Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Más filtros











Base de datos
Intervalo de año de publicación
1.
EuroIntervention ; 15(1): 37-43, 2019 May 20.
Artículo en Inglés | MEDLINE | ID: mdl-30777843

RESUMEN

AIMS: The aim of this study was to evaluate transcatheter aortic valve-in-valve (ViV) implantation performance in rapid deployment (ViVr) vs. conventional (ViVc) surgical heart valves. METHODS AND RESULTS: A multicentre registry was developed as part of the VIVID international registry. A total of 30 ViVr patients (Perceval, n=24, ATS 3f Enable™, n=5, and the INTUITY, n=1) were evaluated and compared with 2,288 ViVc patients. Propensity score (PS) matching was performed to adjust further for bias. Compared with ViVc, ViVr patients presented twice as early after surgical heart valve (SHV) implantation (55.2±36.1 vs. 118.4±57.7 months, p<0.001), were more commonly female (82.8% vs. 41.3%, p<0.001), and had shorter body stature and reduced body weight (p<0.05 for both) prior to PS. Implantation was successful in all ViVr cases and, compared with ViVc, was associated with equally favourable haemodynamic outcomes (mean gradient: 14.6±8.3 vs. 16.2±8.9 mmHg, p=0.356; regurgitation ≥mild: 3.7% vs. 5.2%, p=0.793). Periprocedural complication rates were similar and low in both groups. There was no coronary obstruction event in any ViVr case; one patient (3.6%) died during one year of follow-up. CONCLUSIONS: ViVr appears effective, safe and associated with favourable haemodynamic outcome.


Asunto(s)
Estenosis de la Válvula Aórtica , Bioprótesis , Implantación de Prótesis de Válvulas Cardíacas , Prótesis Valvulares Cardíacas , Válvula Aórtica , Femenino , Humanos , Diseño de Prótesis , Resultado del Tratamiento
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA