos , Valor Predictivo de las Pruebas , Embarazo , Complicaciones del Embarazo/sangre , Complicaciones del Embarazo/mortalidad , Factores de Riesgo , Resultado del Tratamiento
10.
J Neurol Sci ; 348(1-2): 35-40, 2015 Jan 15.
Artículo en Inglés | MEDLINE | ID: mdl-25498844

RESUMEN

BACKGROUND: Reduced ADAMTS13 activity is seen in thrombotic thrombocytopenic purpura (TTP), and may lead to accumulation of prothrombotic ultra-large von Willebrand factor (ULVWF) multimers in vivo. ADAMTS13 activity and its relationship with VWF antigen (VWF:Ag) levels and platelet function in 'non-TTP related' TIA or ischaemic stroke has not been comprehensively studied. METHODS: In this prospective pilot observational analytical case-control study, ADAMTS13 activity and VWF:Ag levels were quantified in platelet poor plasma in 53 patients in the early phase (≤ 4 weeks) and 34 of these patients in the late phase (≥ 3 months) after TIA or ischaemic stroke on aspirin. Data were compared with those from 22 controls not on aspirin. The impact of ADAMTS13 on platelet function in whole blood was quantified by measuring Collagen-ADP (C-ADP) and Collagen-Epinephrine closure times on a platelet function analyser (PFA-100(®)). RESULTS: Median ADAMTS13 activity was significantly reduced in the early phase (71.96% vs. 95.5%, P <0.01) but not in the late phase after TIA or stroke compared with controls (86.3% vs. 95.5%, P=0.19). There was a significant inverse relationship between ADAMTS13 activity and VWF:Ag levels in the early phase (r=-0.31; P=0.024), but not in the late phase after TIA or stroke (P=0.74). There was a positive correlation between ADAMTS13 activity and C-ADP closure times in early phase patients only, likely mediated via VWF:Ag levels. DISCUSSION: ADAMTS13 activity is reduced and VWF:Ag expression is increased within 4 weeks of TIA or ischaemic stroke onset, and can promote enhanced platelet adhesion and aggregation in response to stimulation with collagen and ADP via VWF-mediated pathways. These data improve our understanding of the dynamic haemostatic and thrombotic profiles of ischaemic cerebrovascular disease (CVD) patients, and are important in view of the potential future role that ADAMTS13 may have to play as an anti-thrombotic agent in CVD.


Asunto(s)
Proteínas ADAM/metabolismo , Plaquetas/fisiología , Ataque Isquémico Transitorio/sangre , Accidente Cerebrovascular/sangre , Factor de von Willebrand/metabolismo , Proteína ADAMTS13 , Anciano , Anciano de 80 o más Años , Femenino , Estudios de Seguimiento , Humanos , Masculino , Persona de Mediana Edad , Proyectos Piloto , Pruebas de Función Plaquetaria , Factores de Tiempo
11.
Expert Rev Hematol ; 7(2): 169-72, 2014 Apr.
Artículo en Inglés | MEDLINE | ID: mdl-24606388

RESUMEN

The current mainstay of treatment of thrombotic antiphospholipid syndrome (APS) is long term warfarin; however, the optimal antithrombotic treatment for APS-related ischaemic stroke or transient ischaemic attacks (TIA) remains uncertain, as does the optimal intensity of anticoagulation. The risk of bleeding with increasing anticoagulant intensity needs to be balanced against the risk of profound permanent disability and death, or irreversible neurological deterioration as a result of recurrent stroke/TIA. Several experts recommend a target INR of 3.5 (range 3.0­4.0) for stroke associated with persistent antiphospholipid antibodies (aPL) which meet International consensus Updated Sapporo (Sydney) classification criteria, with a similar approach in patients with aPL-associated TIA. However, current guidelines recommend a target INR of 2.5 (2.0­3.0) in these patients. Prospective adequately powered clinical studies are required to determine the optimal antithrombotic approach including the potential role of oral direct inhibitors of coagulation for patients with aPL-associated stroke.


Asunto(s)
Anticuerpos Antifosfolípidos/uso terapéutico , Anticoagulantes/uso terapéutico , Accidente Cerebrovascular/tratamiento farmacológico , Administración Oral , Anticuerpos Antifosfolípidos/efectos adversos , Anticuerpos Antifosfolípidos/inmunología , Anticoagulantes/efectos adversos , Síndrome Antifosfolípido/tratamiento farmacológico , Hemorragia/etiología , Humanos , Inhibidores de Proteasas/uso terapéutico , Índice de Severidad de la Enfermedad , Accidente Cerebrovascular/etiología , Warfarina/efectos adversos , Warfarina/uso terapéutico
12.
Br J Haematol ; 164(6): 858-66, 2014 Mar.
Artículo en Inglés | MEDLINE | ID: mdl-24372446

RESUMEN

Complement dysregulation is key in the pathogenesis of atypical Haemolytic Uraemic Syndrome (aHUS), but no clear role for complement has been identified in Thrombotic Thrombocytopenic Purpura (TTP). We aimed to assess complement activation and cytokine response in acute antibody-mediated TTP. Complement C3a and C5a and cytokines (interleukin (IL)-2, IL-4, IL-6, IL-10, tumour necrosis factor, interferon-γ and IL-17a) were measured in 20 acute TTP patients and 49 remission cases. Anti-ADAMTS13 immunoglobulin G (IgG) subtypes were measured in acute patients in order to study the association with complement activation. In acute TTP, median C3a and C5a were significantly elevated compared to remission, C3a 63·9 ng/ml vs. 38·2 ng/ml (P < 0·001) and C5a 16·4 ng/ml vs. 9·29 ng/ml (P < 0·001), respectively. Median IL-6 and IL-10 levels were significantly higher in the acute vs. remission groups, IL-6: 8 pg/ml vs. 2 pg/ml (P = 0·003), IL-10: 6 pg/ml vs. 2 pg/ml (P < 0·001). C3a levels correlated with both anti-ADAMTS13 IgG (rs = 0·604, P = 0·017) and IL-10 (rs = 0·692, P = 0·006). No anti-ADAMTS13 IgG subtype was associated with higher complement activation, but patients with the highest C3a levels had 3 or 4 IgG subtypes present. These results suggest complement anaphylatoxin levels are higher in acute TTP cases than in remission, and the complement response seen acutely may relate to anti-ADAMTS13 IgG antibody and IL-10 levels.


Asunto(s)
Activación de Complemento/inmunología , Citocinas/inmunología , Síndrome Hemolítico-Urémico/inmunología , Púrpura Trombocitopénica Trombótica/inmunología , Proteínas ADAM/inmunología , Adolescente , Adulto , Anciano , Síndrome Hemolítico Urémico Atípico , Estudios de Cohortes , Femenino , Síndrome Hemolítico-Urémico/patología , Humanos , Masculino , Persona de Mediana Edad , Estudios Prospectivos , Púrpura Trombocitopénica Trombótica/patología , Adulto Joven
13.
J Clin Pathol ; 65(11): 1024-30, 2012 Nov.
Artículo en Inglés | MEDLINE | ID: mdl-22851510

RESUMEN

BACKGROUND: The Sysmex XN haematology instrument performs automatic reflex testing, depending on sample results. A nucleated red blood cell (NRBC) count is provided on all samples. The instrument has a smaller footprint (34%) than previous Sysmex XE analysers. METHODS: An evaluation comparing all results to the Sysmex XE-2100 and manual microscopic differential and morphology (n=390) was performed followed by a workflow study of 1000 samples to compare speed of operation and number of blood films reviews required from both systems. RESULTS: The new features on the instrument are: (1) white cell and NRBC channel, all samples include the NRBC count; (2) white cell precursor channel: false positive flags for blasts, abnormal lymphocytes and atypical lymphocytes are reduced significantly without a statistical increase of false negatives; (3) low white cell count mode: suggested setting of <0.5×10(9)/l. An extended count is more precise and provides an accurate differential. Fluorescent platelet count is performed in a dedicated channel. If the red cell or platelet size histograms are abnormal or if the platelet count is low, then a fluorescent platelet count is automatically performed. Good correlation with the XE-2100 and manual differential was found and the improved results compared to the reference flow cytometric analysis for platelet counts, especially below 30×10(9)/l (XE-2100, R(2)=0.500; XN, R(2)=0.875). CONCLUSION: The XN showed reduced sample turnaround time of 10% and reduced number of blood films for examination, 49% less than the XE-2100 without loss of sensitivity with more precise and accurate results on low cell counts.


Asunto(s)
Recuento de Células Sanguíneas/instrumentación , Diseño de Equipo , Enfermedades Hematológicas/diagnóstico , Hematología/instrumentación , Ensayos Analíticos de Alto Rendimiento/instrumentación , Eritroblastos/citología , Hematología/métodos , Humanos , Laboratorios/normas , Factores de Tiempo
16.
Thromb Res ; 130(5): 780-4, 2012 Nov.
Artículo en Inglés | MEDLINE | ID: mdl-22265675

RESUMEN

BACKGROUND: Activated protein C resistance (APCR) within a thrombin generation test (TGT) system is associated with an increased risk of venous thromboembolism (VTE). However, application of the TGT is restricted by the analytical platforms used to monitor thrombin generation. Using a routine coagulation analyser we have developed an automated chromogenic TGT that is sensitive to defects in the protein C pathway. METHOD: The TGT was performed on a TOP500 analyser, in the presence and absence of Protac. The reaction was monitored using a substrate with slow kinetics for thrombin (S-2444). Results were expressed as the area under the curve normalised ratio (AUCnr). Assay results were compared with Coatest APCR (expressed as APC-ratio [CoAPCr]). PATIENTS: Samples were obtained from 35 healthy subjects and 91 patients with previous history of VTE. Of these patients, 19, 17, and 9 had heterozygous factor V Leiden (FVL), antiphospholipid syndrome (APS), and protein C/protein S deficiencies (PC/PS) respectively. RESULTS: Inter-assay imprecision in the presence and absence of Protac were 20% and <5% respectively. There was a significant difference between the AUCnr of normals (median [IQR]: 2.8 [2.4-4.7]) compared to: FVL (1.0 [0.7-1.2]); PC/PS (1.1 [0.9-1.2]); and APS (1.1 [0.8-1.4]); p<0.001 for each comparison. No significant difference was seen between the AUCnr of normals and other VTE patients. The detection rate of AUCnr and CoAPCr were: 100% and 56% for FVL; 88% and 44% for PC/PS; and 64% and 45% for APS respectively. CONCLUSION: The automated TGT exhibited good sensitivity to defects in the protein C pathway.


Asunto(s)
Pruebas de Coagulación Sanguínea/métodos , Proteína C/metabolismo , Trombina/biosíntesis , Adolescente , Adulto , Anciano , Anciano de 80 o más Años , Femenino , Humanos , Masculino , Persona de Mediana Edad , Proteína C/genética , Factores de Riesgo , Trombina/genética , Trombina/metabolismo , Adulto Joven
17.
Thromb Res ; 130(5): 775-9, 2012 Nov.
Artículo en Inglés | MEDLINE | ID: mdl-22245224

RESUMEN

INTRODUCTION: Total hip/knee replacement surgery (THR/TKR respectively) is associated with an increased risk of venous thromboembolism. Dabigatran is recommended as a thromboprophylactic agent post orthopaedic surgery. The aim of this study was to assess the post-operative (Day-1 and Day-2) effect of prophylactic Dabigatran on: the thrombin generation (TG) assay; prothrombin fragment 1.2 (F1.2); thrombin-antithrombin complexes (TAT); D-dimer (D-D); and other coagulation parameters. METHODS AND SAMPLES: Nineteen patients (12 THR, 7 TKR) who received 110 mg dabigatran 4 hours post-operatively, then 220 mg the following day, were recruited. Blood was collected: pre-operatively (Pre-); peri-operatively (Peri-); 19 hours after 110 mg dabigatran (Day-1); and 17 hours after 220 mg dabigatran (Day-2). The TG assay was measured using the Calibrated Automated Thrombogram and a low concentration of tissue factor. Other coagulation parameters measured included activated partial thromboplastin time (APTT), thrombin-time (TT), ecarin-clotting time (ECT) and Hemoclot tests. RESULTS: From Pre- to Peri-, ETP/peak-thrombin, F1.2, TAT and D-D increased significantly. From Peri- to Day-1 and Day-2: TAT reduced progressively; D-D increased; F1.2 did not change significantly; lag-time and time-to-peak prolonged; ETP/Peak-thrombin increased spuriously, due to Dabigatran interfering with the α-2 macroglobulin:thrombin complex in the TG assay. APTT, TT, ECT and Hemoclot increased progressively post-operatively; good correlations were seen between these tests. CONCLUSION: The effect of dabigatran on the TG assay, showed a spurious increase in ETP and Peak-thrombin due to its interference with the TG assay. Dabigatran reduced TAT, but not F1.2, suggesting that thrombin was still being generated after surgery, but was blocked by Dabigatran.


Asunto(s)
Antitrombinas/uso terapéutico , Artroplastia de Reemplazo de Cadera/métodos , Artroplastia de Reemplazo de Rodilla/métodos , Bencimidazoles/uso terapéutico , Coagulación Sanguínea/efectos de los fármacos , Trombina/biosíntesis , beta-Alanina/análogos & derivados , Anciano , Dabigatrán , Femenino , Humanos , Masculino , Persona de Mediana Edad , beta-Alanina/uso terapéutico
18.
Am J Clin Pathol ; 137(1): 65-74, 2012 Jan.
Artículo en Inglés | MEDLINE | ID: mdl-22180479

RESUMEN

A knowledge of the limitations of automated platelet counting is essential for the effective care of thrombocytopenic patients and management of platelet stocks for transfusion. For this study, 29 external quality assessment specimen pools with platelet counts between 5 and 64 × 10(9)/L were distributed to more than 1,100 users of 23 different hematology analyzer models. The same specimen pools were analyzed by the international reference method (IRM) for platelet counting at 3 reference centers. The IRM values were on average lower than the all-methods median values returned by the automated analyzers. The majority (~67%) of the automated analyzer results overestimated the platelet count compared with the IRM, with significant differences in 16.5% of cases. Performance differed between analyzer models. The observed differences may depend in part on the nature of the survey material and analyzer technology, but the findings have implications for the interpretation of platelet counts at levels of clinical decision making.


Asunto(s)
Plaquetas/citología , Recuento de Plaquetas/instrumentación , Recuento de Plaquetas/métodos , Humanos , Internacionalidad , Laboratorios/normas , Recuento de Plaquetas/normas , Garantía de la Calidad de Atención de Salud/normas , Estándares de Referencia , Reproducibilidad de los Resultados , Reino Unido
20.
Am J Clin Pathol ; 136(2): 309-16, 2011 Aug.
Artículo en Inglés | MEDLINE | ID: mdl-21757605

RESUMEN

Hematology analyzers generate suspect flags in the presence of abnormal cells. False-positive rates for flags are high on all analyzers. Sysmex, Kobe, Japan, has developed new software for its XE-5000 with improved algorithms for flagging blast cells, abnormal lymphocytes or lymphoblasts, and atypical lymphocytes. This study evaluated the efficiency of these flags in 1,002 samples. The XE-5000 was compared with the XE-2100 (Sysmex) and microscopic examination of cell morphologic features. On the XE-2100, the blast flag demonstrated 90 false-positives, 13 true-positives, and 3 false-negatives. The values on the XE-5000 were 27 false-positives, 14 true-positives, and 2 false-negatives. The abnormal lymphocyte/lymphoblast flag was assessed with the atypical lymphocyte flag. The XE-2100 showed 114 false-positives, 23 true-positives, and 20 false-negatives, and on the XE-5000, there were 45 false-positives, 22 true-positives, and 21 false-negatives. This more specific flagging reduces the number of films that require manual review.


Asunto(s)
Eficiencia , Neoplasias Hematológicas/diagnóstico , Hematología/instrumentación , Hematología/métodos , Laboratorios/normas , Algoritmos , Reacciones Falso Positivas , Humanos , Recuento de Leucocitos , Sensibilidad y Especificidad , Programas Informáticos