2.
Crit Care ; 28(1): 75, 2024 03 14.
Artículo en Inglés | MEDLINE | ID: mdl-38486268

RESUMEN

BACKGROUND: Flow starvation is a type of patient-ventilator asynchrony that occurs when gas delivery does not fully meet the patients' ventilatory demand due to an insufficient airflow and/or a high inspiratory effort, and it is usually identified by visual inspection of airway pressure waveform. Clinical diagnosis is cumbersome and prone to underdiagnosis, being an opportunity for artificial intelligence. Our objective is to develop a supervised artificial intelligence algorithm for identifying airway pressure deformation during square-flow assisted ventilation and patient-triggered breaths. METHODS: Multicenter, observational study. Adult critically ill patients under mechanical ventilation > 24 h on square-flow assisted ventilation were included. As the reference, 5 intensive care experts classified airway pressure deformation severity. Convolutional neural network and recurrent neural network models were trained and evaluated using accuracy, precision, recall and F1 score. In a subgroup of patients with esophageal pressure measurement (ΔPes), we analyzed the association between the intensity of the inspiratory effort and the airway pressure deformation. RESULTS: 6428 breaths from 28 patients were analyzed, 42% were classified as having normal-mild, 23% moderate, and 34% severe airway pressure deformation. The accuracy of recurrent neural network algorithm and convolutional neural network were 87.9% [87.6-88.3], and 86.8% [86.6-87.4], respectively. Double triggering appeared in 8.8% of breaths, always in the presence of severe airway pressure deformation. The subgroup analysis demonstrated that 74.4% of breaths classified as severe airway pressure deformation had a ΔPes > 10 cmH2O and 37.2% a ΔPes > 15 cmH2O. CONCLUSIONS: Recurrent neural network model appears excellent to identify airway pressure deformation due to flow starvation. It could be used as a real-time, 24-h bedside monitoring tool to minimize unrecognized periods of inappropriate patient-ventilator interaction.


Asunto(s)
Aprendizaje Profundo , Respiración Artificial , Adulto , Humanos , Respiración Artificial/métodos , Inteligencia Artificial , Pulmón , Ventiladores Mecánicos
3.
Crit Care ; 27(1): 188, 2023 05 15.
Artículo en Inglés | MEDLINE | ID: mdl-37189173

RESUMEN

BACKGROUND: Intensive Care Unit (ICU) COVID-19 survivors may present long-term cognitive and emotional difficulties after hospital discharge. This study aims to characterize the neuropsychological dysfunction of COVID-19 survivors 12 months after ICU discharge, and to study whether the use of a measure of perceived cognitive deficit allows the detection of objective cognitive impairment. We also explore the relationship between demographic, clinical and emotional factors, and both objective and subjective cognitive deficits. METHODS: Critically ill COVID-19 survivors from two medical ICUs underwent cognitive and emotional assessment one year after discharge. The perception of cognitive deficit and emotional state was screened through self-rated questionnaires (Perceived Deficits Questionnaire, Hospital Anxiety and Depression Scale and Davidson Trauma Scale), and a comprehensive neuropsychological evaluation was carried out. Demographic and clinical data from ICU admission were collected retrospectively. RESULTS: Out of eighty participants included in the final analysis, 31.3% were women, 61.3% received mechanical ventilation and the median age of patients was 60.73 years. Objective cognitive impairment was observed in 30% of COVID-19 survivors. The worst performance was detected in executive functions, processing speed and recognition memory. Almost one in three patients manifested cognitive complaints, and 22.5%, 26.3% and 27.5% reported anxiety, depression and post-traumatic stress disorder (PTSD) symptoms, respectively. No significant differences were found in the perception of cognitive deficit between patients with and without objective cognitive impairment. Gender and PTSD symptomatology were significantly associated with perceived cognitive deficit, and cognitive reserve with objective cognitive impairment. CONCLUSIONS: One-third of COVID-19 survivors suffered objective cognitive impairment with a frontal-subcortical dysfunction 12 months after ICU discharge. Emotional disturbances and perceived cognitive deficits were common. Female gender and PTSD symptoms emerged as predictive factors for perceiving worse cognitive performance. Cognitive reserve emerged as a protective factor for objective cognitive functioning. TRIAL REGISTRATION: ClinicalTrials.gov Identifier: NCT04422444; June 9, 2021.


Asunto(s)
COVID-19 , Trastornos por Estrés Postraumático , Femenino , Humanos , Masculino , Persona de Mediana Edad , Cognición , COVID-19/epidemiología , Demografía , Unidades de Cuidados Intensivos , Alta del Paciente , Estudios Retrospectivos , Trastornos por Estrés Postraumático/complicaciones , Trastornos por Estrés Postraumático/epidemiología , Trastornos por Estrés Postraumático/psicología , Sobrevivientes
4.
Crit Care Med ; 50(7): e619-e629, 2022 07 01.
Artículo en Inglés | MEDLINE | ID: mdl-35120043

RESUMEN

OBJECTIVES: To characterize clusters of double triggering and ineffective inspiratory efforts throughout mechanical ventilation and investigate their associations with mortality and duration of ICU stay and mechanical ventilation. DESIGN: Registry-based, real-world study. BACKGROUND: Asynchronies during invasive mechanical ventilation can occur as isolated events or in clusters and might be related to clinical outcomes. SUBJECTS: Adults requiring mechanical ventilation greater than 24 hours for whom greater than or equal to 70% of ventilator waveforms were available. INTERVENTIONS: We identified clusters of double triggering and ineffective inspiratory efforts and determined their power and duration. We used Fine-Gray's competing risk model to analyze their effects on mortality and generalized linear models to analyze their effects on duration of mechanical ventilation and ICU stay. MEASUREMENTS AND MAIN RESULTS: We analyzed 58,625,796 breaths from 180 patients. All patients had clusters (mean/d, 8.2 [5.4-10.6]; mean power, 54.5 [29.6-111.4]; mean duration, 20.3 min [12.2-34.9 min]). Clusters were less frequent during the first 48 hours (5.5 [2.5-10] vs 7.6 [4.4-9.9] in the remaining period [p = 0.027]). Total number of clusters/d was positively associated with the probability of being discharged alive considering the total period of mechanical ventilation (p = 0.001). Power and duration were similar in the two periods. Power was associated with the probability of being discharged dead (p = 0.03), longer mechanical ventilation (p < 0.001), and longer ICU stay (p = 0.035); cluster duration was associated with longer ICU stay (p = 0.027). CONCLUSIONS: Clusters of double triggering and ineffective inspiratory efforts are common. Although higher numbers of clusters might indicate better chances of survival, clusters with greater power and duration indicate a risk of worse clinical outcomes.


Asunto(s)
Enfermedad Crítica , Ventiladores Mecánicos , Adulto , Enfermedad Crítica/terapia , Humanos , Respiración Artificial
5.
J Pers Med ; 11(12)2021 Nov 29.
Artículo en Inglés | MEDLINE | ID: mdl-34945732

RESUMEN

This study focuses on the application of a non-immersive virtual reality (VR)-based neurocognitive intervention in critically ill patients. Our aim was to assess the feasibility of direct outcome measures to detect the impact of this digital therapy on patients' cognitive and emotional outcomes. Seventy-two mechanically ventilated adult patients were randomly assigned to the "treatment as usual" (TAU, n = 38) or the "early neurocognitive stimulation" (ENRIC, n = 34) groups. All patients received standard intensive care unit (ICU) care. Patients in the ENRIC group also received adjuvant neurocognitive stimulation during the ICU stay. Outcome measures were a full neuropsychological battery and two mental health questionnaires. A total of 42 patients (21 ENRIC) completed assessment one month after ICU discharge, and 24 (10 ENRIC) one year later. At one-month follow-up, ENRIC patients had better working memory scores (p = 0.009, d = 0.363) and showed up to 50% less non-specific anxiety (11.8% vs. 21.1%) and depression (5.9% vs. 10.5%) than TAU patients. A general linear model of repeated measures reported a main effect of group, but not of time or group-time interaction, on working memory, with ENRIC patients outperforming TAU patients (p = 0.008, ηp2 = 0.282). Our results suggest that non-immersive VR-based neurocognitive stimulation may help improve short-term working memory outcomes in survivors of critical illness. Moreover, this advantage could be maintained in the long term. An efficacy trial in a larger sample of participants is feasible and must be conducted.

6.
Sci Rep ; 11(1): 16014, 2021 08 06.
Artículo en Inglés | MEDLINE | ID: mdl-34362950

RESUMEN

The ideal moment to withdraw respiratory supply of patients under Mechanical Ventilation at Intensive Care Units (ICU), is not easy to be determined for clinicians. Although the Spontaneous Breathing Trial (SBT) provides a measure of the patients' readiness, there is still around 15-20% of predictive failure rate. This work is a proof of concept focused on adding new value to the prediction of the weaning outcome. Heart Rate Variability (HRV) and Cardiopulmonary Coupling (CPC) methods are evaluated as new complementary estimates to assess weaning readiness. The CPC is related to how the mechanisms regulating respiration and cardiac pumping are working simultaneously, and it is defined from HRV in combination with respiratory information. Three different techniques are used to estimate the CPC, including Time-Frequency Coherence, Dynamic Mutual Information and Orthogonal Subspace Projections. The cohort study includes 22 patients in pressure support ventilation, ready to undergo the SBT, analysed in the 24 h previous to the SBT. Of these, 13 had a successful weaning and 9 failed the SBT or needed reintubation -being both considered as failed weaning. Results illustrate that traditional variables such as heart rate, respiratory frequency, and the parameters derived from HRV do not differ in patients with successful or failed weaning. Results revealed that HRV parameters can vary considerably depending on the time at which they are measured. This fact could be attributed to circadian rhythms, having a strong influence on HRV values. On the contrary, significant statistical differences are found in the proposed CPC parameters when comparing the values of the two groups, and throughout the whole recordings. In addition, differences are greater at night, probably because patients with failed weaning might be experiencing more respiratory episodes, e.g. apneas during the night, which is directly related to a reduced respiratory sinus arrhythmia. Therefore, results suggest that the traditional measures could be used in combination with the proposed CPC biomarkers to improve weaning readiness.


Asunto(s)
Frecuencia Cardíaca , Unidades de Cuidados Intensivos/estadística & datos numéricos , Respiración Artificial/métodos , Respiración , Desconexión del Ventilador/métodos , Anciano , Femenino , Humanos , Masculino , Persona de Mediana Edad , Estudios Prospectivos
7.
Respir Care ; 66(9): 1389-1397, 2021 Sep.
Artículo en Inglés | MEDLINE | ID: mdl-34230215

RESUMEN

BACKGROUND: This was a pilot study to analyze the effects of tracheostomy on patient-ventilator asynchronies and respiratory system mechanics. Data were extracted from an ongoing prospective, real-world database that stores continuous output from ventilators and bedside monitors. Twenty adult subjects were on mechanical ventilation and were tracheostomized during an ICU stay: 55% were admitted to the ICU for respiratory failure and 35% for neurologic conditions; the median duration of mechanical ventilation before tracheostomy was 12 d; and the median duration of mechanical ventilation was 16 d. METHODS: We compared patient-ventilator asynchronies (the overall asynchrony index and the rates of specific asynchronies) and respiratory system mechanics (respiratory-system compliance and airway resistance) during the 24 h before tracheostomy versus the 24 h after tracheostomy. We analyzed possible differences in these variables among the subjects who underwent surgical versus percutaneous tracheostomy. To compare longitudinal changes in the variables, we used linear mixed-effects models for repeated measures along time in different observation periods. A total of 920 h of mechanical ventilation were analyzed. RESULTS: Respiratory mechanics and asynchronies did not differ significantly between the 24-h periods before and after tracheostomy: compliance of the respiratory system median (IQR) (47.9 [41.3 - 54.6] mL/cm H2O vs 47.6 [40.9 - 54.3] mL/cm H2O; P = .94), airway resistance (9.3 [7.5 - 11.1] cm H2O/L/s vs 7.0 [5.2 - 8.8] cm H2O/L/s; P = .07), asynchrony index (2.0% [1.1 - 3.6%] vs 4.1% [2.3 - 7.6%]; P = .09), ineffective expiratory efforts (0.9% [0.4 - 1.8%] vs 2.2% [1.0 - 4.4%]; P = .08), double cycling (0.5% [0.3 - 1.0%] vs 0.9% [0.5 - 1.9%]; P = .24), and percentage of air trapping (7.6% [4.2 - 13.8%] vs 10.6% [5.9 - 19.2%]; P = .43). No differences in respiratory mechanics or patient-ventilator asynchronies were observed between percutaneous and surgical procedures. CONCLUSIONS: Tracheostomy did not affect patient-ventilator asynchronies or respiratory mechanics within 24 h before and after the procedure.


Asunto(s)
Traqueostomía , Ventiladores Mecánicos , Adulto , Humanos , Pulmón , Proyectos Piloto , Estudios Prospectivos , Respiración Artificial , Mecánica Respiratoria
8.
IEEE Trans Biomed Eng ; 68(3): 1005-1014, 2021 03.
Artículo en Inglés | MEDLINE | ID: mdl-32746073

RESUMEN

Surface electromyography (sEMG) can be used for the evaluation of respiratory muscle activity. Recording sEMG involves the use of surface electrodes in a bipolar configuration. However, electrocardiographic (ECG) interference and electrode orientation represent considerable drawbacks to bipolar acquisition. As an alternative, concentric ring electrodes (CREs) can be used for sEMG acquisition and offer great potential for the evaluation of respiratory muscle activity due to their enhanced spatial resolution and simple placement protocol, which does not depend on muscle fiber orientation. The aim of this work was to analyze the performance of CREs during respiratory sEMG acquisitions. Respiratory muscle sEMG was applied to the diaphragm and sternocleidomastoid muscles using a bipolar and a CRE configuration. Thirty-two subjects underwent four inspiratory load spontaneous breathing tests which was repeated after interchanging the electrode positions. We calculated parameters such as (1) spectral power and (2) median frequency during inspiration, and power ratios of inspiratory sEMG without ECG in relation to (3) basal sEMG without ECG (Rins/noise), (4) basal sEMG with ECG (Rins/cardio) and (5) expiratory sEMG without ECG (Rins/exp). Spectral power, Rins/noise and Rins/cardio increased with the inspiratory load. Significantly higher values (p < 0.05) of Rins/cardio and significantly higher median frequencies were obtained for CREs. Rins/noise and Rins/exp were higher for the bipolar configuration only in diaphragm sEMG recordings, whereas no significant differences were found in the sternocleidomastoid recordings. Our results suggest that the evaluation of respiratory muscle activity by means of sEMG can benefit from the remarkably reduced influence of cardiac activity, the enhanced detection of the shift in frequency content and the axial isotropy of CREs which facilitates its placement.


Asunto(s)
Diafragma , Músculos Respiratorios , Electrocardiografía , Electrodos , Electromiografía , Humanos , Músculo Esquelético
10.
Crit Care ; 24(1): 618, 2020 10 21.
Artículo en Inglés | MEDLINE | ID: mdl-33087171

RESUMEN

BACKGROUND: ICU patients undergoing invasive mechanical ventilation experience cognitive decline associated with their critical illness and its management. The early detection of different cognitive phenotypes might reveal the involvement of diverse pathophysiological mechanisms and help to clarify the role of the precipitating and predisposing factors. Our main objective is to identify cognitive phenotypes in critically ill survivors 1 month after ICU discharge using an unsupervised machine learning method, and to contrast them with the classical approach of cognitive impairment assessment. For descriptive purposes, precipitating and predisposing factors for cognitive impairment were explored. METHODS: A total of 156 mechanically ventilated critically ill patients from two medical/surgical ICUs were prospectively studied. Patients with previous cognitive impairment, neurological or psychiatric diagnosis were excluded. Clinical variables were registered during ICU stay, and 100 patients were cognitively assessed 1 month after ICU discharge. The unsupervised machine learning K-means clustering algorithm was applied to detect cognitive phenotypes. Exploratory analyses were used to study precipitating and predisposing factors for cognitive impairment. RESULTS: K-means testing identified three clusters (K) of patients with different cognitive phenotypes: K1 (n = 13), severe cognitive impairment in speed of processing (92%) and executive function (85%); K2 (n = 33), moderate-to-severe deficits in learning-memory (55%), memory retrieval (67%), speed of processing (36.4%) and executive function (33.3%); and K3 (n = 46), normal cognitive profile in 89% of patients. Using the classical approach, moderate-to-severe cognitive decline was recorded in 47% of patients, while the K-means method accurately classified 85.9%. The descriptive analysis showed significant differences in days (p = 0.016) and doses (p = 0.039) with opioid treatment in K1 vs. K2 and K3. In K2, there were more women, patients were older and had more comorbidities (p = 0.001) than in K1 or K3. Cognitive reserve was significantly (p = 0.001) higher in K3 than in K1 or K2. CONCLUSION: One month after ICU discharge, three groups of patients with different cognitive phenotypes were identified through an unsupervised machine learning method. This novel approach improved the classical classification of cognitive impairment in ICU survivors. In the exploratory analysis, gender, age and the level of cognitive reserve emerged as relevant predisposing factors for cognitive impairment in ICU patients. TRIAL REGISTRATION: ClinicalTrials.gov Identifier:NCT02390024; March 17,2015.


Asunto(s)
Cognición/fisiología , Unidades de Cuidados Intensivos/estadística & datos numéricos , Fenotipo , Factores de Tiempo , Anciano , Estudios de Cohortes , Femenino , Humanos , Unidades de Cuidados Intensivos/organización & administración , Tiempo de Internación/estadística & datos numéricos , Masculino , Persona de Mediana Edad , Estudios Prospectivos , Respiración Artificial
12.
Sci Rep ; 10(1): 13911, 2020 08 17.
Artículo en Inglés | MEDLINE | ID: mdl-32807815

RESUMEN

Patient-ventilator asynchronies can be detected by close monitoring of ventilator screens by clinicians or through automated algorithms. However, detecting complex patient-ventilator interactions (CP-VI), consisting of changes in the respiratory rate and/or clusters of asynchronies, is a challenge. Sample Entropy (SE) of airway flow (SE-Flow) and airway pressure (SE-Paw) waveforms obtained from 27 critically ill patients was used to develop and validate an automated algorithm for detecting CP-VI. The algorithm's performance was compared versus the gold standard (the ventilator's waveform recordings for CP-VI were scored visually by three experts; Fleiss' kappa = 0.90 (0.87-0.93)). A repeated holdout cross-validation procedure using the Matthews correlation coefficient (MCC) as a measure of effectiveness was used for optimization of different combinations of SE settings (embedding dimension, m, and tolerance value, r), derived SE features (mean and maximum values), and the thresholds of change (Th) from patient's own baseline SE value. The most accurate results were obtained using the maximum values of SE-Flow (m = 2, r = 0.2, Th = 25%) and SE-Paw (m = 4, r = 0.2, Th = 30%) which report MCCs of 0.85 (0.78-0.86) and 0.78 (0.78-0.85), and accuracies of 0.93 (0.89-0.93) and 0.89 (0.89-0.93), respectively. This approach promises an improvement in the accurate detection of CP-VI, and future study of their clinical implications.


Asunto(s)
Entropía , Respiración Artificial/instrumentación , Respiración Artificial/métodos , Ventiladores Mecánicos , APACHE , Anciano , Automatización , Femenino , Humanos , Masculino , Persona de Mediana Edad , Reología
13.
Respir Care ; 65(6): 847-869, 2020 Jun.
Artículo en Inglés | MEDLINE | ID: mdl-32457175

RESUMEN

Mechanical ventilation in critically ill patients must effectively unload inspiratory muscles and provide safe ventilation (ie, enhancing gas exchange, protect the lungs and the diaphragm). To do that, the ventilator should be in synchrony with patient's respiratory rhythm. The complexity of such interplay leads to several concerning issues that clinicians should be able to recognize. Asynchrony between the patient and the ventilator may induce several deleterious effects that require a proper physiological understanding to recognize and manage them. Different tools have been developed and proposed beyond the careful analysis of the ventilator waveforms to help clinicians in the decision-making process. Moreover, appropriate handling of asynchrony requires clinical skills, physiological knowledge, and suitable medication management. New technologies and devices are changing our daily practice, from automated real-time recognition of asynchronies and their distribution during mechanical ventilation, to smart alarms and artificial intelligence algorithms based on physiological big data and personalized medicine. Our goal as clinicians is to provide care of patients based on the most accurate and current knowledge, and to incorporate new technological methods to facilitate and improve the care of the critically ill.


Asunto(s)
Respiración Artificial/métodos , Mecánica Respiratoria/fisiología , Ventiladores Mecánicos , Enfermedad Crítica , Humanos , Ventilación Pulmonar/fisiología
14.
Stem Cell Reports ; 13(1): 207-220, 2019 07 09.
Artículo en Inglés | MEDLINE | ID: mdl-31231023

RESUMEN

In vitro surrogate models of human cardiac tissue hold great promise in disease modeling, cardiotoxicity testing, and future applications in regenerative medicine. However, the generation of engineered human cardiac constructs with tissue-like functionality is currently thwarted by difficulties in achieving efficient maturation at the cellular and/or tissular level. Here, we report on the design and implementation of a platform for the production of engineered cardiac macrotissues from human pluripotent stem cells (PSCs), which we term "CardioSlice." PSC-derived cardiomyocytes, together with human fibroblasts, are seeded into large 3D porous scaffolds and cultured using a parallelized perfusion bioreactor with custom-made culture chambers. Continuous electrical stimulation for 2 weeks promotes cardiomyocyte alignment and synchronization, and the emergence of cardiac tissue-like properties. These include electrocardiogram-like signals that can be readily measured on the surface of CardioSlice constructs, and a response to proarrhythmic drugs that is predictive of their effect in human patients.


Asunto(s)
Miocardio , Ingeniería de Tejidos , Andamios del Tejido , Técnicas de Cultivo Celular por Lotes , Fenómenos Biomecánicos , Reactores Biológicos , Diferenciación Celular , Células Cultivadas , Fenómenos Electrofisiológicos , Humanos , Células Madre Pluripotentes Inducidas/citología , Células Madre Pluripotentes Inducidas/metabolismo , Miocardio/citología , Miocardio/metabolismo
15.
IEEE J Biomed Health Inform ; 23(5): 1964-1971, 2019 09.
Artículo en Inglés | MEDLINE | ID: mdl-30530375

RESUMEN

The use of wearable devices in clinical routines could reduce healthcare costs and improve the quality of assessment in patients with chronic respiratory diseases. The purpose of this study is to evaluate the capacity of a Shimmer3 wearable device to extract reliable cardiorespiratory parameters from surface diaphragm electromyography (EMGdi). Twenty healthy volunteers underwent an incremental load respiratory test whilst EMGdi was recorded with a Shimmer3 wearable device (EMGdiW). Simultaneously, a second EMGdi (EMGdiL), inspiratory mouth pressure (Pmouth) and lead-I electrocardiogram (ECG) were recorded via a standard wired laboratory acquisition system. Different cardiorespiratory parameters were extracted from both EMGdiW and EMGdiL signals: heart rate, respiratory rate, respiratory muscle activity, and mean frequency of EMGdi signals. Alongside these, similar parameters were also extracted from reference signals (Pmouth and ECG). High correlations were found between the data extracted from the EMGdiW and the reference signal data: heart rate (R = 0.947), respiratory rate (R = 0.940), respiratory muscle activity (R = 0.877), and mean frequency (R = 0.895). Moreover, similar increments in EMGdiW and EMGdiL activity were observed when Pmouth was raised, enabling the study of respiratory muscle activation. In summary, the Shimmer3 device is a promising and cost-effective solution for the ambulatory monitoring of respiratory muscle function in chronic respiratory diseases.


Asunto(s)
Electromiografía/instrumentación , Frecuencia Cardíaca/fisiología , Frecuencia Respiratoria/fisiología , Procesamiento de Señales Asistido por Computador/instrumentación , Dispositivos Electrónicos Vestibles , Adulto , Electrocardiografía/instrumentación , Electrocardiografía/métodos , Electromiografía/métodos , Diseño de Equipo , Femenino , Humanos , Masculino , Adulto Joven
16.
Annu Int Conf IEEE Eng Med Biol Soc ; 2019: 2344-2347, 2019 Jul.
Artículo en Inglés | MEDLINE | ID: mdl-31946370

RESUMEN

The electrical activity of the diaphragm measured by surface electromyography (sEMGdi) provides indirect information on neural respiratory drive. Moreover, it allows evaluating the ventilatory pattern from the onset and offset (ntoff) estimation of the neural inspiratory time. sEMGdi amplitude variation was quantified using the fixed sample entropy (fSampEn), a less sensitive method to the interference from cardiac activity. The detection of the ntoff is controversial, since it is located in an intermediate point between the maximum value and the cessation of sEMGdi inspiratory activity, evaluated by the fSampEn. In this work ntoff detection has been analyzed using thresholds between 40% and 100 % of the fSampEn peak. Furthermore, fSampEn was evaluated analyzing the r parameter from 0.05 to 0.6, using a m equal to 1 and a sliding window size equal to 250 ms. The ntoff has been compared to the offset time (toff) obtained from the airflow during a controlled respiratory protocol varying the fractional inspiratory time from 0.54 to 0.18 whilst the respiratory rate was constant at 16 bpm. Results show that the optimal threshold values were between 66.0 % to 77.0 % of the fSampEn peak value. r values between 0.25 to 0.50 were found suitable to be used with the fSampEn.


Asunto(s)
Diafragma , Respiración Artificial , Electromiografía , Entropía , Frecuencia Respiratoria
17.
Entropy (Basel) ; 21(3)2019 Mar 07.
Artículo en Inglés | MEDLINE | ID: mdl-33266973

RESUMEN

To optimize long-term nocturnal non-invasive ventilation in patients with chronic obstructive pulmonary disease, surface diaphragm electromyography (EMGdi) might be helpful to detect patient-ventilator asynchrony. However, visual analysis is labor-intensive and EMGdi is heavily corrupted by electrocardiographic (ECG) activity. Therefore, we developed an automatic method to detect inspiratory onset from EMGdi envelope using fixed sample entropy (fSE) and a dynamic threshold based on kernel density estimation (KDE). Moreover, we combined fSE with adaptive filtering techniques to reduce ECG interference and improve onset detection. The performance of EMGdi envelopes extracted by applying fSE and fSE with adaptive filtering was compared to the root mean square (RMS)-based envelope provided by the EMG acquisition device. Automatic onset detection accuracy, using these three envelopes, was evaluated through the root mean square error (RMSE) between the automatic and mean visual onsets (made by two observers). The fSE-based method provided lower RMSE, which was reduced from 298 ms to 264 ms when combined with adaptive filtering, compared to 301 ms provided by the RMS-based method. The RMSE was negatively correlated with the proposed EMGdi quality indices. Following further validation, fSE with KDE, combined with adaptive filtering when dealing with low quality EMGdi, indicates promise for detecting the neural onset of respiratory drive.

18.
Sci Rep ; 8(1): 16921, 2018 11 16.
Artículo en Inglés | MEDLINE | ID: mdl-30446712

RESUMEN

The current gold standard assessment of human inspiratory muscle function involves using invasive measures of transdiaphragmatic pressure (Pdi) or crural diaphragm electromyography (oesEMGdi). Mechanomyography is a non-invasive measure of muscle vibration associated with muscle contraction. Surface electromyogram and mechanomyogram, recorded transcutaneously using sensors placed over the lower intercostal spaces (sEMGlic and sMMGlic respectively), have been proposed to provide non-invasive indices of inspiratory muscle activation, but have not been directly compared to gold standard Pdi and oesEMGdi measures during voluntary respiratory manoeuvres. To validate the non-invasive techniques, the relationships between Pdi and sMMGlic, and between oesEMGdi and sEMGlic were measured simultaneously in 12 healthy subjects during an incremental inspiratory threshold loading protocol. Myographic signals were analysed using fixed sample entropy (fSampEn), which is less influenced by cardiac artefacts than conventional root mean square. Strong correlations were observed between: mean Pdi and mean fSampEn |sMMGlic| (left, 0.76; right, 0.81), the time-integrals of the Pdi and fSampEn |sMMGlic| (left, 0.78; right, 0.83), and mean fSampEn oesEMGdi and mean fSampEn sEMGlic (left, 0.84; right, 0.83). These findings suggest that sMMGlic and sEMGlic could provide useful non-invasive alternatives to Pdi and oesEMGdi for the assessment of inspiratory muscle function in health and disease.


Asunto(s)
Electromiografía , Contracción Muscular , Mecánica Respiratoria , Músculos Respiratorios/fisiología , Adulto , Femenino , Voluntarios Sanos , Humanos , Masculino
19.
Annu Int Conf IEEE Eng Med Biol Soc ; 2018: 3342-3345, 2018 Jul.
Artículo en Inglés | MEDLINE | ID: mdl-30441104

RESUMEN

The relationship between surface diaphragm mechanomyography (sMMGdi), as a noninvasive measure of inspiratory muscle mechanical activation, and crural diaphragm electromyography (oesEMGdi), as the invasive gold standard measure of diaphragm electrical activation, had not previously been examined. To investigate this relationship, oesEMGdi and sMMGdi were measured simultaneously in 6 healthy subjects during an incremental inspiratory threshold loading protocol, and analyzed using fixed sample entropy (fSampEn). A positive curvilinear relationship was observed between mean fSampEn sMMGdi and oesEMGdi (r = 0.67). Accordingly, an increasing electromechanical ratio was also observed with increasing inspiratory load. These findings suggest that sMMGdi could provide useful noninvasive measures of inspiratory muscle mechanical activation.


Asunto(s)
Diafragma , Electromiografía , Entropía , Voluntarios Sanos , Humanos
20.
Annu Int Conf IEEE Eng Med Biol Soc ; 2018: 3350-3353, 2018 Jul.
Artículo en Inglés | MEDLINE | ID: mdl-30441106

RESUMEN

The assessment of respiratory muscle activity by surface electromyography (sEMG) is a promising noninvasive technique for the diagnosis and monitoring of chronic obstructive pulmonary disease. The diaphragm is the most important muscle in breathing, although in forced inspiration other muscles, such as sternocleidomastoid, are activated and contribute to the respiratory process. The measurement of the sEMG in these muscles (sEMGdi and sEMGscm, respectively) by means of two electrodes in conventional bipolar configuration (BEs) is a common practice to evaluate the respiratory muscle activity and allows to indirectly quantify the level of muscular activation. However, the resulting signals are usually contaminated by electrocardiographic (ECG) activity, hindering the assessment of the activity of these muscles. sEMG signals can also be recorded using concentric ring electrodes (CREs). CREs have greater spatial resolution and attenuate distant bioelectrical interferences. In this scenario, the objective of this work has been to evaluate the applicability of CREs for the acquisition of sEMGdi and sEMGscm. For this purpose, both sEMG signals were recorded simultaneously with BEs and CREs in healthy subjects while performing an inspiratory load protocol. To evaluate the effect of the cardiac interference, the ratio between the mean power in inspiratory segments without ECG and the mean power in expiratory segments with ECG (Rcardio) was calculated. Additionally, the ratio between the mean power in inspiratory segments without ECG and the mean power in expiratory segments without ECG (Rinex) was also calculated. The results revealed that the Rcardio and bandwidth is greater in sEMG signals acquired with the CREs, while the Rinex is higher in the signals acquired with BEs. These results suggest that the use of CREs is a recommended alternative for the acquisition of sEMG in muscles with high cardiac interference, such as the diaphragm muscle.


Asunto(s)
Músculos Respiratorios , Diafragma , Electrocardiografía , Electrodos , Electromiografía
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA
...