Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters










Database
Main subject
Language
Publication year range
1.
Sci Rep ; 14(1): 14020, 2024 06 18.
Article in English | MEDLINE | ID: mdl-38890408

ABSTRACT

The study assessed the impact of procedural errors on the remote dielectric sensing system (ReDS), a non-invasive lung fluid assessment technology, in an Asian cohort. Healthy volunteers underwent ReDS measurements following manufacturer's instructions, with two consecutive measurements one minute apart. A subset of 20 participants had modified procedure settings. Reliability was measured using intraclass correlation coefficient (ICC). The study included 86 healthy volunteers, and all ReDS measurements fell within the recommended normal range. The intra-rater reliability of ReDS measurements was excellent, with an ICC of 0.968. Among the subset of 20 subjects, deviations in height and weight did not significantly affect ReDS values. However, deviations in chest size by ± 3 cm had a noticeable impact on ReDS measures, and incorrect station selection led to fluctuations in ReDS readings. In conclusion, the ReDS system demonstrated excellent intra-rater reliability and applicability in an Asian cohort. Procedural errors, such as chest size measurement and station selection, significantly influenced ReDS measurements. Adherence to standardized operating procedures is crucial to ensure accurate and consistent results. These findings highlight the importance of adherence to manufacturer instructions when utilizing ReDS for lung fluid assessment, thereby enhancing its reliability and clinical applicability.


Subject(s)
Lung , Humans , Male , Female , Adult , Lung/physiology , Reproducibility of Results , Remote Sensing Technology/methods , Healthy Volunteers , Young Adult , Middle Aged , Body Fluids , Electric Impedance
SELECTION OF CITATIONS
SEARCH DETAIL
...