BMC Res Notes ; 11(1): 701, 2018 Oct 04.
Artículo en Inglés | MEDLINE | ID: mdl-30286790

RESUMEN

OBJECTIVE: To synthesize lessons from comparison of results obtained from the immuno-chromatographic SD Bioline testing method and the chemiluminescent micro-particle immunoassay Architect in the diagnosis of syphilis at Livingstone Central hospital laboratory. RESULTS: The specificity and sensitivity of SD Bioline syphilis 3.0 against the chemiluminescent immunoassay using the Architect syphilis Treponema pallidum (TP) was 85.3% and 91.3% respectively with substantial agreement between the two test methods (88%, ĸ  = 0.76; p < 0.0005). We recommend further comprehensive study with a larger sample size and clinical details to ascertain the validity of our findings. We also recommend using a non-treponemal test with the current treponemal tests being used to aid diagnosis.


Asunto(s)
Anticuerpos Antibacterianos/sangre , Cromatografía/normas , Inmunoensayo/normas , Mediciones Luminiscentes/normas , Serodiagnóstico de la Sífilis/normas , Sífilis/diagnóstico , Treponema pallidum/inmunología , Humanos , Sensibilidad y Especificidad , Sífilis/sangre , Serodiagnóstico de la Sífilis/métodos