Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Más filtros











Base de datos
Intervalo de año de publicación
1.
Clin Chem Lab Med ; 61(7): 1235-1244, 2023 06 27.
Artículo en Inglés | MEDLINE | ID: mdl-36815732

RESUMEN

OBJECTIVES: In clinical pratice, tau protein measurement generally relies on immunoassays (IAs), whose major drawback is the lack of results comparability due to differences in selectivity and/or calibration. This underlines the importance of establishing a traceability chain for total tau (t-tau) measurements. The objective of this work is to develop a higher order candidate reference measurement procedure (RMP) for the absolute quantification of t-tau in cerebrospinal fluid (CSF). METHODS: To calibrate the candidate RMP and establish metrological traceability to the SI units, a primary calibrator consisting in a highly purified recombinant protein was sourced. Its purity was evaluated by liquid chromatography coupled with high resolution mass spectrometry (LC-HRMS) and the protein mass fraction in solution was certified by amino acid analysis (AAA). An isotopically-labelled homologue was obtained to develop a candidate RMP by isotope dilution mass spectrometry (IDMS) for t-tau absolute quantification in CSF. Calibration blends and quality control (QC) materials were gravimetrically prepared and subjected to the same preparation workflow as CSF samples, followed by LC-HRMS analysis in Parallel Reaction Monitoring (PRM) mode. RESULTS: A primary calibrator has been developed and an IDMS candidate RMP has been validated for CSF t-tau. The candidate RMP was used to certify t-tau concentration in three pools of CSF (low, medium, high). CONCLUSIONS: The candidate RMP will pave the road towards global standardization of CSF t-tau measurements. Together with commutable Certified Reference Materials (CRMs), it will allow evaluating and improving the accuracy and comparability of results provided by IAs.


Asunto(s)
Espectrometría de Masas en Tándem , Proteínas tau , Humanos , Cromatografía Liquida/métodos , Espectrometría de Masas en Tándem/métodos , Estándares de Referencia , Aminoácidos/análisis , Calibración
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA