1.
Biomed Tech (Berl) ; 62(2): 213-223, 2017 Apr 01.
Artículo en Inglés | MEDLINE | ID: mdl-28306515

RESUMEN

This paper presents a decentralized safety concept for networked intensive care setups, for which a decentralized network of sensors and actuators is realized by embedded microcontroller nodes. It is evaluated for up to eleven medical devices in a setup for automated acute respiratory distress syndrome (ARDS) therapy. In this contribution we highlight a blood pump supervision as exemplary safety measure, which allows a reliable bubble detection in an extracorporeal blood circulation. The approach is validated with data of animal experiments including 35 bubbles with a size between 0.05 and 0.3 ml. All 18 bubbles with a size down to 0.15 ml are successfully detected. By using hidden Markov models (HMMs) as statistical method the number of necessary sensors can be reduced by two pressure sensors.


Asunto(s)
Alarmas Clínicas , Cuidados Críticos/métodos , Circulación Extracorporea/métodos , Monitoreo Fisiológico/métodos , Síndrome de Dificultad Respiratoria/diagnóstico , Síndrome de Dificultad Respiratoria/terapia , Terapia Asistida por Computador/métodos , Animales , Retroalimentación Fisiológica , Humanos , Monitoreo Fisiológico/instrumentación , Seguridad del Paciente , Reconocimiento de Normas Patrones Automatizadas/métodos , Resultado del Tratamiento