JACC Cardiovasc Interv ; 15(4): 368-377, 2022 02 28.
Artículo en Inglés | MEDLINE | ID: mdl-35210043

RESUMEN

OBJECTIVES: This study sought to determine the degree of Evolut (Medtronic) leaflet pinning, diameter expansion, leaflet overhang, and performance at different implant depths of the balloon-expandable Sapien 3 (S3, Edwards Lifesciences LLC) transcatheter heart valve (THV) within the Evolut THV. BACKGROUND: Preservation of coronary access and flow is a major factor when considering the treatment of failed Evolut THVs. METHODS: An in vitro study was performed with 20-, 23-, 26-, and 29-mm S3 THVs deployed within 23-, 26-, 29-, and 34-mm Evolut R THVs, respectively. The S3 outflow was positioned at various depths at node 4, 5, and 6 of the Evolut R. Neoskirt height, leaflet overhang, performance, and Evolut R valve housing diameter expansion were assessed under physiological conditions as per ISO 5840-3 standard. RESULTS: The neoskirt height for the Evolut R was shorter when the S3 outflow was positioned at node 4 compared with node 6 (node 4 height for 23 mm = 16.3 mm, 26 mm = 17.1 mm, 29 mm = 18.3 mm, and 34 mm = 19.9 mm vs node 6 height for 23 mm = 23.9 mm, 26 mm = 23.4 mm, 29 mm = 24.7 mm, and 34 mm = 27 mm Evolut R). All configurations exhibited acceptable hydrodynamic performance irrespective of the degree of leaflet overhang, except the 29-mm S3 implanted in 34-mm Evolut R at node 4 (regurgitant fraction >20%). The valve housing radius of the index Evolut R increased when the S3 was implanted, with the increase ranging from 0 to 2.5 mm. CONCLUSIONS: Placement of the S3 at a lower implant position within an index Evolut R reduces the neoskirt height with no significant compromise to S3 valve function despite a higher degree of leaflet overhang. Low S3 implantation may facilitate future coronary access after redo transcatheter aortic valve replacement.


Asunto(s)
Estenosis de la Válvula Aórtica , Prótesis Valvulares Cardíacas , Reemplazo de la Válvula Aórtica Transcatéter , Válvula Aórtica/diagnóstico por imagen , Válvula Aórtica/cirugía , Estenosis de la Válvula Aórtica/cirugía , Humanos , Diseño de Prótesis , Reemplazo de la Válvula Aórtica Transcatéter/efectos adversos , Resultado del Tratamiento
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA
...