1.
Radiat Prot Dosimetry ; 161(1-4): 417-21, 2014 Oct.
Artículo en Inglés | MEDLINE | ID: mdl-24255173

RESUMEN

Inside an IBA proton therapy centre, secondary neutrons are produced due to nuclear interactions of the proton beam with matter mainly inside the cyclotron, the beam line, the treatment nozzle and the patient. Accurate measurements of the neutron ambient dose equivalent H*(10) in such a facility require the use of a detector that has a good sensitivity for neutrons ranging from thermal energies up to 230 MeV, such as for instance the WENDI-2 detector. WENDI-2 measurements have been performed at the Westdeutsches Protonentherapiezentrum Essen, at several positions around the cyclotron room and around a gantry treatment room operated in two different beam delivery modes: Pencil Beam Scanning and Double Scattering. These measurements are compared with Monte Carlo simulation results for the neutron H*(10) obtained with MCNPX 2.5.0 and GEANT4 9.6.


Asunto(s)
Neutrones , Terapia de Protones/métodos , Radiometría/instrumentación , Radiometría/métodos , Algoritmos , Calibración , Simulación por Computador , Ciclotrones , Arquitectura y Construcción de Instituciones de Salud , Alemania , Humanos , Método de Montecarlo , Protones , Dosis de Radiación , Dosificación Radioterapéutica , Reproducibilidad de los Resultados , Dispersión de Radiación
2.
Radiat Prot Dosimetry ; 154(3): 340-5, 2013.
Artículo en Inglés | MEDLINE | ID: mdl-22972796

RESUMEN

The WENDI-II rem meter is one of the most popular neutron dosemeters used to assess a useful quantity of radiation protection, namely the ambient dose equivalent. This is due to its high sensitivity and its energy response that approximately follows the conversion function between neutron fluence and ambient dose equivalent in the range of thermal to 5 GeV. The simulation of the WENDI-II response function with the Geant4 toolkit is then perfectly suited to compare low- and high-energy hadronic models provided by this Monte Carlo code. The results showed that the thermal treatment of hydrogen in polyethylene for neutron <4 eV has a great influence over the whole detector range. Above 19 MeV, both Bertini Cascade and Binary Cascade models show a good correlation with the results found in the literature, while low-energy parameterised models are not suitable for this application.


Asunto(s)
Modelos Estadísticos , Método de Montecarlo , Neutrones , Radiometría/instrumentación , Radiometría/métodos , Programas Informáticos , Simulación por Computador , Diseño Asistido por Computadora , Diseño de Equipo , Análisis de Falla de Equipo , Dosis de Radiación , Reproducibilidad de los Resultados , Dispersión de Radiación , Sensibilidad y Especificidad