1.
Aust Endod J ; 2024 May 15.
Artículo en Inglés | MEDLINE | ID: mdl-38747371

RESUMEN

This study assessed the mechanical properties of the Only One File Blue and the Reciproc Blue instruments. A total of 80 new 25 mm reciprocating NiTi instruments (25/.08v) were evaluated for their mechanical performance (n = 40 per group). Cyclic fatigue resistance, torsional fatigue, flexural resistance and buckling resistance tests were conducted. Statistical analysis was employed with a significance level set at 5%. The results indicated no statistically significant differences in resistance to cyclic fatigue neither in the fragment length between the instruments (p > 0.05). However, the Reciproc Blue instrument exhibited greater torque to fracture and a larger rotation angle than the Only One File Blue instrument (p < 0.05). The Only One File Blue instrument demonstrated higher flexibility but lower resistance to buckling compared to the Reciproc Blue instrument (p < 0.05). In conclusion, the tested instruments exhibit differences in mechanical properties, with the Reciproc Blue instrument generally presenting greater advantages than the Only One File Blue.

2.
Aust Endod J ; 2024 May 08.
Artículo en Inglés | MEDLINE | ID: mdl-38715465

RESUMEN

This study assessed canal preparation effects on disinfection and dentin preservation. Thirty mandibular incisors were paired into two experimental groups (n = 10). Following contamination, the initial microbial sample was collected. Instruments 30/0.03 (Group 1) and 30/0.05 (Group 2) were employed and a second sample was obtained. Canals were enlarged using instruments 40/0.03 and 40/0.05, respectively, and a third sample was collected. Final irrigation was performed, and sample S4 obtained. A final scan evaluated volume, surface area, unprepared areas, removed dentin and dentin thickness. Data were analysed using Student t-test, Mann-Whitney, Kruskal-Wallis and Dunn tests. A significant difference was observed between S1 and other time points (p < 0.05). Comparison between groups showed no differences in bacterial loads and in the percentage of microbial reduction (p > 0.05). Group 2 exhibited greater reduction in dentin thickness than group 1 in the mesial aspect of the root (p < 0.05). Instrument 30/0.03 might provide effective disinfection and safety during mandibular incisors canal preparation.

3.
Int Endod J ; 57(5): 601-616, 2024 May.
Artículo en Inglés | MEDLINE | ID: mdl-38376108

RESUMEN

AIM: To compare eight large- and low-tapered heat-treated reciprocating instruments regarding their design, metallurgy, mechanical properties, and irrigation flow through an in silico model. METHODOLOGY: A total of 472 new 25-mm E-Flex Rex (25/.04 and 25/.06), Excalibur (25/.05), Procodile (25/.06), Reciproc Blue R25 (25/.08v), WaveOne Gold Primary (25/.07v), and Univy Sense (25/.04 and 25/.06) instruments were evaluated regarding their design (stereomicroscopy, scanning electron microscopy, and 3D surface scanning), metallurgy (energy-dispersive X-ray spectroscopy and differential scanning calorimetry), and mechanical performance (cyclic fatigue, torsional resistance, cutting ability, bending and buckling resistance). Computational fluid dynamics assessment was also conducted to determine the irrigation flow pattern, apical pressure, and wall shear stress in simulated canal preparations. Kruskal-Wallis and one-way anova post hoc Tukey tests were used for statistical comparisons (α = 5%). RESULTS: Instruments presented variations in blade numbers, helical angles, and tip designs, with all featuring non-active tips, symmetrical blades, and equiatomic nickel-titanium ratios. Cross-sectional designs exhibited an S-shaped geometry, except for WaveOne Gold. Univy 25/.04 and Reciproc Blue displayed the smallest and largest core diameters at D3. Univy 25/.04 and E-Flex Rec 25/.04 demonstrated the longest time to fracture (p < .05). Reciproc Blue and Univy 25/.04 exhibited the highest and lowest torque to fracture, respectively (p < .05). Univy 25/.04 and Reciproc Blue had the highest rotation angles, whilst E-Flex Rec 25/.06 showed the lowest angle (p < .05). The better cutting ability was observed with E-Flex Rec 25/.06, Procodile, Excalibur, and Reciproc Blue (p > .05). Reciproc R25 and E-Flex Rec showed the highest buckling resistance values (p < .05), with WaveOne Gold being the least flexible instrument. The impact of instruments' size and taper on wall shear stress and apical pressure did not follow a distinct pattern, although Univy 25/.04 and E-Flex Rec 25/.06 yielded the highest and lowest values for both parameters, respectively. CONCLUSIONS: Low-tapered reciprocating instruments exhibit increased flexibility, higher time to fracture, and greater angles of rotation, coupled with reduced maximum bending loads and buckling strength compared to large-tapered instruments. Nevertheless, low-tapered systems also exhibit lower maximum torque to fracture and inferior cutting ability, contributing to a narrower apical canal enlargement that may compromise the penetration of irrigants in that region.


Asunto(s)
Instrumentos Dentales , Titanio , Estudios Transversales , Diseño de Equipo , Ensayo de Materiales , Estrés Mecánico , Titanio/química , Preparación del Conducto Radicular , Metalurgia
4.
Aust Endod J ; 2024 Jan 19.
Artículo en Inglés | MEDLINE | ID: mdl-38243635

RESUMEN

This study evaluated the impact of canal irrigation using surfactants associated to NaOCl on the percentage of voids of root canal filling through micro-computed tomographic (micro-CT) analysis. 27 mandibular mesial roots with Vertucci type II canal configuration with isthmus were selected and scanned in a micro-CT device. The specimens were assigned to three groups (n = 9), according to the irrigation solution during instrumentation: 2.5% NaOCl without additives (control), 2.5% NaOCl with 0.1% Benzalkonium Chloride and 2.5% NaOCl with 0.1% Tween 80. After obturation, the specimens were rescanned, and the percentage of voids were calculated. Data were analysed using one-way ANOVA with a significance level of 5%. The percentage of voids were similar among the groups (p > 0.05). None of the specimens showed completely voids-free areas. The addition of surfactants to NaOCl irrigation did not improve the quality of root canal filling of mandibular mesial root canals.

6.
Int Endod J ; 57(2): 178-194, 2024 Feb.
Artículo en Inglés | MEDLINE | ID: mdl-37966374

RESUMEN

AIM: To assess the periapical alveolar bone pattern and the serum levels of proinflammatory cytokines, biochemical markers and metabolites in rats subjected to chronic alcohol and nicotine consumption and induced apical periodontitis. METHODOLOGY: Twenty-eight male Wistar rats were divided into four groups: Control, Alcohol, Nicotine and Alcohol+Nicotine. The alcohol groups were exposed to self-administration of a 25% alcohol solution, while the other groups were given only filtered water. The nicotine groups received daily intraperitoneal injections of a nicotine solution (0.19 µL of nicotine/mL), whereas the other groups received saline solution. Periapical lesions were induced by exposing the pulps of the left mandibular first molars for 28 days. After euthanasia, the mandibles were removed and the percentage bone volume, bone mineral density, trabecular thickness, trabecular separation and trabecular number of the periapical bone were measured using micro-computed tomography images. Serum samples were collected for analysis of proinflammatory cytokines (IL-1ß, IL-4, IL-6 and TNF-α), biochemical and metabolomic analysis. Statistical analysis was performed with a significance level of 5%. Nonparametric data were analysed using the Kruskal-Wallis test followed by Dunn's test, while one-way anova followed by Tukey's test was performed for parametric data. RESULTS: The groups exposed to alcohol or nicotine consumption exhibited an altered bone pattern indicating lower bone density and higher levels of IL-1ß, IL-6 and TNF-α compared to the Control group (p < .05). Significant differences were observed among the groups in the levels of aspartate aminotransferase (AST), alanine aminotransferase (ALT), alkaline phosphatase, cholesterol, triglycerides, urea, creatinine, albumin, uric acid, bilirubin and calcium. Metabolomic analysis revealed significant differences in glycine, phosphocholine, lysine, lactate, valine, pyruvate and lipids (CH2 CH2 CO), n(CH2 ) and n(CH3 ). Most of these parameters were even more altered in the simultaneous consumption of both substances compared to single consumption. CONCLUSION: Alcohol and nicotine chronic consumption altered several metabolic markers, impaired liver and kidney function, increased the production of systemic proinflammatory mediators and harmed the periapical bone microarchitecture in the presence of apical periodontitis. The simultaneous consumption of alcohol and nicotine intensified these detrimental effects.


Asunto(s)
Nicotina , Periodontitis Periapical , Ratas , Masculino , Animales , Ratas Wistar , Nicotina/farmacología , Microtomografía por Rayos X , Interleucina-6 , Factor de Necrosis Tumoral alfa , Etanol , Interleucina-1beta
7.
Aust Endod J ; 50(1): 60-68, 2024 Apr.
Artículo en Inglés | MEDLINE | ID: mdl-37876340

RESUMEN

This study evaluated the impact of long-term storage conditions (medium and time) on the development of experimental dentinal microcracks through micro-computed tomography. Sixty freshly extracted premolars were stored in formalin, water, or dry conditions (n = 20) and scanned after 72 h, 30 days, 6 months and 3 years of extraction. The effect of the storage medium and time on the occurrence of dentinal defects was statistically evaluated. A total of 211 000 images were screened revealing the existence of 11 519 slices with dentinal defects. Dry conditions significantly contributed to the development of new defects in all time points. During the 3-year follow-up period, no new defects were detected in the teeth that were stored in water and, in a single tooth, in the formalin group, after 6 months of storage.


Asunto(s)
Fracturas de los Dientes , Humanos , Dentina/diagnóstico por imagen , Microtomografía por Rayos X , Preparación del Conducto Radicular , Agua , Formaldehído
8.
Restor Dent Endod ; 48(4): e34, 2023 Nov.
Artículo en Inglés | MEDLINE | ID: mdl-38053776

RESUMEN

This systematic review aimed to investigate whether minimally invasive root canal preparation ensures higher fracture resistance compared to conventional root canal preparation in endodontically treated teeth (ETT). A comprehensive search strategy was conducted on the "PubMed, Web of Science, and Scopus" databases, alongside reference and hand searches, with language restrictions applied. Two independent reviews selected pertinent laboratory studies that explored the effect of minimally invasive root canal preparation on fracture resistance, in comparison to larger preparation counterparts. The quality of the studies was assessed, and the risk of bias was categorized as low, moderate, or high. The electronic search yielded a total of 1,767 articles. After applying eligibility criteria, 8 studies were included. Given the low methodological quality of these studies and the large variability of fracture resistance values, the impact of reduced apical size and/or taper on the fracture resistance of the ETT can be considered uncertain. This systematic review could not reveal sufficient evidence regarding the effect of minimally invasive preparation on increasing fracture resistance of ETT, primarily due to the inherent limitations of the studies and the moderate risk of bias.

9.
Restor Dent Endod ; 48(4): e37, 2023 Nov.
Artículo en Inglés | MEDLINE | ID: mdl-38053784

RESUMEN

This study aimed to compare the disinfectant ability of chlorhexidine (CHX) gel and sodium hypochlorite (NaOCl). Systematic searches were conducted from inception until December 8th, 2022 (MEDLINE/PubMed, Cochrane Library, Web of Science, Scopus, Embase, and Grey Literature databases). Only randomized clinical trials were included. The revised Cochrane risk of bias tools for randomized trials were used to assess the quality of studies. Meta-analyses were performed. The overall quality of evidence was assessed through the Grading of Recommendations Assessment, Development, and Evaluation tool. Six studies were included. Five had a low risk of bias and 1 had some concerns. Three studies assessed bacterial reduction. Two were included in the meta-analysis for bacterial reduction (mean difference, 75.03 [confidence interval, CI, -271.15, 421.22], p = 0.67; I2 = 74%); and 3 in the meta-analysis for cultivable bacteria after chemomechanical preparation (odds ratio, 1.03 [CI, 0.20, 5.31], P = 0.98; I2 = 49%). Five studies assessed endotoxin reduction. Three were included in a meta-analysis (mean difference, 20.59 [CI, -36.41, 77.59], p = 0.48; I2 = 74%). There seems to be no difference in the disinfectant ability of CHX gel and NaOCl, but further research is necessary.

10.
J Endod ; 49(12): 1722-1732, 2023 Dec.
Artículo en Inglés | MEDLINE | ID: mdl-37717910

RESUMEN

INTRODUCTION: This study aimed to assess the canal transportation with 6 preparation systems. METHODS: Sixty curved mesial roots of mandibular molars were scanned, and their root canals (n = 20 per group) were prepared with XP-endo Shaper, BioRace, ProTaper Next, Twisted File Adaptive, Reciproc, or Reciproc Blue systems. After preparation, a new scan was conducted, and transportation was determined by comparing the geometric center of the canal in 13,542 cross-sectional slices. The transport direction frequencies were recorded, and linear transport were compared using a Univariate GLM model and Tukey honestly significant difference tests (α = 5 %). RESULTS: Transportation was affected by the preparation system and root third (P < .05), with no significant effect observed for the root canal (P > .05). The Twisted File Adaptive had the highest transport (P < .05), and the lowest transports were observed with XP-endo Shaper, ProTaper Next, and BioRace (P < .05). Reciproc Blue and Reciproc showed intermediate results (P < .05). The apical and coronal third exhibited the lowest and highest transportation, respectively (P < .05). A significant interaction was observed between the preparation system and root third (P < .05). In the coronal third, transportation was mostly toward the disto-inside direction, whereas in the middle and apical thirds, transportation was predominantly toward the mesio-outside direction. CONCLUSIONS: XP-endo Shaper had the lowest mean linear transportation followed by the ProTaper Next, BioRace, Reciproc Blue, and Reciproc systems. The Twisted File Adaptive system had the worst performance. The direction of transportation was generally toward the inner aspect of the canal at the coronal third and toward the outer aspect of the curve at the middle and apical thirds.


Asunto(s)
Cavidad Pulpar , Preparación del Conducto Radicular , Cavidad Pulpar/diagnóstico por imagen , Preparación del Conducto Radicular/métodos , Microtomografía por Rayos X/métodos , Estudios Transversales , Tratamiento del Conducto Radicular , Diseño de Equipo
11.
Restor Dent Endod ; 48(3): e25, 2023 Aug.
Artículo en Inglés | MEDLINE | ID: mdl-37675447

RESUMEN

Objectives: The purpose of this study was to evaluate the influence of endodontic access cavities design on the removal of calcium hydroxide medication of the apical third of mandibular incisor root canal walls and dentinal tubules with different cleaning protocols: EDDY sonic activation, Er,Cr:YSGG laser-activated irrigation, or conventional irrigation with IrriFlex. Materials and Methods: Seventy-eight extracted human mandibular incisors were assigned to 6 experimental groups (n = 13) according to the endodontic access cavity and cleaning protocol for calcium hydroxide removal: traditional access cavity (TradAC)/EDDY; ultraconservative access cavity performed in the incisal edge (UltraAC.Inc)/EDDY; TradAC/Er,Cr:YSGG; UltraAC.Inc/Er,Cr:YSGG; TradAC/IrriFlex; or UltraAC.Inc/IrriFlex. Confocal laser scanning microscopy images were used to measure the non-penetration percentage, maximum residual calcium hydroxide penetration depth, and penetration area at 2 and 4 mm from the apex. Data were statistically analyzed using Shapiro-Wilk and WRS2 package for 2-way comparison of non-normally distributed parameters (depth of penetration, area of penetration, and percentage of non-penetration) according to cavity and cleaning protocol with the significance level set at 5%. Results: The effect of cavity and cleaning protocol interactions on penetration depth, penetration area and non-penetration percentage was not found statistically significant at 2 and 4 mm levels (p > 0.05). Conclusions: The present study demonstrated that TradAC or UltraAC.Inc preparations with different cleaning protocols in extracted mandibular incisors did not influence the remaining calcium hydroxide at 2 and 4 mm from the apex.

12.
Int Endod J ; 56(11): 1412-1428, 2023 Nov.
Artículo en Inglés | MEDLINE | ID: mdl-37698946

RESUMEN

AIM: To compare the design, metallurgy, and mechanical properties of four heat-treated reciprocating instruments coupled with the evaluation of the irrigation flow using an in silico model. METHODOLOGY: New EdgeOne Fire Primary, Easy-File Flex Regular 25, WaveOne Gold Primary and Reciproc Blue R25 instruments (n = 124) were initially evaluated regarding their design through stereomicroscopy, scanning electron microscopy and 3D surface scanning. In addition, energy-dispersive X-ray spectroscopy was utilized to determine their elemental composition, and differential scanning calorimetry tests to evaluate their phase transformation temperatures. Their mechanical performance was further assessed through torsional and bending tests. Using scans obtained from a real tooth and the instruments, a computational fluid dynamics assessment was conducted to determine the irrigation flow pattern, apical pressure, and wall shear stress in simulated canal preparation. Mood's median and One-way anova post hoc Tukey tests were used for statistical comparisons (α = 5%). RESULTS: Reciproc Blue exhibited a superior number of blades (n = 8), whereas EdgeOne Fire had the highest overall volume (4.38 mm3 ) and surface area (32.32 mm2 ). At the 3-mm axial level, EdgeOne Fire displayed the lowest core diameter (0.13 mm), while Reciproc Blue had the highest (0.16 mm). All blades were symmetrical, and the tips of the instruments were non-active but differed from each other. The most irregular surfaces were observed in EdgeOne Fire and Easy-File Flex. All instruments were manufactured from nickel-titanium alloys and exhibited distinct phase transformation temperatures. WaveOne Gold and Reciproc Blue demonstrated the highest maximum torque values (1.87 and 1.62 N cm, respectively), while the lowest was observed on EdgeOne Fire (1.21 N cm) (p < .05). The most flexible (p < .05) were EdgeOne Fire (angle of rotation: 602.6°; maximum bending load: 251.4 g.f) and Reciproc Blue (533.2° and 235.6 g.f). There were no significant differences observed in the irrigation flow among the four domains generated by the tested instruments. CONCLUSIONS: Despite observing variations in the design, phase transformation temperatures, and in the torsional and bending test outcomes among the four heat-treated reciprocating instruments, no significant differences were found in the irrigation flow pattern among the different groups in the simulated root canal preparations.


Asunto(s)
Instrumentos Dentales , Calor , Diseño de Equipo , Estrés Mecánico , Preparación del Conducto Radicular , Titanio/química , Metalurgia , Oro , Ensayo de Materiales
13.
J Endod ; 2023 Aug 25.
Artículo en Inglés | MEDLINE | ID: mdl-37633518

RESUMEN

INTRODUCTION: Bisphosphonates are antiresorptive drugs used worldwide to treat systemic bone pathologies. This study aimed to assess the impact of zoledronic acid on the progression of induced apical periodontitis and the expression of cytokines interleukin (IL)-1ß, IL-10, IL-6, and tumor necrosis factor alpha (TNF-α) in a mouse model. METHODS: Sixteen female isogenic BALB/c mice 6 weeks of age were distributed into 2 groups: mice with induced apical periodontitis (the AP group, n = 8) and mice with induced apical periodontitis treated with zoledronic acid (the AP-ZA group, n = 8). The AP-ZA group received a dose of 125 µg/kg zoledronic acid diluted in sterile saline solution administered intraperitoneally once a week for 4 weeks before pulp exposure, whereas the AP group received only saline solution. Pulp exposures were performed on the maxillary first molars for the induction of apical periodontitis, and mice were euthanized after 7 and 21 days. The jaws were collected; scanned using micro-computed tomographic imaging; and processed for polymerase chain reaction analysis of IL-1ß, IL-10, IL-6, and TNF-α. The Student t test was performed for parametric data, and Mann-Whitney U tests were used for nonparametric data. The level of significance was set at 5%. RESULTS: Micro-computed tomographic imaging revealed higher bone resorption in the AP group compared with the AP-ZA group at both time points (P < .05). Real-time polymerase chain reaction demonstrated higher TNF-α expression in the AP group at both time points and higher IL-6 and IL-1ß expression in the AP group at the 7- and 21-day time points, respectively, compared with the AP-ZA group (P < .05). No differences were observed regarding IL-10 expression between the groups. CONCLUSIONS: Zoledronic acid had significant anti-inflammatory and antiresorptive effects on apical periodontitis in mice.

14.
Aust Endod J ; 49(3): 700-710, 2023 Dec.
Artículo en Inglés | MEDLINE | ID: mdl-37489612

RESUMEN

This study evaluated unprepared canal areas, volume of dentine removal, and dentine thickness after the progressive enlargement of the buccal canals of 22 maxillary molars, with and without the MB2, using instruments with the same tip size (0.25 mm) but 4 different tapers (0.03, 0.05, 0.06 and 0.08 v). Micro-CT scans were performed after each instrument. Data were compared using a general linear mixed model post hoc comparisons with Kenward-Roger for Wald F tests (α = 5%). After each treatment step, it was observed an increase in the percentage volume of removed dentine and a decrease in the unprepared areas and dentine thickness at the pericervical region of all canals (p < 0.05). The percentage reduction of dentine thickness in MB2 canal was higher than in the mesiobuccal and distobuccal canals (p < 0.05). Knowledge of pre-operative dimensions and dentine thickness before enlargement of buccal canals of maxillary molars is an important factor to avoid excessive dentine removal.


Asunto(s)
Cavidad Pulpar , Diente Molar , Humanos , Microtomografía por Rayos X , Cavidad Pulpar/diagnóstico por imagen , Diente Molar/diagnóstico por imagen , Hipertrofia , Conocimiento , Preparación del Conducto Radicular
15.
BMC Oral Health ; 23(1): 385, 2023 06 12.
Artículo en Inglés | MEDLINE | ID: mdl-37308853

RESUMEN

BACKGROUND: Considering the irregular shape of the root canal, removing inflamed pulp and granulation tissue completely from internal resorption cavities during chemomechanical preparation can be challenging. This study aimed to evaluate the effectiveness of passive ultrasonic irrigation (PUI) compared to mechanical activation with Easy Clean in the removal of organic tissue from simulated areas of internal root resorption. METHODS: The root canals of 72 extracted single-rooted teeth with oval canals were instrumented with Reciproc R25 instruments. After root canal preparations, the specimens were split longitudinally, and semicircular cavities were prepared using a round bur on each half of the roots. Samples obtained from bovine muscle tissue were weighed and adapted into semicircular cavities. The roots were reassembled and joined, and the teeth were divided into six groups (n = 12) according to the irrigation protocol: Sodium hypochlorite (NaOCl) without activation; NaOCl + PUI; NaOCl + Easy Clean; distilled water without activation; distilled water + PUI; and distilled water + Easy Clean. After irrigation protocols, the teeth were disassembled, and the remaining organic tissue was weighed. Data were analyzed by two-way ANOVA and Tukey's post hoc test (p < 0.05). RESULTS: None of the experimental protocols totally removed the bovine tissue from simulated cavities. Tissue weight reduction was significantly affected by the activation method (p < 0.05) and by irrigation solution (p < 0.05). Groups with NaOCl irrigation presented higher tissue weight loss when compared to distilled water, for all irrigation methods (p < 0.05). The use of Easy Clean resulted in the greatest tissue weight loss (42.0%-Distilled water/45.5%- NaOCl) compared to those of PUI (33.3%-Distilled water/37.7%- NaOCl) and no activation (33.4%-Distilled water/38.8%- NaOCl) (p < 0.05). However, no differences were observed between PUI and no activation groups (p > 0.05). CONCLUSIONS: Mechanical activation with Easy Clean enhanced organic tissue removal from simulated internal resorption more effectively than PUI. Easy Clean for agitation of the irrigating solution is effective in removing simulated organic tissues from artificial internal resorption cavities, being an alternative to the use of PUI.


Asunto(s)
Caries Dental , Resorción Radicular , Humanos , Animales , Bovinos , Análisis de Varianza , Pulpa Dental , Preparación del Conducto Radicular
16.
J Endod ; 49(5): 487-495, 2023 May.
Artículo en Inglés | MEDLINE | ID: mdl-36848949

RESUMEN

INTRODUCTION: This study aimed to compare the in vivo accuracy and precision of 3 electronic apex locators (EALs) in determining the position of the major foramen using micro-computed tomographic (micro-CT) technology. METHODS: After access preparation of 23 necrotic or vital teeth from 5 patients, canals were negotiated, and hand files were used to determine the position of the foramen with 3 EALs: Propex Pixi (Dentsply Maillefer, Ballaigues, Switzerland), Woodpex III (Woodpecker Medical Instrument Co, Guilin, China), and Root ZX II (J Morita, Tokyo, Japan). After fixing the silicon stop to the file, teeth were extracted and scanned in a micro-CT device with and without the instrument inserted into the canal. Data sets were coregistered, and the accuracy and precision of the EALs were determined at a tolerance level of ±0.5 mm by measuring the distance from the tip of the instruments to a tangential line crossing the margins of the foramen. Statistical comparisons were performed using Friedman with post hoc related samples sign and Spearman tests (α = 5%). RESULTS: A significant difference was detected comparing the accuracy of Root ZX II (100%), Woodpex III (86.96%), and Propex Pixi (52.17%) (P < .05). There was a lack of significance in the relationship between the pulp status and the accuracy of the tested EALs (P > .05). Propex Pixi was significantly less precise than Root ZX II (P < .05), whereas no difference was found between Woodpex III and Root ZX II or Propex Pixi (P > .05). CONCLUSIONS: EALs presented similar precision, but Woodpex III and Root ZX II showed better accuracy to determine the position of the apical major foramen than Propex Pixi.


Asunto(s)
Cavidad Pulpar , Preparación del Conducto Radicular , Humanos , Cavidad Pulpar/diagnóstico por imagen , Odontometría , Ápice del Diente/diagnóstico por imagen , Electrónica
17.
Int Endod J ; 56(4): 530-542, 2023 Apr.
Artículo en Inglés | MEDLINE | ID: mdl-36508297

RESUMEN

AIM: To compare design, metallurgy and mechanical performance of the ProTaper (PT) Ultimate system with instruments of similar dimensions from the ProGlider, PT Gold and PT Universal systems. METHODOLOGY: New PT Ultimate instruments (n = 248) were compared with instruments of similar dimensions from ProGlider (n = 31), PT Gold (n = 155) and PT Universal (n = 155) systems regarding their number of spirals, helical angle, blade symmetry, tip geometry, surface finishing, nickel/titanium ratio, phase transformation temperatures and mechanical performance. One-way anova and nonparametric Mood's median tests were used for statistical comparison (α = 5%). RESULTS: All instruments had symmetrical blades without radial lands or flat sides, similar surface finishing and an almost equiatomic nickel/titanium ratio, whilst the number of spirals, helical angles and the tip geometry were different. PT Ultimate instruments showed 3 distinct heat treatments that matched with the colour of their metal wire. Slider and ProGlider instruments had similar R-phase start (Rs) and R-phase finish (Rf) temperatures. SX, F1, F2, F3 and Shaper instruments showed equivalent heat treatments (Rs ~45.6°C and Rf ~28.3°C) that were similar to their PT Gold counterparts (Rs ~47.9°C and Rf ~28.2°C), but completely distinct to the PT Universal ones (Rs ~16.2°C and Rf ~-18.2°C). Amongst the PT Ultimate instruments, the lowest maximum torques were observed in the SX (0.44 N cm), Slider (0.45 N cm) and Shaper (0.60 N cm) instruments, whilst the highest was noted in the FXL (4.90 N cm). PT Ultimate Slider and ProGlider had similar torsional (~0.40 N cm) and bending loads (~145.0 gf) (p = 1.000), whilst the other PT Ultimate instruments showed statistically significantly lower maximum torque, higher angle of rotation and lower bending load (higher flexibility) than their counterparts of the PT Universal and PT Gold systems. CONCLUSIONS: The PT Ultimate system comprises instruments with 3 distinct heat treatments that showed similar phase transformation temperatures to their heat-treated analogues. PT Ultimate instruments presented lower torsional strength and superior flexibility than their counterparts, whilst maximum torque, angle of rotation and bending loads progressively increased with their sizes.


Asunto(s)
Níquel , Titanio , Calor , Falla de Equipo , Ensayo de Materiales , Aleaciones Dentales , Preparación del Conducto Radicular , Diseño de Equipo , Torsión Mecánica
18.
Aust Endod J ; 49 Suppl 1: 515-527, 2023 Sep.
Artículo en Inglés | MEDLINE | ID: mdl-36514265

RESUMEN

This systematic review assessed the clinical influence of centric and eccentric engine-driven instrument design for reduction in microbial load in infected root canal system. The literature search was conducted in electronic databases (PubMed, Cochrane Library, Scopus, Lilacs and Google Scholar) and grey literature till June 2022. The search strategy followed the PRISMA 2020 guidelines. Qualitative and quantitative synthesis was conducted based on the guidelines of the Cochrane Handbook. The risk of bias was assessed using the revised Cochrane criteria and quality of evidence was conducted using the Grading of Recommendation Assessment, Development and Evaluation tool. Out of 28 papers, only five papers met the inclusion criteria of this review. Studies showed reduction in microbial load after instrumentation using centric or eccentric instruments with no statistically significant difference in the meta-analysis. In conclusion, there is low-grade evidence suggesting that microbial reduction is similar using centric and eccentric instruments.


Asunto(s)
Tratamiento del Conducto Radicular , Bases de Datos Factuales , Ensayos Clínicos Controlados Aleatorios como Asunto
19.
Aust Endod J ; 49(2): 256-261, 2023 Aug.
Artículo en Inglés | MEDLINE | ID: mdl-35770610

RESUMEN

This study aimed to evaluate the shaping ability of XP-endo Shaper, TruNatomy and EdgeFile X3 during the preparation of resin-printed mandibular molar mesial root canals. Thirty-three resin-based mandibular mesial roots with two canals, obtained from extracted tooth cone-beam computed tomography (CBCT) image and printed on a three-dimensional (3D) printer, were divided into three experimental groups according to the different nickel-titanium (NiTi) systems used for root canal preparation. The specimens were scanned using CBCT imaging before and after root canal preparation. Then images were registered using a dedicated software and changes in the canal area, volume, untouched canal surface and the maximum and minimum dentine wall wear were calculated. The XP-endo Shaper instruments showed improved shaping ability with lower untouched root canal surface and better preservation of root canal anatomy during the preparation of resin-printed mandibular mesial root canals compared with TruNatomy and EdgeFile X3 systems.


Asunto(s)
Endometriosis , Níquel , Humanos , Femenino , Titanio , Microtomografía por Rayos X , Preparación del Conducto Radicular , Cavidad Pulpar/diagnóstico por imagen , Diente Molar/diagnóstico por imagen , Diseño de Equipo
20.
Aust Endod J ; 49 Suppl 1: 194-201, 2023 Sep.
Artículo en Inglés | MEDLINE | ID: mdl-36401559

RESUMEN

This study assessed the influence of pH variation on the volumetric change of EndoSequence BC Sealer compared to AH Plus Jet. Eighteen single-rooted teeth were uniformly prepared and filled with one of the sealers. After 24-h storing, samples were scanned in a micro-CT device. Then, roots were immersed in 20 ml phosphate-buffered saline (PBS) with pH of 5, 7 or 12 at 37°C for 7 and 30 days and rescanned. Statistical analysis was performed using Shapiro-Wilk's test, one-way ANOVA and Tukey's or Student's t-tests (p < 0.05). Differences between sealers were observed only at neutral pH and after 30 days of acidic pH exposure, with a higher loss of EndoSequence (p < 0.05). After 7 days, alkaline pH reduced and increased the volumetric loss of EndoSequence and AH Plus, respectively, compared to neutral pH (p < 0.05). It can be concluded that the pH directly influenced the volumetric change of both tested root canal sealers in different ways.


Asunto(s)
Resinas Epoxi , Materiales de Obturación del Conducto Radicular , Humanos , Microtomografía por Rayos X , Compuestos de Calcio , Silicatos , Concentración de Iones de Hidrógeno , Ensayo de Materiales , Cavidad Pulpar/diagnóstico por imagen
SELECCIÓN DE REFERENCIAS