Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Más filtros










Base de datos
Intervalo de año de publicación
1.
Appl Radiat Isot ; 86: 7-12, 2014 Apr.
Artículo en Inglés | MEDLINE | ID: mdl-24457420

RESUMEN

In the past few years, two Cerenkov methods were developed to make activity measurements of high-energy beta emitters in liquid scintillation counters with two or three photomultiplier tubes (PMTs) possible. Both methods are based on a free parameter model and make use of the Frank and Tamm theory for the emission of Cerenkov light. In this article, additional effects are discussed and further improvements are presented. The dependence of the refractive index of water on the wavelength can now be taken into account, which has also an influence on the upper limit of the wavelength region for the production of Cerenkov light. In addition, the dependence of the PMT response on the wavelength is taken into account. Finally, it is possible to take a potential asymmetry of efficiencies in a system with three PMTs into account. To this end, three free parameters are assigned to each individual PMT and then determined by means of a downhill simplex optimization algorithm. The computed counting efficiencies for a triple-to-double coincidence ratio (TDCR) system were compared with experimental data for (32)P, (89)Sr, and (90)Y.