Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters










Database
Language
Publication year range
1.
ACS Biomater Sci Eng ; 2024 Aug 06.
Article in English | MEDLINE | ID: mdl-39107258

ABSTRACT

Poly(methyl methacrylate) (PMMA) bone cements have been widely used in orthopedics; thanks to their excellent mechanical properties, biocompatibility, and chemical stability. Barium sulfate and zirconia are usually added into PMMA bone cement to enhance the X-ray radiopacity, while the mechanical strength, radiopacity, and biocompatibility are not well improved. In this study, an insoluble and corrosion-resistant ceramic, tantalum carbide (TaC), was added into the PMMA bone cement as radiopacifies, significantly improving the mechanical, radiopaque, biocompatibility, and osteogenic performance of bone cement. The TaC-PMMA bone cement with varied TaC contents exhibits compressive strength over 100 MPa, higher than that of the commercial 30% BaSO4-PMMA bone cement. Intriguingly, when the TaC content reaches 20%, the radiopacity is equivalent to the commercial bone cement with 30% of BaSO4 in PMMA. The cytotoxicity and osteogenic performance indicate that the incorporation of TaC not only enhances the osteogenic properties of PMMA but also does not reduce cell viability. This study suggests that TaC could be a superior and multifunctional radio-pacifier for PMMA bone cement, offering a promising avenue for improving patient outcomes in orthopedic applications.