1.
J Appl Clin Med Phys ; 24(9): e14053, 2023 Sep.
Artículo en Inglés | MEDLINE | ID: mdl-37247259

RESUMEN

PURPOSE: The CyberKnife quality assurance (QA) program relies mainly on the use of radiochromic film (RCF). We aimed at evaluating high-resolution arrays of detectors as an alternative to films for CyberKnife machine QA. METHODS: This study will test the SRS Mapcheck (Sun Nuclear, Melbourne, Florida, USA) diode array and its own software, which allows three tests of the CyberKnife QA program to be performed. The first one is a geometrical accuracy test based on the delivery of two orthogonal beams (Automated Quality Assurance, AQA). Besides comparing the constancy and repeatability of both methods, known errors will be introduced to check their sensitivity. The second checks the constancy of the iris collimator field sizes (Iris QA). Changes in the field sizes will be introduced to study the array sensitivity. The last test checks the correct positioning of the multileaf collimator (MLC). It will be tested introducing known systematic displacements to whole banks and to single leaves. RESULTS: The results of the RCF and diode array were equivalent (maximum differences of 0.18 ± 0.14 mm) for the AQA test, showing the array a higher reproducibility. When known errors were introduced, both methods behaved linearly with similar slopes. Regarding Iris QA, the array measurements are highly linear when changes in the field sizes are introduced. Linear regressions show slopes of 0.96-1.17 with r2 above 0.99 in all field sizes. Diode array seems to detect changes of 0.1 mm. In MLC QA, systematic errors of the whole bank of leaves were not detected by the array, while single leaf errors were detected. CONCLUSIONS: The diode array is sensitive and accurate in the AQA and Iris QA tests, which give us the possibility of substituting RCF with a diode array. QA would be performed faster than using the film procedure, obtaining reliable results. Regarding the MLC QA, the inability to detect systematic displacements make it difficult to confidently use the detector.


Asunto(s)
Radioterapia de Intensidad Modulada , Programas Informáticos , Humanos , Reproducibilidad de los Resultados , Garantía de la Calidad de Atención de Salud/métodos , Radioterapia de Intensidad Modulada/métodos , Planificación de la Radioterapia Asistida por Computador , Dosificación Radioterapéutica
2.
J Appl Clin Med Phys ; 24(8): e14006, 2023 Aug.
Artículo en Inglés | MEDLINE | ID: mdl-37097754

RESUMEN

PURPOSE: To study and analyze the effect of the use of a thin guide-wire instead of a Foley catheter for urethral definition in prostate stereotactic body radiation therapy (SBRT) treatments and to compare treatment parameters in both situations. MATERIAL AND METHODS: Thirty-seven prostate SBRT patients were employed in this study. A Foley catheter was employed in nine of them, and a guide-wire was employed in the other 28 patients. For each of the 28 patients in which the guide-wire was employed, a comparison between urethral positions in both situations was performed, allowing for a margin definition of the urethra when a Foley catheter was employed. Displacements of the prostate during treatment were obtained, allowing for an analysis of prostate positions in both situations. Also, different treatment parameters such as the number of treatment interruptions, couch movements performed, and x-rays needed were gathered. RESULTS: Large differences between urethral positions can be found in the anterior-posterior (AP) directions compared to those in the lateral (LAT) direction. Differences are also larger in areas closer to the base of the prostate, where margins applied in the case of using a Foley catheter are 16 mm with a mean displacement of 6 mm in the posterior direction. No differences in the treatment parameters were found during treatment in both situations. The difference found in absolute prostate pitch rotations suggests that the Foley catheter provokes a shift of the prostate position, which does not occur when employing the guide-wire. CONCLUSIONS: Foley catheters shift the urethral position, making them a wrong surrogate of the urethra when no catheters are present. The margins needed to assess uncertainties introduced by the use of a Foley catheter are larger than those usually applied. The use of a Foley catheter did not present any additional difficulty during treatment delivery in terms of images employed or interruptions produced.


Asunto(s)
Neoplasias de la Próstata , Radiocirugia , Masculino , Humanos , Uretra , Próstata , Neoplasias de la Próstata/radioterapia , Neoplasias de la Próstata/cirugía , Cateterismo Urinario
3.
J Med Imaging Radiat Sci ; 51(1): 88-94, 2020 03.
Artículo en Inglés | MEDLINE | ID: mdl-31987765

RESUMEN

PURPOSE: The purpose of the study was to compare the three most common technologies available in digital mammography based in the evaluation of average glandular dose and contrast noise ratio (CNR). METHOD: The average glandular dose was estimated with a direct digital detector (aSe) with a pixel size of 0.85 µm, a photon counter with a pixel size of 50 µm and a computerized radiography (CR) system. A semiconductor detector was used to measure the input kerma to the detector, and the average glandular dose was calculated from the conversion factors dependent on the anode filter combination, half value layer and breast glandularity. RESULTS: The average glandular dose obtained with polymethyl methacrylate and CNR of mean thickness 4.5 cm using direct digital equipment was 1.02 mGy and CNR = 7.4; using the photon counter it was 0.43 mGy and CNR = 4.7 in C100 mode, and 0.64 mGy and CNR = 5.7 in C120 mode; and using the photostimulable CR, the estimated value was 1.65 mGy and CNR = 5.1. CONCLUSIONS: The photon counter offers a lower average glandular dose than the other two devices with adequate image quality (CNR). The CR equipment offers a similar CNR value but delivers a higher dose than the new generations of available mammograms.


Asunto(s)
Neoplasias de la Mama/diagnóstico por imagen , Mamografía/métodos , Radiometría/métodos , Femenino , Humanos , Fotones , Dosis de Radiación
4.
Med Phys ; 46(10): 4622-4630, 2019 Oct.
Artículo en Inglés | MEDLINE | ID: mdl-31370096

RESUMEN

PURPOSE: To propose adaptive setup protocols using Bayesian statistics that facilitate, based on a prediction of coverage probability, making a decision on which patients should follow daily imaging prior to treatment delivery. MATERIALS AND METHODS: The suitability of the treatment margins was assessed combining interfraction variability measurements of the first days of treatment with previous data gathered from our patient population. From this information, we decided if a patient needs an online imaging protocol to perform daily isocenter correction before each treatment fraction. We applied our method to five different datasets. Protocol parameters were selected from each dataset based on coverage probability, the expected imaging workload of the treatment unit, and the accuracy of patient classification. Time trends were assessed and included in the proposed protocols. To validate the accuracy of the protocols, they were applied to a validation dataset of prostate cancer patients. RESULTS: Adaptive setup protocols lead expected population coverage >97% in all datasets analyzed when time trends were considered. The reduction in imaging workload ranged from 40% in lung treatments to 28.5% in prostate treatments. Results of the protocol on the validation dataset were very similar to those previously predicted. CONCLUSIONS: The adaptive setup protocols based on Bayesian statistics presented in this study enable the optimization of imaging workload in the treatment unit ensuring that appropriate dose coverage remains unchanged.


Asunto(s)
Planificación de la Radioterapia Asistida por Computador/métodos , Radioterapia Guiada por Imagen , Teorema de Bayes , Fraccionamiento de la Dosis de Radiación , Humanos , Masculino , Neoplasias de la Próstata/radioterapia