Am J Infect Control ; 51(2): 194-198, 2023 Feb.
Artículo en Inglés | MEDLINE | ID: mdl-35659562

RESUMEN

BACKGROUND: We analyzed the qualitative fit test results of the filtering facepiece respirators (FFRs) used at our institution to determine their performance and utility. METHODS: We retrospectively analyzed 12,582 qualitative fit testing results for several FFR models among 8,809 health care workers (HCWs). RESULTS: The overall failure rate for HCWs was 15.2%. Nearly one-third (2933/8809, 33.3%) had multiple FFRs fit tested. HCW sex was a statistically significant indicator of fit testing failure (χ2 = 29.9, df = 1, P < .001), with women having a 44% higher likelihood (OR, 1.4; 95% CI: 1.27-1.65) than men. There were statistically significant differences in the failure rate across FFRs (Fig 4, F[5, 12475] = 8.4, P < .001). Fluidshield 46867S had a significantly higher failure rate (49%) than the 3M 1860 (P = .012), 3M 1860S (P < .001), 3M 8210 (P < .001), and Safelife (P < .001) FFRs. CONCLUSIONS: There was a large degree of variation in fit testing results for the FFRs tested. Although we were unable to find an FFR that fit more than 95% of the HCWs successfully, we identified poorly functioning FFRs that can help our institution with decision-making and budgeting for acquisition and stocking appropriate FFRs.


Asunto(s)
Exposición Profesional , Dispositivos de Protección Respiratoria , Masculino , Humanos , Femenino , Respiradores N95 , Estudios Retrospectivos , Ventiladores Mecánicos