Heart Vessels ; 36(5): 638-645, 2021 May.
Article in English | MEDLINE | ID: mdl-33389064

ABSTRACT

The 4-Fr catheter system is not recommended for invasive functional assessment of coronary artery stenosis, because it tends to distort the aortic waveform. This study aimed to identify the incidence of aortic waveform distortion and a feasible method for correct diagnosis of coronary artery stenosis with a 4-Fr catheter. We retrospectively investigated 178 lesions with intermediate coronary artery stenosis. Non-hyperemic distal coronary artery pressure (Pd) and aortic pressure (Pa) were measured with a 4-Fr diagnostic or 6-Fr guiding catheter before and after saline flush. The mean Pd/mean Pa (Pd/Pa) and instantaneous wave-free ratio (iFR) were calculated before and after flushing. We compared the effect of flushing on the changes in Pd/Pa and iFR between the 4-Fr diagnostic and 6-Fr guiding catheters. Using the 4-Fr diagnostic catheter, there was a significant decrease in incidence of aortic waveform distortion from 42.0% (47 lesions) before flushing to 1.8% (2 lesions) after flushing (p < 0.001); the incidence was only 3.0% before saline flush and decreased to 0% after saline flush when using the 6-Fr guiding catheter. The presence of aortic waveform distortion influenced the iFR when the 4-Fr system was used. Functional measurements with the 4-Fr diagnostic catheter require adequate saline flush to remove the influence of aortic waveform distortion.


Subject(s)
Cardiac Catheterization/methods , Coronary Angiography/methods , Coronary Stenosis/physiopathology , Fractional Flow Reserve, Myocardial/physiology , Aged , Coronary Stenosis/diagnosis , Coronary Vessels/physiopathology , Female , Follow-Up Studies , Humans , Male , ROC Curve , Retrospective Studies , Severity of Illness Index
SELECTION OF CITATIONS
SEARCH DETAIL