Your browser doesn't support javascript.
loading
Montrer: 20 | 50 | 100
Résultats 1 - 1 de 1
Filtrer
Plus de filtres











Base de données
Gamme d'année
1.
Int J Hematol ; 103(5): 530-6, 2016 May.
Article de Anglais | MEDLINE | ID: mdl-26872908

RÉSUMÉ

Myocardial iron quantification remains limited to 1.5 T systems with T2* measurement. The present study aimed at comparing myocardial T2* values at 1.5 T to T1 and T2 mapping at 3.0 T in patients with iron overload and healthy controls. A total of 17 normal volunteers and seven patients with a history of myocardial iron overload were prospectively enrolled. Mid-interventricular septum T2*, native T1 and T2 times were quantified on the same day, using a multi-echo gradient-echo sequence at 1.5 T and T1 and T2 mapping sequences at 3.0 T, respectively. Subjects with myocardial iron overload (T2* < 20 ms) in comparison with those without had significantly lower mean myocardial T1 times (868.9 ± 120.2 vs. 1170.3 ± 25.0 ms P = 0.005 respectively) and T2 times (34.9 ± 4.7 vs. 45.1 ± 2.0 ms P = 0.007 respectively). 3 T T1 and T2 times strongly correlated with 1.5 T, T2* times (Pearson's r = 0.95 and 0.91 respectively). T1 and T2 measures presented less variability than T2* in inter- and intra-observer analysis. Native myocardial T1 and T2 times at 3 T correlate closely with T2* times at 1.5 T and may be useful for myocardial iron overload quantification.


Sujet(s)
Surcharge en fer/anatomopathologie , Fer/analyse , Imagerie par résonance magnétique , Myocarde/anatomopathologie , Adolescent , Adulte , Femelle , Humains , Imagerie par résonance magnétique/méthodes , Mâle , Adulte d'âge moyen , Valeur prédictive des tests , Reproductibilité des résultats
SÉLECTION CITATIONS
DÉTAIL DE RECHERCHE