1.
Magn Reson Med ; 2024 Apr 22.
Artículo en Inglés | MEDLINE | ID: mdl-38649922

RESUMEN

PURPOSE: We examined magnetic field dependent SNR gains and ability to capture them with multichannel receive arrays for human head imaging in going from 7 T, the most commonly used ultrahigh magnetic field (UHF) platform at the present, to 10.5 T, which represents the emerging new frontier of >10 T in UHFs. METHODS: Electromagnetic (EM) models of 31-channel and 63-channel multichannel arrays built for 10.5 T were developed for 10.5 T and 7 T simulations. A 7 T version of the 63-channel array with an identical coil layout was also built. Array performance was evaluated in the EM model using a phantom mimicking the size and electrical properties of the human head and a digital human head model. Experimental data was obtained at 7 T and 10.5 T with the 63-channel array. Ultimate intrinsic SNR (uiSNR) was calculated for the two field strengths using a voxelized cloud of dipoles enclosing the phantom or the digital human head model as a reference to assess the performance of the two arrays and field depended SNR gains. RESULTS: uiSNR calculations in both the phantom and the digital human head model demonstrated SNR gains at 10.5 T relative to 7 T of 2.6 centrally, ˜2 at the location corresponding to the edge of the brain, ˜1.4 at the periphery. The EM models demonstrated that, centrally, both arrays captured ˜90% of the uiSNR at 7 T, but only ˜65% at 10.5 T, leading only to ˜2-fold gain in array SNR in going from 7 to 10.5 T. This trend was also observed experimentally with the 63-channel array capturing a larger fraction of the uiSNR at 7 T compared to 10.5 T, although the percentage of uiSNR captured were slightly lower at both field strengths compared to EM simulation results. CONCLUSIONS: Major uiSNR gains are predicted for human head imaging in going from 7 T to 10.5 T, ranging from ˜2-fold at locations corresponding to the edge of the brain to 2.6-fold at the center, corresponding to approximately quadratic increase with the magnetic field. Realistic 31- and 63-channel receive arrays, however, approach the central uiSNR at 7 T, but fail to do so at 10.5 T, suggesting that more coils and/or different type of coils will be needed at 10.5 T and higher magnetic fields.

2.
Magn Reson Med ; 89(6): 2204-2216, 2023 06.
Artículo en Inglés | MEDLINE | ID: mdl-36669882

RESUMEN

PURPOSE: To expand on the previously developed B 1 + $$ {\mathrm{B}}_1^{+} $$ -encoding technique, frequency-modulated Rabi-encoded echoes (FREE), to perform accelerated image acquisition by collecting multiple lines of k-space in an echo train. METHODS: FREE uses adiabatic full-passage pulses and a spatially varying RF field to encode unique spatial information without the use of traditional B0 gradients. The original implementation relied on acquiring single lines of k-space, leading to long acquisitions. In this work, an acceleration scheme is presented in which multiple echoes are acquired in a single shot, analogous to conventional fast spin-echo sequences. Theoretical analysis and computer simulations investigated the feasibility of this approach and presented a framework to analyze important imaging parameters of FREE-based sequences. Experimentally, the multi-echo approach was compared with conventional phase-encoded images of the human visual cortex using a simple surface transceiver coil. Finally, different contrasts demonstrated the clinical versatility of the new accelerated sequence. RESULTS: Images were acquired with an acceleration factor of 3.9, compared with the previous implementation of FREE, without exceeding specific absorption rate limits. Different contrasts can easily be acquired without major modifications, including inversion recovery-type images. CONCLUSION: FREE initially illustrated the feasibility of performing slice-selective 2D imaging of the human brain without the need for a B0 gradient along the y-direction. The multi-echo version maintains the advantages that B 1 + $$ {\mathrm{B}}_1^{+} $$ encoding provides but represents an important step toward improving the clinical feasibility of such sequences. Additional acceleration and more advanced reconstruction techniques could further improve the clinical viability of FREE-based techniques.


Asunto(s)
Encéfalo , Imagen por Resonancia Magnética , Humanos , Imagen por Resonancia Magnética/métodos , Encéfalo/diagnóstico por imagen , Cabeza , Simulación por Computador , Aceleración , Procesamiento de Imagen Asistido por Computador/métodos , Fantasmas de Imagen
3.
Magn Reson Med ; 89(5): 2100-2108, 2023 05.
Artículo en Inglés | MEDLINE | ID: mdl-36517956

RESUMEN

PURPOSE: To correct image distortions that result from nonlinear spatial variation in the transmit RF field amplitude ( B 1 + $$ {B}_1^{+} $$ ) when performing spatial encoding with the method called frequency-modulated Rabi encoded echoes (FREE). THEORY AND METHODS: An algorithm developed to correct image distortion resulting from the use of nonlinear static field (B0 ) gradients in standard MRI is adapted herein to correct image distortion arising from a nonlinear B 1 + $$ {B}_1^{+} $$ -gradient field in FREE. From a B 1 + $$ {B}_1^{+} $$ -map, the algorithm performs linear interpolation and intensity scaling to correct the image. The quality of the distortion correction is evaluated in 1.5T images of a grid phantom and human occipital lobe. RESULTS: An expanded theoretical description of FREE revealed the symmetry between this B 1 + $$ {B}_1^{+} $$ -gradient field spatial-encoding and standard B0 -gradient field spatial-encoding. The adapted distortion-correction algorithm substantially reduced image distortions arising in the spatial dimension that was encoded by the nonlinear B 1 + $$ {B}_1^{+} $$ gradient of a circular surface coil. CONCLUSION: Image processing based on straightforward linear interpolation and intensity scaling, as previously applied in conventional MRI, can effectively reduce distortions in FREE images acquired with nonlinear B 1 + $$ {B}_1^{+} $$ -gradient fields.


Asunto(s)
Procesamiento de Imagen Asistido por Computador , Imagen por Resonancia Magnética , Humanos , Imagen por Resonancia Magnética/métodos , Procesamiento de Imagen Asistido por Computador/métodos , Algoritmos , Fantasmas de Imagen
4.
IEEE Antennas Wirel Propag Lett ; 21(9): 1857-1861, 2022 Sep.
Artículo en Inglés | MEDLINE | ID: mdl-37020750

RESUMEN

In this letter, we evaluate antenna designs for ultra-high frequency and field (UHF) human brain magnetic resonance imaging (MRI) at 10.5 tesla (T). Although MRI at such UHF is expected to provide major signal-to-noise gains, the frequency of interest, 447 MHz, presents us with challenges regarding improved B1 + efficiency, image homogeneity, specific absorption rate (SAR), and antenna element decoupling for array configurations. To address these challenges, we propose the use of both monopole and dipole antennas in a novel hybrid configuration, which we refer to as a mono-dipole hybrid antenna (MDH) array. Compared to an 8-channel dipole antenna array of the same dimensions, the 8-channel MDH array showed an improvement in decoupling between adjacent array channels, as well as ~18% higher B1 + and SAR efficiency near the central region of the phantom based on simulation and experiment. However, the performances of the MDH and dipole antenna arrays were overall similar when evaluating a human model in terms of peak B1 + efficiency, 10 g SAR, and SAR efficiency. Finally, the concept of an MDH array showed an advantage in improved decoupling, SAR, and B1 + near the superior region of the brain for human brain imaging.

5.
Sensors (Basel) ; 21(21)2021 Oct 30.
Artículo en Inglés | MEDLINE | ID: mdl-34770558

RESUMEN

For ultra-high field and frequency (UHF) magnetic resonance imaging (MRI), the associated short wavelengths in biological tissues leads to penetration and homogeneity issues at 10.5 tesla (T) and require antenna transmit arrays for efficiently generated 447 MHz B1+ fields (defined as the transmit radiofrequency (RF) magnetic field generated by RF coils). Previously, we evaluated a 16-channel combined loop + dipole antenna (LD) 10.5 T head array. While the LD array configuration did not achieve the desired B1+ efficiency, it showed an improvement of the specific absorption rate (SAR) efficiency compared to the separate 8-channel loop and separate 8-channel dipole antenna arrays at 10.5 T. Here we compare a 16-channel dipole antenna array with a 16-channel LD array of the same dimensions to evaluate B1+ efficiency, 10 g SAR, and SAR efficiency. The 16-channel dipole antenna array achieved a 24% increase in B1+ efficiency in the electromagnetic simulation and MR experiment compared to the LD array, as measured in the central region of a phantom. Based on the simulation results with a human model, we estimate that a 16-channel dipole antenna array for human brain imaging can increase B1+ efficiency by 15% with similar SAR efficiency compared to a 16-channel LD head array.


Asunto(s)
Imagen por Resonancia Magnética , Ondas de Radio , Encéfalo/diagnóstico por imagen , Diseño de Equipo , Humanos , Fantasmas de Imagen
6.
Magn Reson Med ; 86(6): 3292-3303, 2021 12.
Artículo en Inglés | MEDLINE | ID: mdl-34272898

RESUMEN

PURPOSE: Investigating the designs and effects of high dielectric constant (HDC) materials in the shape of a conformal helmet on the enhancement of RF field and reduction of specific absorption rate at 10.5 T for human brain studies. METHODS: A continuous and a segmented four-piece HDC helmet fit to a human head inside an eight-channel fractionated-dipole array were constructed and studied with a phantom and a human head model using computer electromagnetic simulations. The simulated transmit efficiency and receive sensitivity were experimentally validated using a phantom with identical electric properties and helmet-coil configurations of the computer model. The temporal and spatial distributions of displacement currents on the HDC helmets were analyzed. RESULTS: Using the continuous HDC helmet, simulation results in the human head model demonstrated an average transmit efficiency enhancement of 66%. A propagating displacement current was induced on the continuous helmet, leading to an inhomogeneous RF field enhancement in the brain. Using the segmented four-piece helmet design to reduce this effect, an average 55% and 57% enhancement in the transmit efficiency and SNR was achieved in human head, respectively, along with 8% and 28% reductions in average and maximum local specific absorption rate. CONCLUSION: The HDC helmets enhanced the transmit efficiency and SNR of the dipole array coil in the human head at 10.5 T. The segmentation of the helmet to disrupt the continuity of circumscribing displacement currents in the helmet produced a more uniform distribution of the transmit field and lower specific absorption rate in the human head compared with the continuous helmet design.


Asunto(s)
Dispositivos de Protección de la Cabeza , Imagen por Resonancia Magnética , Encéfalo/diagnóstico por imagen , Diseño de Equipo , Humanos , Fantasmas de Imagen , Ondas de Radio
7.
Magn Reson Med ; 86(3): 1759-1772, 2021 09.
Artículo en Inglés | MEDLINE | ID: mdl-33780032

RESUMEN

PURPOSE: Receive array layout, noise mitigation, and B0 field strength are crucial contributors to SNR and parallel-imaging performance. Here, we investigate SNR and parallel-imaging gains at 10.5 T compared with 7 T using 32-channel receive arrays at both fields. METHODS: A self-decoupled 32-channel receive array for human brain imaging at 10.5 T (10.5T-32Rx), consisting of 31 loops and one cloverleaf element, was co-designed and built in tandem with a 16-channel dual-row loop transmitter. Novel receive array design and self-decoupling techniques were implemented. Parallel imaging performance, in terms of SNR and noise amplification (g-factor), of the 10.5T-32Rx was compared with the performance of an industry-standard 32-channel receiver at 7 T (7T-32Rx) through experimental phantom measurements. RESULTS: Compared with the 7T-32Rx, the 10.5T-32Rx provided 1.46 times the central SNR and 2.08 times the peripheral SNR. Minimum inverse g-factor value of the 10.5T-32Rx (min[1/g] = 0.56) was 51% higher than that of the 7T-32Rx (min[1/g] = 0.37) with R = 4 × 4 2D acceleration, resulting in significantly enhanced parallel-imaging performance at 10.5 T compared with 7 T. The g-factor values of 10.5 T-32 Rx were on par with those of a 64-channel receiver at 7 T (eg, 1.8 vs 1.9, respectively, with R = 4 × 4 axial acceleration). CONCLUSION: Experimental measurements demonstrated effective self-decoupling of the receive array as well as substantial gains in SNR and parallel-imaging performance at 10.5 T compared with 7 T.


Asunto(s)
Encéfalo , Imagen por Resonancia Magnética , Aceleración , Encéfalo/diagnóstico por imagen , Diseño de Equipo , Humanos , Fantasmas de Imagen , Relación Señal-Ruido
8.
Magn Reson Med ; 85(6): 3522-3530, 2021 06.
Artículo en Inglés | MEDLINE | ID: mdl-33464649

RESUMEN

PURPOSE: In this work, we investigated how the position of the radiofrequency (RF) shield can affect the signal-to-noise ratio (SNR) of a receive RF coil. Our aim was to obtain physical insight for the design of a 10.5T 32-channel head coil, subject to the constraints on the diameter of the RF shield imposed by the head gradient coil geometry. METHOD: We used full-wave numerical simulations to investigate how the SNR of an RF receive coil depends on the diameter of the RF shield at ultra-high magnetic field (UHF) strengths (≥7T). RESULTS: Our simulations showed that there is an SNR-optimal RF shield size at UHF strength, whereas at low field the SNR monotonically increases with the shield diameter. For a 32-channel head coil at 10.5T, an optimally sized RF shield could act as a cylindrical waveguide and increase the SNR in the brain by 27% compared to moving the shield as far as possible from the coil. Our results also showed that a separate transmit array between the RF shield and the receive array could considerably reduce SNR even if they are decoupled. CONCLUSION: At sufficiently high magnetic field strength, the design of local RF coils should be optimized together with the design of the RF shield to benefit from both near field and resonant modes.


Asunto(s)
Imagen por Resonancia Magnética , Ondas de Radio , Encéfalo/diagnóstico por imagen , Diseño de Equipo , Cabeza , Fantasmas de Imagen , Relación Señal-Ruido
9.
NMR Biomed ; 34(4): e4472, 2021 04.
Artículo en Inglés | MEDLINE | ID: mdl-33511726

RESUMEN

A 32-channel RF coil was developed for brain imaging of anesthetized non-human primates (rhesus macaque) at 10.5 T. The coil is composed of an 8-channel dipole transmit/receive array, close-fitting 16-channel loop receive array headcap, and 8-channel loop receive array lower insert. The transceiver dipole array is composed of eight end-loaded dipole elements self-resonant at the 10.5 T proton Larmor frequency. These dipole elements were arranged on a plastic cylindrical former, which was split into two to allow for convenient animal positioning. Nested into the bottom of the dipole array former is located an 8-channel loop receive array, which contains 5 × 10 cm2 square loops arranged in two rows of four loops. Arranged in a close-fitting plastic headcap is located a high-density 16-channel loop receive array. This array is composed of 14 round loops 37 mm in diameter and 2 partially detachable, irregularly shaped loops that encircle the ears. Imaging experiments were performed on anesthetized non-human primates on a 10.5 T MRI system equipped with body gradients with a 60 cm open bore. The coil enabled submillimeter (0.58 mm isotropic) high-resolution anatomical and functional imaging as well as tractography of fasciculated axonal bundles. The combination of a close-fitting loop receive array and dipole transceiver array allowed for a higher-channel-count receiver and consequent higher signal-to-noise ratio and parallel imaging gains. Parallel imaging performance supports high-resolution functional MRI and diffusion MRI with a factor of three reduction in sampling. The transceive array elements during reception contributed approximately one-quarter of the signal-to-noise ratio in the lower half of the brain, which was farthest from the close-fitting headcap receive array.


Asunto(s)
Cabeza/diagnóstico por imagen , Imagen por Resonancia Magnética/métodos , Animales , Femenino , Macaca mulatta , Relación Señal-Ruido
10.
IEEE Access ; 8: 203555-203563, 2020.
Artículo en Inglés | MEDLINE | ID: mdl-33747679

RESUMEN

We evaluated a 16-channel loop + dipole (LD) transceiver antenna array with improved specific absorption rate (SAR) efficiency for 10.5 Tesla (T) human head imaging apsplications. Three different array designs with equal inner dimensions were considered: an 8-channel dipole antenna, an 8-channel loop, and a 16-channel LD antenna arrays. Signal-to-noise ratio (SNR) and B1 + efficiency (in units of µT per √W) were simulated and measured in 10.5 T magnetic resonance imaging (MRI) experiments. For the safety validation, 10 g SAR and SAR efficiency (defined as the B1 + over √ (peak 10 g SAR)) were calculated through simulation. Finally, high resolution porcine brain images were acquired with the 16-channel LD antenna array, including a fast turbo-spin echo (TSE) sequence incorporating B1 shimming techniques. Both the simulation and experiments demonstrated that the combined 16-channel LD antenna array showed similar B1 + efficiency compared to the 8-channel dipole antenna and the 8-channel loop arrays in a circular polarized (CP) mode. In a central 2 mm × 2 mm region of the phantom, however, the 16-channel LD antenna array showed an improvement in peak 10 g SAR of 27.5 % and 32.5 % over the 8-channel dipole antenna and the 8-channel loop arrays, respectively. We conclude that the proposed 16-channel head LD antenna array design is capable of achieving ~7% higher SAR efficiency at 10.5 T compared to either the 8-channel loop-only or the 8-channel dipole-only antenna arrays of the same dimensions.

11.
Magn Reson Med ; 82(1): 495-509, 2019 07.
Artículo en Inglés | MEDLINE | ID: mdl-30803023

RESUMEN

PURPOSE: Despite the clear synergy between high channel counts in a receive array and magnetic fields ≥ 7 Tesla, to date such systems have been restricted to a maximum of 32 channels. Here, we examine SNR gains at 7 Tesla in unaccelerated and accelerated images with a 64-receive channel (64Rx) RF coil. METHODS: A 64Rx coil was built using circular loops tiled in 2 separable sections of a close-fitting form; custom designed preamplifier boards were integrated into each coil element. A 16-channel transmitter arranged in 2 rows along the z-axis was employed. The performance of the 64Rx array was experimentally compared to that of an industry-standard 32-channel receive (32Rx) array for SNR in unaccelerated images and for noise amplification under parallel imaging. RESULTS: SNR gains were observed in the periphery but not in the center of the brain in unaccelerated imaging compared to the 32Rx coil. With either 1D or 2D undersampling of k-space, or with slice acceleration together with 1D undersampling of k-space, significant reductions in g-factor noise were observed throughout the brain, yielding effective gains in SNR in the entire brain compared to the 32Rx coil. Task-based FMRI data with 12-fold 2D (slice and phase-encode) acceleration yielded excellent quality functional maps with the 64Rx coil but was significantly beyond the capabilities of the 32Rx coil. CONCLUSION: The results confirm the expectations from modeling studies and demonstrate that whole-brain studies with up to 16-fold, 2D acceleration would be feasible with the 64Rx coil.


Asunto(s)
Encéfalo/diagnóstico por imagen , Procesamiento de Imagen Asistido por Computador/instrumentación , Procesamiento de Imagen Asistido por Computador/métodos , Imagen por Resonancia Magnética/métodos , Neuroimagen/métodos , Adulto , Diseño de Equipo , Femenino , Humanos , Masculino , Fantasmas de Imagen , Relación Señal-Ruido , Adulto Joven
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA
...