RESUMEN
Extrapolation chambers are instruments designed to measure doses of low-energy radiations, mainly beta radiation. In this work, a commercial extrapolation chamber and a homemade extrapolation chamber were applied in measurements using standard radiotherapy X-ray beams. Saturation curves and polarity effect as well as short- and medium-term stabilities were obtained, and these results are within the recommendations of the International Electrotechnical Commission (IEC). The response linearity and the extrapolation curves were also obtained, and they presented good behavior. The results show the usefulness of these extrapolation chambers in low-energy X-ray beams.
RESUMEN
The extrapolation chamber is a parallel-plate ionization chamber that allows variation of its air-cavity volume. In this work, an experimental study and MCNP-4C Monte Carlo code simulations of an ionization chamber designed and constructed at the Calibration Laboratory at IPEN to be used as a secondary dosimetry standard for low-energy X-rays are reported. The results obtained were within the international recommendations, and the simulations showed that the components of the extrapolation chamber may influence its response up to 11.0%.