1.
J Appl Clin Med Phys ; 24(6): e13919, 2023 Jun.
Artículo en Inglés | MEDLINE | ID: mdl-37278646

RESUMEN

To evaluate the image quality, dosimetric properties, setup reproducibility, and planar cine motion detection of a high-resolution brain coil and integrated stereotactic brain immobilization system that constitute a new brain treatment package (BTP) on a low-field magnetic resonance imaging (MRI) linear accelerator (MR-linac). Image quality of the high-resolution brain coil was evaluated with the 17 cm diameter spherical phantom and the American College of Radiology (ACR) Large MRI Phantom. Patient imaging studies approved by the institutional review board (IRB) assisted in selecting image acquisition parameters. Radiographic and dosimetric evaluation of the high-resolution brain coil and the associated immobilization devices was performed using dose calculations and ion chamber measurements. End-to-end testing was performed simulating a cranial lesion in a phantom. Inter-fraction setup variability and motion detection tests were evaluated on four healthy volunteers. Inter-fraction variability was assessed based on three repeat setups for each volunteer. Motion detection was evaluated using three-plane (axial, coronal, and sagittal) MR-cine imaging sessions, where volunteers were asked to perform a set of specific motions. The images were post-processed and evaluated using an in-house program. Contrast resolution of the high-resolution brain coil is superior to the head/neck and torso coils. The BTP receiver coils have an average HU value of 525 HU. The most significant radiation attenuation (3.14%) of the BTP, occurs through the lateral portion of the overlay board where the high-precision lateral-profile mask clips attach to the overlay. The greatest inter-fraction setup variability occurred in the pitch (average 1.08 degree) and translationally in the superior/inferior direction (average 4.88 mm). Three plane cine imaging with the BTP was able to detect large and small motions. Small voluntary motions, sub-millimeter in magnitude (maximum 0.9 mm), from motion of external limbs were detected. Imaging tests, inter-fraction setup variability, attenuation, and end-to-end measurements were quantified and performed for the BTP. Results demonstrate better contrast resolution and low contrast detectability that allows for better visualization of soft tissue anatomical changes relative to head/neck and torso coil systems.


Asunto(s)
Neoplasias Encefálicas , Humanos , Reproducibilidad de los Resultados , Neoplasias Encefálicas/diagnóstico por imagen , Neoplasias Encefálicas/radioterapia , Encéfalo , Imagenología Tridimensional , Imagen por Resonancia Magnética/métodos , Fantasmas de Imagen
2.
Front Oncol ; 13: 1090582, 2023.
Artículo en Inglés | MEDLINE | ID: mdl-36761944

RESUMEN

Objective: Magnetic resonance imaging (MRI) is a standard imaging modality in intracranial stereotactic radiosurgery (SRS) for defining target volumes. However, wide disparities in MRI resolution exist, which could directly impact accuracy of target delineation. Here, sequences with various MRI resolution were acquired on phantoms to evaluate the effect on volume definition and dosimetric consequence for cranial SRS. Materials/Methods: Four T1-weighted MR sequences with increasing 3D resolution were compared, including two Spin Echo (SE) 2D acquisitions with 5mm and 3mm slice thickness (SE5mm, SE3mm) and two gradient echo 3D acquisitions (TFE, BRAVO). The voxel sizes were 0.4×0.4×5.0, 0.5×0.5×3.0, 0.9×0.9×1.25, and 0.4×0.4×0.5 mm3, respectively. Four phantoms with simulated lesions of different shape and volume (range, 0.53-25.0 cm3) were imaged, resulting in 16 total sets of MRIs. Four radiation oncologists provided contours on individual MR image set. All observer contours were compared with ground truth, defined on CT image according to the absolute dimensions of the target structure, using Dice similarity coefficient (DSC), Hausdorff distance (HD), mean distance-to-agreement (MDA), and the ratio between reconstructed and true volume (Ratiovol ). For dosimetric consequence, SRS plans targeting observer volumes were created. The true Paddick conformity index ( C I p a d d i c k t r u e ), calculated with true target volume, was correlated with quality of observer volume. Results: All measures of observer contours improved as increasingly higher MRI resolution was provided from SE5mm to BRAVO. The improvement in DSC, HD and MDA was statistically significant (p<0.01). Dosimetrically, C I p a d d i c k t r u e   strongly correlated with DSC of the planning observer volume (Pearson's r=0.94, p<0.00001). Conclusions: Significant improvement in target definition and reduced inter-observer variation was observed as the MRI resolution improved, which also improved the quality of SRS plans. Results imply that high resolution 3D MR sequences should be used to minimize potential errors in target definition, and multi-slice 2D sequences should be avoided.

3.
Adv Radiat Oncol ; 6(4): 100663, 2021.
Artículo en Inglés | MEDLINE | ID: mdl-33997481

RESUMEN

PURPOSE: Treatment planning of skull based meningiomas can be difficult due to the irregular shaped target volumes and proximity to critical optic structures. This study evaluated the use of HyperArc (HA) radiosurgery optimization and delivery in conjunction with multicriteria optimization (MCO) to create conformal and efficient treatment plans for conventionally fractionated radiation therapy to difficult base-of-skull (BOS) lesions. METHODS AND MATERIALS: Twelve patients with BOS meningioma were retrospectively planned with HA-specific optimization algorithm, stereotactic normal tissue objective (SRS-NTO), and conventional automatic normal tissue objective to evaluate normal brain sparing (mean dose and V20 Gy). MCO was used on both SRS-NTO and automatic normal tissue objective plans to further decrease organ-at-risk doses and target dose maximum to within clinically acceptable constraints. Delivery efficiency was evaluated based on planned monitor units. RESULTS: The SRS-NTO in HA can be used to improve the mid- and low-dose spread to normal brain tissue in the irradiation of BOS meningiomas. Improvement in normal brain sparing can be seen in larger, more irregular shaped lesions and less so in smaller spherical targets. MCO can be used in conjunction with the SRS-NTO to reduce target dose maximum and dose to organ at risk without sacrificing the gain in normal brain sparing. CONCLUSIONS: HA can be beneficial both in treatment planning by using the SRS-NTO and in delivery efficiency through the decrease in monitor units and automated delivery.

4.
Phys Med Biol ; 62(11): 4346-4360, 2017 06 07.
Artículo en Inglés | MEDLINE | ID: mdl-28072395

RESUMEN

The purpose of this study was to develop metrics to evaluate uncertainties in deformable dose accumulation for patients with non-small cell lung cancer (NSCLC). Initial treatment plans (primary) and cone-beam CT (CBCT) images were retrospectively processed for seven NSCLC patients, who showed significant tumor regression during the course of treatment. Each plan was developed with IMRT for 2 Gy × 33 fractions. A B-spline-based DIR algorithm was used to register weekly CBCT images to a reference image acquired at fraction 21 and the resultant displacement vector fields (DVFs) were then modified using a finite element method (FEM). The doses were calculated on each of these CBCT images and mapped to the reference image using a tri-linear dose interpolation method, based on the B-spline and FEM-generated DVFs. Contours propagated from the planning image were adjusted to the residual tumor and OARs on the reference image to develop a secondary plan. For iso-prescription adaptive plans (relative to initial plans), mean lung dose (MLD) was reduced, on average from 17.3 Gy (initial plan) to 15.2, 14.5 and 14.8 Gy for the plans adapted using the rigid, B-Spline and FEM-based registrations. Similarly, for iso-toxic adaptive plans (considering MLD relative to initial plans) using the rigid, B-Spline and FEM-based registrations, the average doses were 69.9 ± 6.8, 65.7 ± 5.1 and 67.2 ± 5.6 Gy in the initial volume (PTV1), and 81.5 ± 25.8, 77.7 ± 21.6, and 78.9 ± 22.5 Gy in the residual volume (PTV21), respectively. Tumor volume reduction was correlated with dose escalation (for isotoxic plans, correlation coefficient = 0.92), and with MLD reduction (for iso-fractional plans, correlation coefficient = 0.85). For the case of the iso-toxic dose escalation, plans adapted with the B-Spline and FEM DVFs differed from the primary plan adapted with rigid registration by 2.8 ± 1.0 Gy and 1.8 ± 0.9 Gy in PTV1, and the mean difference between doses accumulated using the B-spline and FEM DVF's was 1.1 ± 0.6 Gy. As a dose mapping-induced energy change, energy defect in the tumor volume was 20.8 ± 13.4% and 4.5 ± 2.4% for the B-spline and FEM-based dose accumulations, respectively. The energy defect of the B-Spline-based dose accumulation is significant in the tumor volume and highly correlated to the difference between the B-Spline and FEM-accumulated doses with their correlation coefficient equal to 0.79. Adaptive planning helps escalate target dose and spare normal tissue for patients with NSCLC, but deformable dose accumulation may have a significant loss of energy in regressed tumor volumes when using image intensity-based DIR algorithms. The metric of energy defect is a useful tool for evaluation of adaptive planning accuracy for lung cancer patients.


Asunto(s)
Carcinoma de Pulmón de Células no Pequeñas/radioterapia , Neoplasias Pulmonares/radioterapia , Planificación de la Radioterapia Asistida por Computador/métodos , Algoritmos , Carcinoma de Pulmón de Células no Pequeñas/diagnóstico por imagen , Carcinoma de Pulmón de Células no Pequeñas/patología , Tomografía Computarizada de Haz Cónico , Análisis de Elementos Finitos , Humanos , Neoplasias Pulmonares/diagnóstico por imagen , Neoplasias Pulmonares/patología , Dosificación Radioterapéutica , Estudios Retrospectivos , Carga Tumoral
5.
J Appl Clin Med Phys ; 17(4): 268-284, 2016 07 08.
Artículo en Inglés | MEDLINE | ID: mdl-27455505

RESUMEN

2.5 MV electronic portal imaging, available on Varian TrueBeam machines, was characterized using various phantoms in this study. Its low-contrast detectability, spatial resolution, and contrast-to-noise ratio (CNR) were compared with those of conventional 6 MV and kV planar imaging. Scatter effect in large patient body was simulated by adding solid water slabs along the beam path. The 2.5 MV imaging mode was also evaluated using clinically acquired images from 24 patients for the sites of brain, head and neck, lung, and abdomen. With respect to 6 MV, the 2.5 MV achieved higher contrast and preserved sharpness on bony structures with only half of the imaging dose. The quality of 2.5 MV imaging was comparable to that of kV imaging when the lateral separation of patient was greater than 38 cm, while the kV image quality degraded rapidly as patient separation increased. Based on the results of patient images, 2.5 MV imaging was better for cranial and extracranial SRS than the 6 MV imaging.


Asunto(s)
Neoplasias Encefálicas/diagnóstico por imagen , Tomografía Computarizada de Haz Cónico/métodos , Neoplasias de Cabeza y Cuello/diagnóstico por imagen , Procesamiento de Imagen Asistido por Computador/métodos , Neoplasias Pulmonares/diagnóstico por imagen , Fantasmas de Imagen , Radiocirugia/métodos , Neoplasias Encefálicas/cirugía , Neoplasias de Cabeza y Cuello/cirugía , Humanos , Neoplasias Pulmonares/cirugía , Masculino , Aceleradores de Partículas , Dosis de Radiación , Planificación de la Radioterapia Asistida por Computador/métodos , Radioterapia de Intensidad Modulada/métodos