1.
J Neuroimmunol ; 345: 577288, 2020 08 15.
Artículo en Inglés | MEDLINE | ID: mdl-32544754

RESUMEN

For the diagnosis of anti-MAG polyneuropathy the commercial ELISA manufacturer currently recommends a cut-off of 1000 Bühlmann Titer Units (BTU). We analyzed sera from 80 anti-MAG neuropathy patients and 383 controls (with other neuropathies or healthy controls) to assess the ELISA sensitivity and specificity at different thresholds. A better combination of sensitivity/specificity was found at a threshold >1500 BTU than at >1000 BTU. The best value of specificity was obtained at threshold >7000 BTU. There was a diagnostic grey area between 1500 and 7000 BTU in which the clinical phenotypes as well as electrophysiological studies need to be carefully assessed particularly to differentiate CIDP and anti-MAG neuropathy.


Asunto(s)
Autoanticuerpos/sangre , Glicoproteína Asociada a Mielina/sangre , Polineuropatías/sangre , Estudios de Casos y Controles , Ensayo de Inmunoadsorción Enzimática/métodos , Ensayo de Inmunoadsorción Enzimática/normas , Humanos , Polineuropatías/diagnóstico , Estudios Retrospectivos