Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Más filtros











Base de datos
Intervalo de año de publicación
1.
J ISAKOS ; 9(3): 410-414, 2024 Jun.
Artículo en Inglés | MEDLINE | ID: mdl-38266965

RESUMEN

Numerous interference screws of different compositions exist including titanium screws, bioabsorbable screws, and polyetheretherketone (PEEK) screws. PEEK-based implants are frequently used in orthopaedic surgery due to their biocompatibility, similar elastic modulus to cortical bone, and purported negligible risk of osteolysis compared with bioabsorbable screws. In this case report, we present the case of a 48-year-old healthy female who experienced a massive osteolytic reaction in the talus and fibula after 11 weeks following implantation of PEEK-based interference screws during lateral ankle ligament reconstruction. The patient subsequently underwent removal of the PEEK screws and specimens were sent for microbiological and histopathological analysis. The specimens report demonstrated fibrotic tenosynovial soft tissue with patchy chronic inflammation, oedematous reactive changes, and histiocytic reaction, with no evidence of any significant acute inflammation. The patient recovered well and was asymptomatic at 6 months postoperatively. To our knowledge, this is the first case report of a massive osteolytic reaction to PEEK-based interference screws.


Asunto(s)
Benzofenonas , Tornillos Óseos , Peroné , Cetonas , Osteólisis , Polietilenglicoles , Polímeros , Astrágalo , Humanos , Femenino , Persona de Mediana Edad , Osteólisis/etiología , Tornillos Óseos/efectos adversos , Cetonas/efectos adversos , Polietilenglicoles/efectos adversos , Polímeros/efectos adversos , Peroné/cirugía , Astrágalo/cirugía , Materiales Biocompatibles/efectos adversos
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA